GASTROINTESTINAL FIBERSCOPE OLYMPUS GIF TYPE E3 (UDI 04953170307355)Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 04953170307355), received between Oct 2, 2020 and Sep 11, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Failure to Clean Adequately” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 2 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Clean Adequately4
  • Corroded1
  • Mechanical Jam1
  • Microbial Contamination of Device1
  • Poor Quality Image1
  • Thermal Decomposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 11, 2026Thermal Decomposition of Device

    The customer reported a burn on the plastic distal end cover during inspection for use involving an Olympus gastrointestinal fiberscope. There was no patient injury reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50738643 ↗
  • MalfunctionJun 10, 2026Failure to Clean Adequately

    THE CUSTOMER RETURNED THE GASTROINTESTINAL FIBERSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPORT TECHNICIAN INDICATES THAT DIRT IN THE FIELD VISION WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25460154 ↗
  • MalfunctionFeb 16, 2026Corroded; Failure to Clean Adequately

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL FIBERSCOPE HAD DIRT IN THE OPTICAL PART AND RUST UNDER THE LENSES (BACKLIGHT). THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24343645 ↗
  • MalfunctionOct 15, 2025Mechanical Jam; Mechanical Jam

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE GASTROINTESTINAL FIBERSCOPE EXHIBITED A BLOCKED UP/DOWN HANDLE STOPPER. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23297286 ↗
  • MalfunctionOct 15, 2025Poor Quality Image; Poor Quality Image

    IT WAS REPORTED THAT THE SUBJECT DEVICE HAD BLURRY IMAGE. THE ISSUE WAS FOUND DURING SERVICE / MAINTENANCE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23276665 ↗
  • MalfunctionDec 9, 2022Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS THAT THE WORKING CHANNEL WAS BLOCKED ON THE GASTROINTESTINAL FIBERSCOPE. THE ISSUE OCCURRED DURING REPROCESSING AND THERE WAS NO PATIENT HARM ASSOCIATED WITH THE EVENT. THE DEVICE WAS RETURNED AND EVALUATED, AND THE NOZZLE WAS FOUND TO BE CLOGGED WITH FOREIGN MATERIAL. THIS MDR IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING THE DEVICE EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15950011 ↗
  • MalfunctionNov 11, 2022Failure to Clean Adequately; Failure to Clean Adequately

    THE CUSTOMER REPORTED TO OLYMPUS, THE GASTROINTESTINAL FIBERSCOPE HAD AN IMAGE PROBLEM. INSPECTION AND TESTING OF THE RETURNED DEVICE FOUND THERE WAS DIRT ON THE BENDING SECTION RUBBER, THE GRIP, THE RIGHT/LEFT KNOB, AND THE UP/DOWN KNOB, WHICH IS INDICATIVE OF INSUFFICIENT OR INCORRECT REPROCESSING. THERE WERE NO REPORTS OF PATIENT HARM. THIS MEDICAL DEVICE REPORT (MDR) IS BEING SUBMITTED TO CAPTURE THE REPORTABLE MALFUNCTION FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15782630 ↗
  • MalfunctionOct 2, 2020Microbial Contamination of Device; Microbial Contamination of Device

    OLYMPUS MEDICAL SYSTEMS CORP. (OMSC) WAS INFORMED THAT AS A RESULT OF MICROBIOLOGICAL TESTING BY THE USER FACILITY, FOLLOWING MICROBES WERE DETECTED FROM THE SAMPLE COLLECTED FROM THE DEVICE. DISTAL END: STAPHYLOCOCCUS WARNERI (1 CFU) INSTRUMENT CHANNEL: ESCHERICHIA COLI (68 CFU), PSEUDOMONAS AERUGINOSA (41 CFU), SERRATIA MARCESCENS (2 CFU) THE DEVICE HAD BEEN REPROCESSED WITH A STEELCO AUTOMATED ENDOSCOPE REPROCESSOR MODEL EW 2/2, USING NEODISHER SEPTO PAC. THERE WAS NO REPORT OF INFECTION ASSOCIATED WITH THIS REPORT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10620155 ↗

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →