EVIS EXERA SIGMOIDOVIDEOSCOPE OLYMPUS CF TYPE Q160S (UDI 04953170339134)Recalls & reported problems
Olympus Medical Systems Corp. · CF-Q160S LARGE IMAGE VIDEO SIGMOIDOSCOPE · Class 2
All models of EVIS EXERA SIGMOIDOVIDEOSCOPE (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 19
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 19 adverse event reports that list this exact device (DI 04953170339134), received between Apr 19, 2024 and Jul 29, 2026 — classified as 19 malfunctions.
The most frequently reported specific device problem is “No Display/Image” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- No Display/Image9
- Residue After Decontamination4
- Device Reprocessing Problem3
- Communication or Transmission Problem1
- Contamination1
- Erratic or Intermittent Display1
- Improper or Incorrect Procedure or Method1
- Poor Quality Image1
Patient problems
- No Clinical Signs, Symptoms or Conditions19
Recent adverse event reports
The 19 most recent of 19 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 29, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50065964 ↗MalfunctionJul 28, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50057091 ↗MalfunctionJul 27, 2026Erratic or Intermittent Display
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50043270 ↗MalfunctionJun 12, 2026Residue After Decontamination
THE CUSTOMER RETURNED THE SIGMOIDOVIDEOSCOPE TO THE SERVICE CENTER FOR EVALUATION RELATED TO A SEPARATE ISSUE. DURING TESTING, THE SERVICE TECHNICIAN IDENTIFIED THE DEVICE HAD DARK DEBRIS IN LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25488451 ↗MalfunctionMay 12, 2026Residue After Decontamination
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE EXHIBITED WHITE FOREIGN MATERIAL IN AIR-WATER CHANNEL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25148039 ↗MalfunctionJan 28, 2026Contamination
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE HAD BLACK FOREIGN MATERIAL IN THE AIR/WATER NOZZLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24204562 ↗MalfunctionJan 28, 2026No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE HAD NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24203489 ↗MalfunctionJan 2, 2026Residue After Decontamination
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE EXHIBITED WHITE RESIDUE IN THE COVER GLASS AND LIGHT GUIDE ROD. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23946278 ↗MalfunctionDec 26, 2025No Display/Image; No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE EXHIBITED INTERMITTENT IMAGE AND NO IMAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23896761 ↗MalfunctionDec 22, 2025No Display/Image; No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE HAD NO IMAGE WITH BLACK SCREEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23863688 ↗MalfunctionDec 17, 2025No Display/Image; Poor Quality Image; No Display/Image; Poor Quality Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE EXHIBITED NO IMAGE, A BLACK SCREEN, AND HORIZONTAL LINES ON THE SCREEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23828276 ↗MalfunctionNov 25, 2025Communication or Transmission Problem; Communication or Transmission Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THE SUBJECT DEVICE HAD AN ERROR CODE MALFUNCTION (B30). THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23653374 ↗MalfunctionNov 25, 2025No Display/Image; No Display/Image
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE SIGMOIDOVIDEOSCOPE EXHIBITED A LOSS OF IMAGE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23644343 ↗MalfunctionNov 20, 2025Residue After Decontamination; Residue After Decontamination
IT WAS OBSERVED THAT DURING EVALUATION, THE SIGMOIDOVIDEOSCOPE EXHIBITED A DARK RESIDUE ON THE LIGHT GUIDE LENS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23612890 ↗MalfunctionNov 10, 2025No Display/Image; No Display/Image
IT WAS REPORTED THAT THE SUBJECT DEVICE HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23511440 ↗MalfunctionNov 5, 2025No Display/Image; No Display/Image
IT WAS REPORTED THE SIGMOIDOVIDEOSCOPE EXPERIENCED A PROBLEM OF BEING UNABLE TO VIEW SCREEN. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23476859 ↗MalfunctionSep 25, 2024Device Reprocessing Problem; Improper or Incorrect Procedure or Method; Device Reprocessing Problem; Improper or Incorrect Procedure or Method
OLYMPUS WAS INFORMED THAT, DURING AN ONSITE VISIT, THE USER FACILITY STAFF WAS NOT SUCTIONING ANY OF THEIR SCOPES DUE TO LACK OF A SUCTIONING SOURCE. THEY HAVE NOT BEEN REPROCESSING ACCESSORIES SUCH AS THE INJECTION TUBING OR SUCTION CLEANING ADAPTER. THEY HAVE BEEN DEVIATING FROM INSTRUCTIONS FOR USE BY NOT PERFORMING THE STEPS CORRECTLY. FLUSHING WAS BEING PERFORMED WITH A SYRINGE RATHER THAN INJECTION TUBING, THEY HAVE NOT FOLLOWED DELAYED REPROCESSING, AND AT TIMES WERE REPROCESSING THE SCOPES THE FOLLOWING DAY AFTER A PROCEDURE. NO PATIENT INFECTIONS OR INJURIES REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20299667 ↗MalfunctionJul 15, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS REPORTED THE SIGMOIDOVIDEOSCOPE HAD INTERNAL DEFECTS OF CHANNEL BLOCKED WATER CHANNEL. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19745423 ↗MalfunctionApr 19, 2024Device Reprocessing Problem; Device Reprocessing Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION THAT THE GI SCOPE HAD A CLOGGED NOZZLE WITH BLACK DEBRIS INSIDE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19149211 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.