EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F (UDI 04953170339844)Recalls & reported problems
Olympus Medical Systems Corp. · JF-140F DUODENOVIDEOSCOPE (FIXED CAP) · Class 2
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 04953170339844), received between May 19, 2020 and Jul 17, 2026 — classified as 8 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Device Reprocessing Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Device Reprocessing Problem5
- Adverse Event Without Identified Device or Use Problem4
- Microbial Contamination of Device4
- Crack1
- Improper or Incorrect Procedure or Method1
- Peeled/Delaminated1
Patient problems
- Bacterial Infection8
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2026Crack; Peeled/Delaminated
THE CUSTOMER RETURNED THE DUODENOVIDEOSCOPE TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE REPAIR TECHNICIAN INDICATED THAT A CHIP, CRACK, PEELING ON CEMENT ON OB LENS WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25858451 ↗MalfunctionMay 5, 2021Device Reprocessing Problem; Device Reprocessing Problem
IT IS REPORTED IN THE LITERATURE ARTICLE TITLED "RESIDUAL MOISTURE AND WATERBORNE PATHOGENS INSIDE FLEXIBLE ENDOSCOPES: EVIDENCE FROM A MULTISITE STUDY OF ENDOSCOPE DRYING EFFECTIVENESS," 22 DIFFERENT MODEL SCOPES TESTED POSITIVE FOR MICROORGANISMS AFTER REPROCESSING. BACKGROUND: ENDOSCOPY-ASSOCIATED INFECTION TRANSMISSION IS FREQUENTLY LINKED TO INADEQUATE REPROCESSING. RESIDUAL ORGANIC MATERIAL AND MOISTURE MAY FOSTER BIOFILM DEVELOPMENT INSIDE ENDOSCOPES. THIS STUDY EVALUATED THE EFFECTIVENESS OF ENDOSCOPE DRYING AND STORAGE METHODS AND ASSESSED ASSOCIATIONS BETWEEN RETAINED MOISTURE AND CONTAMINATION. METHODS: ENDOSCOPE REPROCESSING, DRYING, AND STORAGE PRACTICES WERE ASSESSED AT 3 HOSPITALS. RESEARCHERS PERFORMED VISUAL EXAMINATIONS AND TESTS TO DETECT FLUID AND CONTAMINATION ON PATIENT-READY ENDOSCOPES. RESULTS: FLUID WAS DETECTED IN 22 OF 45 (49%) ENDOSCOPES. PREVALENCE OF MOISTURE VARIED SIGNIFICANTLY BY SITE (5%; 83%; 85%; P < .001). HIGH ADENOSINE TRIPHOSPHATE LEVELS WERE FOUND IN 22% OF ENDOSCOPES, AND MICROBIAL GROWTH WAS DETECTED IN 71% OF ENDOSCOPES. STENOTROPHOMONAS MALTOPHILIA, CITROBACTER FREUNDII, AND LECANICILLIUM LECANII/VERTICILLIUM DAHLIAE WERE FOUND. RETAINED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11776025 ↗InjuryMay 20, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP (B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. THIS REPORT IS FOR PATIENT 1.
Patient problems: Bacterial Infection
MAUDE report 10075460 ↗InjuryMay 20, 2020Microbial Contamination of Device
OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP (B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. A REVIEW OF THE SERVICE HISTORY INDICATES THE SCOPE WAS PURCHASED ON (B)(6) 1998 AND HAS RECEIVED TWO SERVICES IN THE LAST YEAR. THESE REPAIRS WERE NOT RELATED TO ANY POSITIVE PATIENT OR SCOPE C
Patient problems: Bacterial Infection
MAUDE report 10075383 ↗InjuryMay 19, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP(B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. THIS REPORT CAPTURES PATIENT 2.
Patient problems: Bacterial Infection
MAUDE report 10072982 ↗InjuryMay 19, 2020Improper or Incorrect Procedure or Method; Microbial Contamination of Device
OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP (B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. A REVIEW OF THE SERVICE HISTORY INDICATES THE SCOPE WAS PURCHASED ON (B)(6) 1998 AND HAS RECEIVED TWO SERVICES IN THE LAST YEAR. THESE REPAIRS WERE NOT RELATED TO ANY POSITIVE PATIENT OR SCOPE C
Patient problems: Bacterial Infection
MAUDE report 10072814 ↗InjuryMay 19, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS RECEIVED A COMPLAINT FROM A USER FACILITY STATING THAT THERE WAS CONCERN ABOUT TWO PATIENTS THAT DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PATIENT 2 HAD AN ERCP, THE PATIENT WAS NOT KNOWN TO HAVE CRE PRIOR TO THE ERCP. ON (B)(6) 2020 PATIENT 2 HAD AN ERCP WITH JF-140F, SN (B)(4). THIS PARTICULAR SCOPE WAS USED ON 3 ADDITIONAL PATIENTS WITHOUT ANY EVIDENCE OF TRANSMISSION. ON (B)(6) 2020, PATIENT 2 HAD A URINE CULTURE WHICH WAS POSITIVE FOR CRE. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. PATIENT 2 WAS AN INPATIENT ON THE SAME UNIT WHERE ANOTHER PATIENT TESTED POSITIVE FOR CRE AND SHARED CAREGIVERS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY R
Patient problems: Bacterial Infection
MAUDE report 10072318 ↗InjuryMay 19, 2020Microbial Contamination of Device
OLYMPUS RECEIVED A COMPLAINT FROM A USER FACILITY STATING THAT THERE WAS CONCERN ABOUT TWO PATIENTS THAT DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PATIENT 2 HAD AN ERCP, THE PATIENT WAS NOT KNOWN TO HAVE CRE PRIOR TO THE ERCP. ON 1/30/2020 PATIENT 2 HAD AN ERCP WITH JF-140F, SN (B)(4). THIS PARTICULAR SCOPE WAS USED ON 3 ADDITIONAL PATIENTS WITHOUT ANY EVIDENCE OF TRANSMISSION. ON (B)(6) 2020, PATIENT 2 HAD A URINE CULTURE WHICH WAS POSITIVE FOR CRE. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. PATIENT 2 WAS AN INPATIENT ON THE SAME UNIT WHERE ANOTHER PATIENT TESTED POSITIVE FOR CRE AND SHARED CAREGIVERS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REG
Patient problems: Bacterial Infection
MAUDE report 10071671 ↗InjuryMay 19, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem
OLYMPUS RECEIVED A COMPLAINT FROM A USER FACILITY STATING THAT THERE WAS CONCERN ABOUT TWO PATIENTS THAT DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PATIENT 1 HAD AN ERCP, THE PATIENTS WERE NOT KNOWN TO HAVE CRE PRIOR TO THE ERCPS. ON (B)(6) 2020 PATIENT 1 HAD AN ERCP WITH JF-140F, SN (B)(4). THIS PARTICULAR SCOPE WAS USED ON 3 ADDITIONAL PATIENTS WITHOUT ANY EVIDENCE OF TRANSMISSION. ON (B)(6) 2020, PATIENT 1 HAD BLOOD CULTURES DRAWN WHICH WERE POSITIVE FOR CRE. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. PATIENT 1 WAS AN INPATIENT ON THE SAME UNIT WHERE ANOTHER PATIENT TESTED POSITIVE FOR CRE AND SHARED CAREGIVERS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER F
Patient problems: Bacterial Infection
MAUDE report 10071542 ↗InjuryMay 19, 2020Microbial Contamination of Device
THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO CORRECT SECTION G6; 5-DAY REPORT SELECTED IN ERROR. G6: TYPE OF REPORT ¿ INITIAL (30-DAY).
Patient problems: Bacterial Infection
MAUDE report 10071424 ↗
Other Endoscopic video imaging system/component, gastroenterology-urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.