EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F (UDI 04953170339844)Recalls & reported problems

Olympus Medical Systems Corp. · JF-140F DUODENOVIDEOSCOPE (FIXED CAP) · Class 2

All models of EVIS DUODENOVIDEOSCOPE (3)

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 04953170339844), received between May 19, 2020 and May 5, 2021 — classified as 8 injuries, 1 malfunction.

The most frequently reported specific device problem is “Device Reprocessing Problem” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Device Reprocessing Problem5
  • Adverse Event Without Identified Device or Use Problem4
  • Microbial Contamination of Device4
  • Improper or Incorrect Procedure or Method1

Patient problems

  • Bacterial Infection8
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 5, 2021Device Reprocessing Problem; Device Reprocessing Problem

    IT IS REPORTED IN THE LITERATURE ARTICLE TITLED "RESIDUAL MOISTURE AND WATERBORNE PATHOGENS INSIDE FLEXIBLE ENDOSCOPES: EVIDENCE FROM A MULTISITE STUDY OF ENDOSCOPE DRYING EFFECTIVENESS," 22 DIFFERENT MODEL SCOPES TESTED POSITIVE FOR MICROORGANISMS AFTER REPROCESSING. BACKGROUND: ENDOSCOPY-ASSOCIATED INFECTION TRANSMISSION IS FREQUENTLY LINKED TO INADEQUATE REPROCESSING. RESIDUAL ORGANIC MATERIAL AND MOISTURE MAY FOSTER BIOFILM DEVELOPMENT INSIDE ENDOSCOPES. THIS STUDY EVALUATED THE EFFECTIVENESS OF ENDOSCOPE DRYING AND STORAGE METHODS AND ASSESSED ASSOCIATIONS BETWEEN RETAINED MOISTURE AND CONTAMINATION. METHODS: ENDOSCOPE REPROCESSING, DRYING, AND STORAGE PRACTICES WERE ASSESSED AT 3 HOSPITALS. RESEARCHERS PERFORMED VISUAL EXAMINATIONS AND TESTS TO DETECT FLUID AND CONTAMINATION ON PATIENT-READY ENDOSCOPES. RESULTS: FLUID WAS DETECTED IN 22 OF 45 (49%) ENDOSCOPES. PREVALENCE OF MOISTURE VARIED SIGNIFICANTLY BY SITE (5%; 83%; 85%; P < .001). HIGH ADENOSINE TRIPHOSPHATE LEVELS WERE FOUND IN 22% OF ENDOSCOPES, AND MICROBIAL GROWTH WAS DETECTED IN 71% OF ENDOSCOPES. STENOTROPHOMONAS MALTOPHILIA, CITROBACTER FREUNDII, AND LECANICILLIUM LECANII/VERTICILLIUM DAHLIAE WERE FOUND. RETAINED

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11776025 ↗
  • InjuryMay 20, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem

    OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP (B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. THIS REPORT IS FOR PATIENT 1.

    Patient problems: Bacterial Infection

    MAUDE report 10075460 ↗
  • InjuryMay 20, 2020Microbial Contamination of Device

    OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP (B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. A REVIEW OF THE SERVICE HISTORY INDICATES THE SCOPE WAS PURCHASED ON (B)(6) 1998 AND HAS RECEIVED TWO SERVICES IN THE LAST YEAR. THESE REPAIRS WERE NOT RELATED TO ANY POSITIVE PATIENT OR SCOPE C

    Patient problems: Bacterial Infection

    MAUDE report 10075383 ↗
  • InjuryMay 19, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem

    OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP(B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. THIS REPORT CAPTURES PATIENT 2.

    Patient problems: Bacterial Infection

    MAUDE report 10072982 ↗
  • InjuryMay 19, 2020Improper or Incorrect Procedure or Method; Microbial Contamination of Device

    OLYMPUS RECEIVED A REPORT FROM A USER FACILITY, STATING TWO PATIENTS HAD DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) PROCEDURES 2 DAYS APART THAT USED THE SAME SCOPE. PATIENT 1 HAD AN ERCP (B)(6) 2020 AND PATIENT 2, HAD AN ERCP ON (B)(6) 2020. THE SCOPE WAS NOT USED ON ANY OTHER PATIENT IN BETWEEN THESE TWO PATIENTS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REGARDING THE TREATMENT OF THE INFECTIONS IN EITHER PATIENT. A REVIEW OF THE SERVICE HISTORY INDICATES THE SCOPE WAS PURCHASED ON (B)(6) 1998 AND HAS RECEIVED TWO SERVICES IN THE LAST YEAR. THESE REPAIRS WERE NOT RELATED TO ANY POSITIVE PATIENT OR SCOPE C

    Patient problems: Bacterial Infection

    MAUDE report 10072814 ↗
  • InjuryMay 19, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem

    OLYMPUS RECEIVED A COMPLAINT FROM A USER FACILITY STATING THAT THERE WAS CONCERN ABOUT TWO PATIENTS THAT DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PATIENT 2 HAD AN ERCP, THE PATIENT WAS NOT KNOWN TO HAVE CRE PRIOR TO THE ERCP. ON (B)(6) 2020 PATIENT 2 HAD AN ERCP WITH JF-140F, SN (B)(4). THIS PARTICULAR SCOPE WAS USED ON 3 ADDITIONAL PATIENTS WITHOUT ANY EVIDENCE OF TRANSMISSION. ON (B)(6) 2020, PATIENT 2 HAD A URINE CULTURE WHICH WAS POSITIVE FOR CRE. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. PATIENT 2 WAS AN INPATIENT ON THE SAME UNIT WHERE ANOTHER PATIENT TESTED POSITIVE FOR CRE AND SHARED CAREGIVERS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY R

    Patient problems: Bacterial Infection

    MAUDE report 10072318 ↗
  • InjuryMay 19, 2020Microbial Contamination of Device

    OLYMPUS RECEIVED A COMPLAINT FROM A USER FACILITY STATING THAT THERE WAS CONCERN ABOUT TWO PATIENTS THAT DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PATIENT 2 HAD AN ERCP, THE PATIENT WAS NOT KNOWN TO HAVE CRE PRIOR TO THE ERCP. ON 1/30/2020 PATIENT 2 HAD AN ERCP WITH JF-140F, SN (B)(4). THIS PARTICULAR SCOPE WAS USED ON 3 ADDITIONAL PATIENTS WITHOUT ANY EVIDENCE OF TRANSMISSION. ON (B)(6) 2020, PATIENT 2 HAD A URINE CULTURE WHICH WAS POSITIVE FOR CRE. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. PATIENT 2 WAS AN INPATIENT ON THE SAME UNIT WHERE ANOTHER PATIENT TESTED POSITIVE FOR CRE AND SHARED CAREGIVERS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER FACILITY REG

    Patient problems: Bacterial Infection

    MAUDE report 10071671 ↗
  • InjuryMay 19, 2020Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem; Device Reprocessing Problem; Adverse Event Without Identified Device or Use Problem

    OLYMPUS RECEIVED A COMPLAINT FROM A USER FACILITY STATING THAT THERE WAS CONCERN ABOUT TWO PATIENTS THAT DEVELOPED CARBAPENEM-RESISTANT ENTEROBACTERIACEAE (CRE) FOLLOWING ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP). PATIENT 1 HAD AN ERCP, THE PATIENTS WERE NOT KNOWN TO HAVE CRE PRIOR TO THE ERCPS. ON (B)(6) 2020 PATIENT 1 HAD AN ERCP WITH JF-140F, SN (B)(4). THIS PARTICULAR SCOPE WAS USED ON 3 ADDITIONAL PATIENTS WITHOUT ANY EVIDENCE OF TRANSMISSION. ON (B)(6) 2020, PATIENT 1 HAD BLOOD CULTURES DRAWN WHICH WERE POSITIVE FOR CRE. THE SCOPE WAS DOUBLE MANUALLY CLEANED AND DOUBLE HIGH-LEVEL DISINFECTED BETWEEN PATIENTS AS IS THEIR USUAL PROCESS. PATIENT 1 WAS AN INPATIENT ON THE SAME UNIT WHERE ANOTHER PATIENT TESTED POSITIVE FOR CRE AND SHARED CAREGIVERS. BOTH ISOLATES WERE SENT FOR GENOMIC ANALYSIS AND HAVE MATCHING CHROMOSOMAL PATTERNS. THE USER FACILITY SUSPECTS THAT TRANSMISSION OCCURRED EITHER VIA DUODENOSCOPE OR THROUGH SHARED CAREGIVERS. ALL THE 140 SERIES SCOPES WERE SENT THROUGH ETO STERILIZATION, PER FACILITY PROTOCOL, AND WERE NOT CULTURED PRIOR. THE JF-140F SCOPE HAS BEEN QUARANTINED AT THE USER FACILITY. NO ADDITIONAL INFORMATION COULD BE PROVIDED BY THE USER F

    Patient problems: Bacterial Infection

    MAUDE report 10071542 ↗
  • InjuryMay 19, 2020Microbial Contamination of Device

    THIS SUPPLEMENTAL REPORT IS BEING SUBMITTED TO CORRECT SECTION G6; 5-DAY REPORT SELECTED IN ERROR. G6: TYPE OF REPORT ¿ INITIAL (30-DAY).

    Patient problems: Bacterial Infection

    MAUDE report 10071424 ↗

FDA clearance and classification

The FDA regulates product code FET (endoscopic video imaging system/component, gastroenterology-urology) under 21 CFR 876.1500 as a Class II device. 21 CFR 876.1500 ↗

Devices of this type normally need 510(k) clearance before sale.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F

Has the EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F been recalled?

No FDA recall record in our data names this exact device identifier (DI 04953170339844), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F?

The FDA's MAUDE database lists 9 adverse event reports for this device, received between May 19, 2020 and May 5, 2021. In those reports, the most common specific device problems are “Device Reprocessing Problem” (5), “Microbial Contamination of Device” (4) and “Improper or Incorrect Procedure or Method” (1); the most common patient problem is “Bacterial Infection” (8). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F?

None of the reports are classified as deaths, 8 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Who makes the EVIS DUODENOVIDEOSCOPE OLYMPUS JF TYPE 140F, and is it still sold?

The FDA's device database lists Olympus Medical Systems Corp. as the company responsible for it (brand EVIS DUODENOVIDEOSCOPE, model OLYMPUS JF TYPE 140F). Its record shows it as no longer in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Endoscopic video imaging system/component, gastroenterology-urology devices

All Endoscopic video imaging system/component, gastroenterology-urology devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page