THUNDERBEAT Open Fine Jaw Type X TB-0009OFX (UDI 04953170394577)Recalls & reported problems
Olympus Medical Systems Corp. · THUNDERBEAT Open Fine Jaw Type X · Class 2
All models of THUNDERBEAT Open Fine Jaw Type X (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 04953170394577), received between Oct 19, 2022 and Mar 12, 2026 — classified as 13 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component3
- Peeled/Delaminated3
- Break2
- Fracture2
- Degraded1
- Material Separation1
- Naturally Worn1
Patient problems
- No Clinical Signs, Symptoms or Conditions13
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 12, 2026Degraded
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE ULTRASONIC SURGICAL INSTRUMENT TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24581695 ↗MalfunctionNov 20, 2025Naturally Worn; Naturally Worn
DURING THE DEVICE EVALUATION, IT WAS FOUND THAT THE TISSUE PAD IN THE GRASPING SECTION OF THE THUNDERBEAT DEVICE WAS WORN. NO PATIENTS WERE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23611987 ↗MalfunctionAug 22, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THE THUNDERBEAT JAW CLAMP BECAME DETACHED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22880076 ↗MalfunctionJul 8, 2025Break; Break
IT WAS REPORTED THE TIP OF THE THUNDERBEAT BROKE OFF DURING A PROCEDURE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22436036 ↗MalfunctionJan 30, 2025Break; Break
IT WAS REPORTED THAT UPON USING THE THUNDERBEAT AROUND THE UTERUS DURING A PROCEDURE THE TIP BROKE AND DETACHED. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21276208 ↗MalfunctionJun 20, 2024Material Separation; Material Separation
IT WAS OBSERVED THAT DURING THE DEVICE INSPECTION, THE ULTRASONIC SURGICAL DEVICE EXHIBITED A PART OF THE TISSUE PAD WAS PEELED AWAY. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19576149 ↗MalfunctionMay 6, 2024Peeled/Delaminated; Peeled/Delaminated
THE ISSUE OCCURRED 10 MINUTES FROM THE END OF A THYROIDECTOMY PROCEDURE CAUSING A 15 MINUTE PROCEDURAL DELAY. A NON-DISPOSABLE BIPOLAR FORCEPS WAS USED TO COMPLETE THE PROCEDURE WITHOUT THE ENERGY DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19249956 ↗MalfunctionSep 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THE GRIP OF THE THUNDERBEAT OPEN FINE JAW TYPE X BROKE AT THE BASE. THE ISSUE OCCURRED DURING A THERAPEUTIC HEPATECTOMY PROCEDURE. THE GRIP PART FELL OFF WHEN THE NURSE, WHO TOOK OUT THE INSTRUMENT, WAS CLEANING THE TIP TISSUE PAD WITH GAUZE. THE NURSE WAS CLEANING THE BURN PART. THE GRIP BROKE IMMEDIATELY WITHIN AN HOUR AFTER THE START OF THE OPERATION. A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17791975 ↗MalfunctionSep 21, 2023Detachment of Device or Device Component; Detachment of Device or Device Component
THE CUSTOMER REPORTED THE WIPER JAW OF THE THUNDERBEAT OPEN FINE JAW TYPE X BROKE WHEN THE NURSE WAS WIPING OFF DIRT. THE ISSUE OCCURRED ABOUT EIGHT HOURS AFTER STARTING THE THERAPEUTIC PANCREATECTOMY PROCEDURE. THE CUSTOMER CONFIRMED THE METAL PART OF THE WIPER JAW WAS THE BROKEN INSERTION PART. A SIMILAR DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT HARM OR IMPACT DUE TO THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17791899 ↗MalfunctionAug 9, 2023Fracture; Fracture
AN OLYMPUS TERRITORY MANAGER REPORTED ON BEHALF OF A CUSTOMER, DURING AN UNSPECIFIED PROCEDURE A THYROID THUNDERBEAT DID NOT WORK AS EXPECTED. THE TIP BROKE OFF AND WAS BURNT. THE TIP DID NOT FALL WITHIN THE PATIENT¿S CAVITY. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17494554 ↗MalfunctionJul 6, 2023Peeled/Delaminated; Peeled/Delaminated
THE CUSTOMER REPORTED THAT DURING AN UNSPECIFIED PROCEDURE, THE INSIDE OF THE ISOLATED CLAW CAME LOOSE A BIT WHILE USING THE FIRST DEVICE. THE ULTRASONIC GENERATOR GAVE AN ERROR MESSAGE WHILE USING ANOTHER DEVICE AND THE DEVICE MUST BE REPLACED. THERE WAS NO PATIENT OR USER INJURY REPORTED DUE TO THE EVENT. NO ADDITIONAL INFORMATION WAS PROVIDED REGARDING THE REPORTED ISSUE. THE SUBJECT DEVICE WAS RETURNED TO AN OLYMPUS SERVICE CENTER FOR EVALUATION. UPON INSPECTION AND TESTING OF THE RETURNED DEVICE, IT WAS OBSERVED THAT THE TISSUE PAD IN THE GRASPING SECTION WAS WORN AWAY, AND A PART OF THE TISSUE PAD WAS PEELED AWAY. THIS REPORT IS BEING SUBMITTED FOR THE MALFUNCTION FOUND DURING EVALUATION (WORN AND PEELED TISSUE PAD).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17274062 ↗MalfunctionApr 24, 2023Fracture; Fracture
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, DURING AN EVENTRATION PROCEDURE AFTER FIVE MINUTES F USING THE THUNDERBEAT DEVICE, THE DEVICE TIP BROKE OFF. THE PROCEDURE WAS COMPLETED WITH A NEW SIMILAR DEVICE. THERE IS NO CONSEQUENCE, HARM OR ADVERSE IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16795583 ↗MalfunctionOct 19, 2022Peeled/Delaminated; Peeled/Delaminated
AS REPORTED FOR THIS EVENT BY THE CUSTOMER, APPROXIMATELY TWO HOURS INTO A THERAPEUTIC OPEN LIVER RESECTION PROCEDURE, AN UNKNOWN OUTPUT ERROR WAS RECEIVED FOR THE DEVICE WHEN SHORT-PITCH OUTPUT WAS CONTINUED FOR ABOUT 10 MINUTES. THE DOCTOR CONTINUED TO USE THE DEVICE BUT THE ERROR MESSAGES CONTINUED. THE DEVICE WAS REMOVED FROM THE PATIENT BODY FOR CHECKING. THE DEVICE TISSUE PAD WAS PEELED BACK. THE PROCEDURE WAS COMPLETED WITH A NEW SIMILAR DEVICE. THERE IS NO HARM OR ADVERSE IMPACT TO THE PATIENT. THE SETTINGS AT THE TIME WERE SEAL AND CUT1, SEAL 3. THE INTELLIGENT TISSUE MONITORING (ITM) SETTING WERE ON DURING THE PROCEDURE. THE DEVICE WAS INSPECTED PRIOR TO USE WITH NO ANOMALY NTOED. PER THE DOCTOR, THE CASE WAS COMPLEX COMPARED TO NORMAL SUCH CASES. THE DOCTOR IS A REGULAR USER OF THE DEVICE OFJ TYPE OF THUNDERBEAT. THIS DEVICE IS OFX TYPE WHICH IS NEW AND IMPROVED. THE DEVICE WAS BEING USED TO DEMONSTRATE THE QUALITIES OF THE DEVICE. A FALSE OSCILLATION WAS PERFORMED WHICH CONTRIBUTED TO THE EVENT. THE DOCTOR NOW UNDERSTANDS THE DEVICE USAGE AND WHAT LED TO THE ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15633141 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page