ULTRASONIC BIPOLAR GENERATOR USG-410Recalls & reported problems

Olympus Medical Systems Corp. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 04953170406881), received between May 27, 2022 and Nov 18, 2024 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 2 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Break1
  • Detachment of Device or Device Component1
  • Electrical /Electronic Property Problem1
  • Output Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Burn(s)1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 18, 2024Break; Break

    IT WAS REPORTED THAT DURING AN INTRODUCTION OF THE INSTRUMENT, THE ULTRASONIC BIPOLAR GENERATOR PROBE BROKE OFF BEFORE ITS ACTUAL USE. THE INTENDED THERAPEUTIC PROCEDURE WAS COMPLETED WITH A SIMILAR DEVICE WITH NO DELAY. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20706481 ↗
  • MalfunctionNov 11, 2024Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE GENERATOR HAD VARIOUS ERROR CODES DURING USE (U504, U508, U510, U511 & U601). THE DEVICE WAS USED IN COMBINATION WITH A THUNDERBEAT WHERE THE ACTIVE BRANCH BROKE AND FELL OFF INTO THE PATIENT'S ABDOMEN. IT WAS ALSO REPORTED THAT THE TRANSDUCER CABLES WERE NEW. ADDITIONAL INFORMATION HAS BEEN REQUESTED BUT NOTHING RECEIVED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20658134 ↗
  • InjuryAug 22, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT AFTER USING THE ULTRASONIC GENERATOR FOR A THERAPEUTIC GYNECOLOGICAL LAPAROSCOPY THE PATIENT HAD A BURN SCAR ON HER NECK. IT WAS BELIEVED THAT SOMETHING METALLIC HAD COME INTO CONTACT WITH THE NECK. THE PATIENT WAS CURRENTLY BEING TREATED AT THE DERMATOLOGY DEPARTMENT OF THE FACILITY. THE PATIENT'S HOSPITALIZATION WAS NOT PROLONGED DUE TO THE INCIDENT. THERE WAS NO FURTHER IMPACT REPORTED.

    Patient problems: Burn(s)

    MAUDE report 20035056 ↗
  • MalfunctionSep 28, 2022Output Problem; Output Problem

    IT WAS REPORTED THE (LOANER) DEVICE (USG-410) EXHIBITED AN ERROR MESSAGE OF "PLEASE RESTART THE GENERATOR" EACH TIME THE SURGEON ACTIVATED SEAL AND CUT. ACCORDING TO THE REPORTER, SEAL WAS ABSOLUTELY WORKING FINE. THE OLYMPUS REPRESENTATIVE WHO WAS PRESENT DURING THE THERAPEUTIC TOTAL LAPAROSCOPIC HYSTERECTOMY (TLH) PROCEDURE TURNED THE DEVICE OFF SEVERAL TIMES HOWEVER, THE SURGEON WAS REPORTED LOST HIS PATIENCE WITH THE DISRUPTION AND CHANGED TRANSDUCER, CHANGED DEVICE TO A THIRD PARTY ETHICON (HARMONIC DEVICE). THE INTENDED PROCEDURE WAS COMPLETED USING THE THIRD PARTY DEVICE . THERE WAS A TEN (10) MINUTES DELAY IN THE PROCEDURE DUE TO THE CUSTOMER/OLYMPUS REPRESENTATIVE TRIED TO GET THE DEVICE (USG-410) TO WORK, HOWEVER, THE DELAY WAS REPORTED NOT CLINICALLY RELEVANT, PATIENT ACCORDING TO THE REPORTER IS ABSOLUTELY FINE. THERE WAS NO PATIENT HARM, NO USER INJURY REPORTED DUE TO THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15499252 ↗
  • MalfunctionMay 27, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    THE CUSTOMER REPORTED TO OLYMPUS, APPROXIMATELY TWO (2) HOURS INTO A THERAPEUTIC HYSTERECTOMY PROCEDURE, THE ULTRASONIC BIPOLAR GENERATOR (USG-410) DISPLAYED TWO (2) PROBE DAMAGE ERROR CODES, U504 AND U601, AND STARTED BEEPING. THE THUNDERBEAT DEVICE WAS REMOVED FROM THE PATIENT AND THE PROBE JAWS WERE CLEANED WITH WET GAUZE, THE ELECTRODE WAS TESTED, THE CABLE WAS CHECKED AND IT WAS CONFIRMED THERE WAS NO CONTACT WITH METAL. AFTER PERFORMING TROUBLESHOOTING, THE CUSTOMER REPORTED THE SYSTEM PASSED TESTING. THE PHYSICIAN PROCEEDED TO INSERT THE THUNDERBEAT PROBE BACK INTO THE PATIENT¿S ABDOMEN. AT THE FIRST ACTIVATION OF THE THUNDERBEAT, THE JAWS BROKE INSIDE THE PATIENT. THE SURGEON IMMEDIATELY REMOVED THE THUNDERBEAT DEVICE AND THE BROKEN PIECE FROM THE ABDOMEN. THERE WAS NO PATIENT OR USER HARM REPORTED DUE TO THE EVENT. THE INTENDED PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE SAME USG-410, BUT WITH A DIFFERENT THUNDERBEAT. IT IS UNKNOWN HOW THE BROKEN PIECE WAS RETRIEVED AND IF THERE WAS A PROCEDURAL DELAY. THIS EVENT INCLUDES TWO REPORTS: THIS REPORT IS FOR THE ULTRASONIC BIPOLAR GENERATOR, USG-410, S/N (B)(4), PATIENT IDENTIFIER (B)(6). THE THUNDERBEAT, TB-0535FCS, LOT KR14

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14524147 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →