4K Camera Head OLYMPUS CH-S700-XZ-EARecalls & reported problems
Olympus Medical Systems Corp. · Class 2
All models of 4K CAMERA HEAD (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Olympus camera head is used with endoscopic video imaging systems. The 26 reports mostly describe communication failures, leaks, and missing or distorted images; there were no deaths or injuries. No recalls name this device, and reports rose sharply to 25 in the last 12 months from 1 in the previous 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 29
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 04953170430268), received between Oct 30, 2025 and Sep 28, 2026 — classified as 29 malfunctions.
The most frequently reported specific device problem is “Communication or Transmission Problem” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Communication or Transmission Problem11
- No Display/Image7
- Gas/Air Leak5
- Leak/Splash4
- Output Problem3
- Degraded2
- Material Split, Cut or Torn2
- Failure to Clean Adequately1
Patient problems
- No Clinical Signs, Symptoms or Conditions29
Recent adverse event reports
The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 28, 2026Poor Quality Image; Gas/Air Leak
The customer reported poor image quality during inspection for use involving an Olympus autoclavable camera head. During the initial evaluation of the returned device, a failed leak test was observed. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50917289 ↗MalfunctionSep 24, 2026No Display/Image; Leak/Splash
The customer reported that the Olympus camera head had no image during inspection for use. During the initial evaluation of the device, it was observed that the device failed the leak test. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50884392 ↗MalfunctionSep 22, 2026No Display/Image
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50861669 ↗MalfunctionSep 15, 2026No Display/Image; Communication or Transmission Problem
It was reported that the autoclavable camera head displayed E226 (Incompatible Scope) and a white-out image. The issue was identified during inspection for use. There were no reports of patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50776665 ↗MalfunctionSep 10, 2026No Display/Image; No Device Output
The customer reported no light and no image involving an Olympus autoclavable camera head. There were no reports of patient harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50706847 ↗MalfunctionAug 26, 2026Image Display Error/Artifact; Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50455635 ↗MalfunctionAug 19, 2026Communication or Transmission Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50332439 ↗MalfunctionAug 17, 2026Gas/Air Leak
The customer returned the autoclavable camera head, which is an Olympus asset with no reported complaint. During the evaluation of the device, failed water leak test on cable was observed. The Service Repair Technician indicated that the most likely cause was failed water leak test due to cut on cable. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50274575 ↗MalfunctionAug 17, 2026Degraded; Communication or Transmission Problem; Gas/Air Leak; Optical Distortion; Material Split, Cut or Torn
The customer reported the camera head had Distorted Image. During the initial evaluation of the device, it was observed that Smashed connector, failed leak test, image quality error (error E238) shown on monitor. No patient harm was reported.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50273469 ↗MalfunctionJul 28, 2026
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50055640 ↗MalfunctionJul 16, 2026Gas/Air Leak
THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE TECHNICAL ASSISTANCE INDICATES THAT A FAILED LEAK TEST WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25842830 ↗MalfunctionJul 16, 2026Degraded
THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TEAM INDICATES THAT DAMAGE CABLE INSULATION WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25841863 ↗MalfunctionJun 10, 2026Output Problem
THE CUSTOMER REPORTED IR MODE IS NOT WORKING DURING AN UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS AUTOCLAVABLE CAMERA HEAD. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25462120 ↗MalfunctionJun 6, 2026Communication or Transmission Problem
THE CUSTOMER REPORTED THAT COMMUNICATION ERROR (E238) OCCURRED EACH TIME THE OLYMPUS AUTOCLAVABLE CAMERA HEAD WAS PLUGGED IN. THERE WAS NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25420647 ↗MalfunctionJun 2, 2026Leak/Splash
THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING EVALUATION OF THE DEVICE LEAKAGE WAS FOUND. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE FAILURE IS DUE TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25364676 ↗MalfunctionMay 29, 2026Gas/Air Leak
THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TEAM INDICATES THAT FAILED LEAK TEST WAS FOUND. IT WAS DETERMINED THAT THE DAMAGED INSULATION CORD NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25331449 ↗MalfunctionMay 19, 2026No Display/Image
IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25215023 ↗MalfunctionMay 6, 2026Leak/Splash
IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE AUTOCLAVEABLE CAMERA HEAD FAILED THE LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25095179 ↗MalfunctionApr 25, 2026Communication or Transmission Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE AUTOCLAVE CAMERA HEAD HAD SCOPE COMMUNICATION ERROR (E226) ON THE SCREEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24992045 ↗MalfunctionApr 9, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD DISPLAYED A COMMUNICATION ERROR MESSAGE MENTIONING THE CAMERA. THE ISSUE WAS FOUND DURING A THERAPEUTIC LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, WHICH WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24829276 ↗MalfunctionMar 31, 2026Product Quality Problem
IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD DID NOT WORK. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24742292 ↗MalfunctionMar 30, 2026Communication or Transmission Problem
IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD HAD E238 SCOPE COMMUNICATION ERROR. THERE WERE NO REPORTS OF PATIENT HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24715133 ↗MalfunctionMar 10, 2026Communication or Transmission Problem
NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24561246 ↗MalfunctionJan 7, 2026No Display/Image; Output Problem
IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED NO IMAGE, ERROR 226 AND FAULTY VIDEO CABLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23991829 ↗MalfunctionJan 6, 2026No Display/Image; Communication or Transmission Problem
IT WAS REPORTED THE CAMERA HEAD HAD A SCOPE COMMUNICATION ERROR ERROR E238, NO IMAGE, AND DID NOT WORK AT OUT OF THE BOX FAILURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23976156 ↗
Other Endoscopic Video Imaging System/Component, Gastroenterology-Urology devices
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EBOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,522 reports
- 4K CAMERA HEAD OLYMPUS CH-S400-XZ-EAOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology2,375 reports
- HD CAMERA HEAD OTV-S7ProH-HD-L08EOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1,424 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology529 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology4 deaths413 reports
- DUODENOVIDEOSCOPE TJF-150Olympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology1 recall189 reports
- VISERA ELITE II VIDEO SYSTEM CENTER OLYMPUS OTV-S300Olympus Medical Systems Corp. · Endoscopic Video Imaging System/Component, Gastroenterology-Urology158 reports
- VIDEO SYSTEM OTV-SIOlympus Medical Systems Corp. · Endoscopic video imaging system/component, gastroenterology-urology143 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page