Quick summary

This Olympus camera head is used with endoscopic video imaging systems. The 26 reports mostly describe communication failures, leaks, and missing or distorted images; there were no deaths or injuries. No recalls name this device, and reports rose sharply to 25 in the last 12 months from 1 in the previous 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
29
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 29 adverse event reports that list this exact device (DI 04953170430268), received between Oct 30, 2025 and Sep 28, 2026 — classified as 29 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 2 reportsDec 2025: 0 reportsJan 2026: 3 reportsFeb 2026: 0 reportsMar 2026: 3 reportsApr 2026: 2 reportsMay 2026: 3 reportsJun 2026: 3 reportsJul 2026: 3 reportsAug 2026: 4 reportsSep 2026: 5 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Communication or Transmission Problem11
  • No Display/Image7
  • Gas/Air Leak5
  • Leak/Splash4
  • Output Problem3
  • Degraded2
  • Material Split, Cut or Torn2
  • Failure to Clean Adequately1

Patient problems

  • No Clinical Signs, Symptoms or Conditions29

Recent adverse event reports

The 25 most recent of 29 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 28, 2026Poor Quality Image; Gas/Air Leak

    The customer reported poor image quality during inspection for use involving an Olympus autoclavable camera head. During the initial evaluation of the returned device, a failed leak test was observed. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917289 ↗
  • MalfunctionSep 24, 2026No Display/Image; Leak/Splash

    The customer reported that the Olympus camera head had no image during inspection for use. During the initial evaluation of the device, it was observed that the device failed the leak test. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884392 ↗
  • MalfunctionSep 22, 2026No Display/Image

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50861669 ↗
  • MalfunctionSep 15, 2026No Display/Image; Communication or Transmission Problem

    It was reported that the autoclavable camera head displayed E226 (Incompatible Scope) and a white-out image. The issue was identified during inspection for use. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50776665 ↗
  • MalfunctionSep 10, 2026No Display/Image; No Device Output

    The customer reported no light and no image involving an Olympus autoclavable camera head. There were no reports of patient harm.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50706847 ↗
  • MalfunctionAug 26, 2026Image Display Error/Artifact; Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50455635 ↗
  • MalfunctionAug 19, 2026Communication or Transmission Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50332439 ↗
  • MalfunctionAug 17, 2026Gas/Air Leak

    The customer returned the autoclavable camera head, which is an Olympus asset with no reported complaint. During the evaluation of the device, failed water leak test on cable was observed. The Service Repair Technician indicated that the most likely cause was failed water leak test due to cut on cable. There was no patient involvement.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50274575 ↗
  • MalfunctionAug 17, 2026Degraded; Communication or Transmission Problem; Gas/Air Leak; Optical Distortion; Material Split, Cut or Torn

    The customer reported the camera head had Distorted Image. During the initial evaluation of the device, it was observed that Smashed connector, failed leak test, image quality error (error E238) shown on monitor. No patient harm was reported.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50273469 ↗
  • MalfunctionJul 28, 2026

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50055640 ↗
  • MalfunctionJul 16, 2026Gas/Air Leak

    THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE TECHNICAL ASSISTANCE INDICATES THAT A FAILED LEAK TEST WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25842830 ↗
  • MalfunctionJul 16, 2026Degraded

    THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TEAM INDICATES THAT DAMAGE CABLE INSULATION WAS FOUND. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25841863 ↗
  • MalfunctionJun 10, 2026Output Problem

    THE CUSTOMER REPORTED IR MODE IS NOT WORKING DURING AN UNSPECIFIED PROCEDURE INVOLVING AN OLYMPUS AUTOCLAVABLE CAMERA HEAD. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25462120 ↗
  • MalfunctionJun 6, 2026Communication or Transmission Problem

    THE CUSTOMER REPORTED THAT COMMUNICATION ERROR (E238) OCCURRED EACH TIME THE OLYMPUS AUTOCLAVABLE CAMERA HEAD WAS PLUGGED IN. THERE WAS NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25420647 ↗
  • MalfunctionJun 2, 2026Leak/Splash

    THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD WHICH IS AN OLYMPUS ASSET WITH NO REPORTED COMPLAINT. DURING EVALUATION OF THE DEVICE LEAKAGE WAS FOUND. THE SERVICE REPAIR TECHNICIAN INDICATED THAT THE MOST LIKELY CAUSE OF THE FAILURE IS DUE TO COMPONENT FAILURE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25364676 ↗
  • MalfunctionMay 29, 2026Gas/Air Leak

    THE CUSTOMER RETURNED THE AUTOCLAVABLE CAMERA HEAD TO THE SERVICE CENTER FOR A SEPARATE ISSUE. DURING THE DEVICE ANALYSIS, THE SERVICE TEAM INDICATES THAT FAILED LEAK TEST WAS FOUND. IT WAS DETERMINED THAT THE DAMAGED INSULATION CORD NEEDED TO BE REPLACED. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25331449 ↗
  • MalfunctionMay 19, 2026No Display/Image

    IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD HAD NO IMAGE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25215023 ↗
  • MalfunctionMay 6, 2026Leak/Splash

    IT WAS OBSERVED THAT DURING DEVICE EVALUATION, THE AUTOCLAVEABLE CAMERA HEAD FAILED THE LEAK TEST. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25095179 ↗
  • MalfunctionApr 25, 2026Communication or Transmission Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE AUTOCLAVE CAMERA HEAD HAD SCOPE COMMUNICATION ERROR (E226) ON THE SCREEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24992045 ↗
  • MalfunctionApr 9, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD DISPLAYED A COMMUNICATION ERROR MESSAGE MENTIONING THE CAMERA. THE ISSUE WAS FOUND DURING A THERAPEUTIC LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, WHICH WAS COMPLETED WITH A SIMILAR DEVICE. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24829276 ↗
  • MalfunctionMar 31, 2026Product Quality Problem

    IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD DID NOT WORK. THE ISSUE WAS FOUND DURING REPROCESSING. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24742292 ↗
  • MalfunctionMar 30, 2026Communication or Transmission Problem

    IT WAS REPORTED THAT THE AUTOCLAVABLE CAMERA HEAD HAD E238 SCOPE COMMUNICATION ERROR. THERE WERE NO REPORTS OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24715133 ↗
  • MalfunctionMar 10, 2026Communication or Transmission Problem

    NO ADDITIONAL INFORMATION RECEIVED FROM CUSTOMER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24561246 ↗
  • MalfunctionJan 7, 2026No Display/Image; Output Problem

    IT WAS OBSERVED THAT DURING THE DEVICE EVALUATION, THE CAMERA HEAD EXHIBITED NO IMAGE, ERROR 226 AND FAULTY VIDEO CABLE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23991829 ↗
  • MalfunctionJan 6, 2026No Display/Image; Communication or Transmission Problem

    IT WAS REPORTED THE CAMERA HEAD HAD A SCOPE COMMUNICATION ERROR ERROR E238, NO IMAGE, AND DID NOT WORK AT OUT OF THE BOX FAILURE. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23976156 ↗

Other Endoscopic Video Imaging System/Component, Gastroenterology-Urology devices