Skip to content

Logi - Laparoscopic Instrument System 120-7500Recalls & reported problems

Surgical Innovations Limited · LogiCut Mini Scissor Length 320mm · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 05051986000411), received between Jan 4, 2019 and Aug 25, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1
  • Material Integrity Problem1
  • Smoking1

Patient problems

  • Burn, Thermal1
  • Insufficient Information1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • OtherAug 25, 2026Material Integrity Problem; Insufficient Device Problem Information

    Patient problems: Insufficient Information

    MAUDE report 50434469 ↗
  • MalfunctionJan 4, 2019Smoking

    THE SCISSOR WAS BEING USED TO DIATHERMY A SECTION OF ENDOMETRIOSIS AND THE CONSULTANT NOTICED SMOKE COMING FROM THE SHAFT OF THE INSTRUMENT AWAY FROM THE SITE OF DIATHERMY WHERE THE UTERUS WAS TOUCHING THE SHAFT. THEREFORE, THE CONSULTANT WAS AWARE THE INSTRUMENT MUST HAVE SHORTED OUT OF VIEW AND CAUSED A BURN TO THE UTERUS WHERE THE SHAFT WAS IN CONTACT WITH IT. ELECTRICAL SHORT CIRCUIT THROUGH THE FAULTY INSULATION CAUSED A SUPERFICIAL BURN TO THE UTERINE SEROSA. THIS WILL HAVE NO CONSEQUENCE TO THE PATIENT BUT IF THE BURN HAD BEEN ON THE BOWEL, BLADDER URETER OR FALLOPIAN TUBES, SERIOUS HARM COULD HAVE RESULTED.

    Patient problems: Burn, Thermal

    MAUDE report 8216986 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →

Sources and updates