Skip to content

NeuroTherm™ RF-DGP-SRecalls & reported problems

Neurotherm, Inc. · Grounding Pad · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 05415067023629), received between Oct 8, 2019 and Dec 9, 2024 — classified as 4 injuries.

The most frequently reported specific device problem is “Grounding Malfunction” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 2 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Grounding Malfunction1
  • Improper or Incorrect Procedure or Method1
  • Overheating of Device1
  • Temperature Problem1

Patient problems

  • Contact Dermatitis2
  • Partial thickness (Second Degree) Burn2

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryDec 9, 2024Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT DURING A RADIOFREQUENCY LESIONING PROCEDURE PATIENT EXPERIENCED BURNING SENSATION AT GROUNDING PAD SITE. IN TURN, THE PROCEDURE WAS STOPPED AND GROUNDING PAD WAS REMOVED. RED IRRITATION/ABRASION AT SITE OF GROUNDING PAD WAS NOTED.

    Patient problems: Contact Dermatitis

    MAUDE report 20883829 ↗
  • InjuryDec 9, 2024Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem; Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT HAD MULTIPLE YELLOW COLOR LIKE BURNS ON THEIR SKIN FOLLOWING AN RFA PROCEDURE. THE PATIENT DOES NOT FEEL PAIN AND ANY BURNING SENSATIONS. NO FURTHER INFORMATION IS AVAILABLE AT THIS TIME.

    Patient problems: Contact Dermatitis

    MAUDE report 20878855 ↗
  • InjuryNov 4, 2022Grounding Malfunction; Overheating of Device; Grounding Malfunction; Overheating of Device

    DURING THE PROCEDURE, THE PATIENT EXPERIENCED A GROUNDING PAD BURN WITH REDDENED SKIN AND BLISTERS. IT WAS NOTED THAT THE PATIENT WAS HAIRY, AND THE AREA WAS NOT SHAVED PRIOR TO PLACEMENT OF THE GROUNDING PAD.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 15735451 ↗
  • InjuryOct 8, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    DURING A RIGHT SIDE SIMPLICITY ABLATION PROCEDURE A SECOND DEGREE PATIENT BURN OCCURRED. FOLLOWING THE PROCEDURE, WHEN THE GROUNDING PAD WAS REMOVED A SECOND DEGREE BURN WAS NOTED ON THE LEFT THIGH WHERE THE PAD WAS LOCATED. THE GROUNDING PAD WAS NOT OVERLAPPING WITH ANY OTHER PAD AND THERE WERE NO WRINKLES OR EDGES LIFTED, HOWEVER; THE PATIENT WAS HAIRY AT THE PAD LOCATION AND THE SKIN WAS NOT PREPPED PRIOR TO PAD PLACEMENT. NO INTERVENTION WAS REQUIRED TO TREAT THE BURN.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 9167096 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page