CATHAY GP202D-ACRecalls & reported problems

Cathay Manufacturing Corp. · Monopolar Adult Grounding Pad with Cable · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
11
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 06959006303261), received between Apr 7, 2023 and May 8, 2023 — classified as 2 injuries, 9 malfunctions.

The most frequently reported specific device problem is “Connection Problem” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Connection Problem10
  • Adverse Event Without Identified Device or Use Problem1
  • Grounding Malfunction1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • Full thickness (Third Degree) Burn1

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 8, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT FOLLOWING A RHIZOTOMY PROCEDURE, THE PATIENT EXPERIENCED A THIRD-DEGREE GROUNDING PAD BURN ON THEIR RIGHT POSTERIOR THIGH. AS A RESULT, THE PATIENT IS BEING TREATED AT A WOUND CARE FACILITY WITH MEDICINAL HONEY.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 16892588 ↗
  • InjuryApr 28, 2023Grounding Malfunction; Connection Problem; Grounding Malfunction; Connection Problem

    RELATED MANUFACTURER REFERENCE NUMBER: 2184149-2023-00093. IT WAS REPORTED THAT THE DURING A RF PROCEDURE ON (B)(6) 2023, AFTER COMPLETING THREE LESIONS THE PHYSICIAN RECEIVED AN ERROR MESSAGE STATING THE GROUNDING PAD DETACHED WHEN A PROBE WAS ADDED FOR A FOURTH LESION. TROUBLESHOOTING WAS UNABLE TO RESOLVE THE ISSUE. AS SUCH, THE PROCEDURE WAS COMPLETED WITH A COMPETITORS DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16838230 ↗
  • MalfunctionApr 11, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00150, 3005334138-2023-00152 IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16716570 ↗
  • MalfunctionApr 11, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00150, 3005334138-2023-00151. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16716562 ↗
  • MalfunctionApr 11, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00151, 3005334138-2023-00152. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16716496 ↗
  • MalfunctionApr 7, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00144, 3005334138-2023-00145, 3005334138-2023-00146, 3005334138-2023-00148, 3005334138-2023-00149. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699381 ↗
  • MalfunctionApr 7, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00144, 3005334138-2023-00145, 3005334138-2023-00146, 3005334138-2023-00147, 3005334138-2023-00148. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699380 ↗
  • MalfunctionApr 7, 2023Connection Problem; Connection Problem

    ADDITIONAL INFORMATION RECEIVED INDICATES THAT THE DOCTOR HAD NOT ATTEMPTED TO INSERT THE PROBE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699379 ↗
  • MalfunctionApr 7, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00144, 3005334138-2023-00145, 3005334138-2023-00147, 3005334138-2023-00148,3005334138-2023-00149. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699377 ↗
  • MalfunctionApr 7, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00144, 3005334138-2023-00146, 3005334138-2023-00147, 3005334138-2023-00148, 3005334138-2023-00149. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699376 ↗
  • MalfunctionApr 7, 2023Connection Problem; Connection Problem

    RELATED MANUFACTURER REPORT NUMBER: 3005334138-2023-00145, 3005334138-2023-00146, 3005334138-2023-00147, 3005334138-2023-00148, 3005334138-2023-00149. IT WAS REPORTED THAT DURING A PROCEDURE, A GROUNDING ERROR MESSAGE APPEARED ON AN RFA GENERATOR. TROUBLESHOOTING REVEALED A CONNECTION PROBLEM WITH THE GROUNDING PAD PLUG AND THE GENERATOR. AS A RESULT, THE PROCEDURE COULD NOT BE COMPLETED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16699375 ↗

Other Electrosurgical, Cutting & Coagulation & Accessories devices

All Electrosurgical, Cutting & Coagulation & Accessories devices →