SERFAS Energy 0279250101Recalls & reported problems

Stryker Corporation · Micro-Claw, Small-Joint Probe, Non-suction, Bendable, Max Cut Level 6 · Class 2

All models of SERFAS Energy (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Micro-Claw small-joint probe is a bendable electrosurgical instrument used to cut and coagulate tissue during minimally invasive procedures such as wrist and ankle arthroscopy. The 29 reports mostly describe the ceramic tip or other parts breaking off during surgery, sometimes leaving fragments in the joint or wound that then had to be removed. Reports describe 10 injuries and no deaths. No recalls name this device, and reports fell from 6 in the prior 12 months to 1 in the past year.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
28
Deaths reported
0
Injuries reported
9
Recalls naming it
0

Summary

The FDA's MAUDE database contains 28 adverse event reports that list this exact device (DI 07613327056877), received between Sep 18, 2018 and Aug 29, 2025 — classified as 9 injuries, 19 malfunctions.

The most frequently reported specific device problem is “Break” (18 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break18
  • Material Fragmentation7
  • Fracture6
  • Device Remains Activated2
  • Detachment of Device or Device Component1
  • Failure to Shut Off1

Patient problems

  • No Clinical Signs, Symptoms or Conditions19
  • Insufficient Information6
  • No Known Impact Or Consequence To Patient6
  • No Consequences Or Impact To Patient5
  • Device Embedded In Tissue or Plaque2
  • Foreign Body In Patient2
  • No Code Available1
  • Pain1

Recent adverse event reports

The 25 most recent of 28 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 29, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING THE PROCEDURE; PART OF THE DEVICE BROKE OFF THE MAIN STRUCTURE AND WENT INTO THE PATIENTS WOUND. THIS WAS LATER RETRIEVED DURING THE SAME PERIOPERATIVE EPISODE. A LARGER WOUND WAS MADE TO RETRIEVE THE FOREIGN BODY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22925135 ↗
  • MalfunctionJul 2, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT FOR A PROBE, DURING THE OPERATION, TWO PIECES OF WHITE PLASTIC CAME OFF. THE PLASTIC PARTS WERE REMOVED FROM THE WRIST.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22385369 ↗
  • MalfunctionJun 25, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT PARTS OF THE TIP BROKE OFF DURING PROCEDURE. ALL OF THE BROKEN PIECES WERE RETRIEVED FROM PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22332172 ↗
  • MalfunctionJun 19, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE SURGEON WAS PERFORMING A RIGHT WRIST ARTHROSCOPY, WHILE USING A MICRO CLAW SMALL PROBE FOR CAUTERY. A SMALL PLASTIC PEACE BROKE OFF THE INSTRUMENT. THE PIECE WAS ABLE TO BE REMOVED FROM WRIST. THE INSTRUMENT AND FRAGMENT WERE THEN REMOVED FROM THE STERILE FIELD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22294187 ↗
  • MalfunctionMar 7, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21551347 ↗
  • InjuryDec 6, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT PART OF THE PRODUCT REMAINED IN THE BODY (ANKLE JOINT). THE PATIENT MENTIONED PAIN DURING EXTENSION. AN X-RAY SUGGESTED THE POSSIBILITY OF CERAMIC REMAINING IN THE PATIENT BODY. THE PART WAS REMOVED FROM PATIENT BODY COMPLETELY IN AN ADDITIONAL PROCEDURE.

    Patient problems: Foreign Body In Patient; Pain

    MAUDE report 20871205 ↗
  • InjuryMar 1, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE TIP OF MICRO CLAW MADE OF CERAMIC WAS BROKEN AND FALL OFF. THE BROKEN PIECE WAS LEFT IN PATIENT AND WILL NEED TO BE REMOVED LATER.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18820108 ↗
  • InjuryApr 27, 2023Break; Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE AND AN ADDITIONAL PROCEDURE WAS REQUIRED TO RETRIEVE THE BROKEN PIECE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16826840 ↗
  • InjuryFeb 22, 2023Break; Break

    IT WAS REPORTED THAT THE DEVICE BROKE DURING PROCEDURE AND PIECES WERE POTENTIALLY UNABLE TO BE RETRIEVED.

    Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions

    MAUDE report 16420530 ↗
  • MalfunctionDec 21, 2022Break

    IT WAS REPORTED THE TIP OF THE PROBE BROKE DURING PROCEDURE. IT WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16024288 ↗
  • MalfunctionJun 21, 2022Failure to Shut Off; Device Remains Activated; Failure to Shut Off

    IT WAS REPORTED THAT THE DEVICE WAS CONTINUOUSLY ACTIVATING.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 14754599 ↗
  • MalfunctionJun 21, 2022Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14754583 ↗
  • MalfunctionApr 26, 2022Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14215173 ↗
  • MalfunctionApr 12, 2022Fracture; Break; Fracture

    IT WAS REPORTED THAT THE TIP OF THE PROBE BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14088115 ↗
  • MalfunctionApr 7, 2022Fracture; Break; Fracture

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BREAKING DURING THE PROCEDURE. THE TIP WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14048116 ↗
  • MalfunctionJan 11, 2022Break

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE OFF PRIOR TO ENTERING THE JOINT.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13230170 ↗
  • MalfunctionJun 22, 2021Break

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12040476 ↗
  • MalfunctionDec 16, 2020Break

    IT WAS REPORTED, THAT THE CERAMIC TIP BROKE, DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 11016422 ↗
  • MalfunctionNov 24, 2020Break

    IT WAS REPORTED THAT THE DEVICE TIP BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10891558 ↗
  • MalfunctionNov 24, 2020Break

    IT WAS REPORTED THAT THE CERAMIC TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10891526 ↗
  • MalfunctionJul 1, 2020Fracture; Break; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 10221489 ↗
  • MalfunctionJun 1, 2020Break

    IT WAS REPORTED THAT THE CERAMIC BROKE OFF OF THE DEVICE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 10105169 ↗
  • MalfunctionMay 19, 2020Fracture; Break; Fracture

    IT WAS REPORTED THAT THE TIP BROKE DURING THE PROCEDURE. ALL PIECES WERE RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 10071117 ↗
  • InjuryMar 31, 2020Break; Fracture; Fracture

    IT WAS REPORTED THAT THE DEVICE BROKE AND THE PIECES REMAIN IN THE PATIENT.

    Patient problems: Device Embedded In Tissue or Plaque; No Consequences Or Impact To Patient

    MAUDE report 9903205 ↗
  • InjuryNov 14, 2019Fracture

    IT WAS REPORTED THAT TIP OF THE DEVICE FRACTURED DURING THE PROCEDURE. IT IS NOT CONFIRMED IF THE TIP WAS RETRIEVED.

    Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient

    MAUDE report 9323363 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates