SERFAS Energy 0279351100Recalls & reported problems

Stryker Corporation · 90-S, Suction Probe, Non-bendable, Max Cut Level 11 · Class 2

All models of SERFAS Energy (17)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker SERFAS Energy suction probe is a non-bendable electrosurgical probe used to cut and coagulate tissue during surgery. The 38 reports mostly describe the probe failing to deliver energy, breaking or melting during a procedure, overheating, and arcing; some describe pieces breaking off or the tip coming loose and remaining inside the body, plus burns to the patient. Reports number 38, with no deaths and 7 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
38
Deaths reported
0
Injuries reported
7
Recalls naming it
0

Summary

The FDA's MAUDE database contains 38 adverse event reports that list this exact device (DI 07613327056952), received between May 23, 2018 and May 28, 2026 — classified as 7 injuries, 31 malfunctions.

The most frequently reported specific device problem is “Failure to Deliver Energy” (8 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 3 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 3 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Deliver Energy8
  • Break7
  • Overheating of Device6
  • Adverse Event Without Identified Device or Use Problem3
  • Arcing3
  • Material Fragmentation3
  • Temperature Problem3
  • Material Integrity Problem2

Patient problems

  • No Clinical Signs, Symptoms or Conditions25
  • No Known Impact Or Consequence To Patient7
  • Insufficient Information3
  • Superficial (First Degree) Burn3
  • Burn(s)2
  • Burn, Thermal2
  • Device Embedded In Tissue or Plaque2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 25 most recent of 38 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 28, 2026Material Fragmentation

    IT WAS REPORTED THAT THE TIP OF THE SERFAS DEVICE BECAME DISLODGED AND REMAINED INSIDE THE BODY. IT HAS SINCE BEEN SUCCESSFULLY REMOVED. THE ELECTRODE AT THE TIP OF THE PRODUCT HAD COME LOOSE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25322150 ↗
  • MalfunctionApr 2, 2026Overheating of Device

    IT WAS REPORTED THAT THE PROBE MELTED DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765195 ↗
  • MalfunctionApr 2, 2026Overheating of Device

    IT WAS REPORTED THAT THE PROBE MELTED DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24765192 ↗
  • MalfunctionApr 1, 2026Overheating of Device

    IT WAS REPORTED THAT THE PROBE MELTED DURING A CASE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24758981 ↗
  • MalfunctionMar 18, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE PROBE HEATED UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21638598 ↗
  • MalfunctionMar 18, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE PROBE HEATED UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21638597 ↗
  • MalfunctionMar 18, 2025Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE PROBE HEATED UP.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21634491 ↗
  • MalfunctionFeb 6, 2025Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS REMOVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21326827 ↗
  • MalfunctionFeb 3, 2025Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21299028 ↗
  • MalfunctionFeb 3, 2025Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT/NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21299026 ↗
  • MalfunctionDec 23, 2024Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21007000 ↗
  • MalfunctionDec 18, 2024Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20972869 ↗
  • MalfunctionDec 6, 2024Failure to Deliver Energy; Failure to Cut; Failure to Deliver Energy; Failure to Cut

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20865155 ↗
  • MalfunctionNov 20, 2024Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20740920 ↗
  • MalfunctionNov 14, 2024Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20691685 ↗
  • MalfunctionNov 14, 2024Failure to Deliver Energy; Failure to Deliver Energy

    IT WAS REPORTED THAT THE DEVICE HAD INSUFFICIENT / NO ENERGY DELIVERED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20687041 ↗
  • MalfunctionAug 9, 2024Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT THE TIP OF THE DEVICE BROKE DURING THE PROCEDURE. THE PIECE WAS RETRIEVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19956563 ↗
  • MalfunctionAug 9, 2024Electrostatic Discharge; Electrostatic Discharge

    IT WAS REPORTED THAT THERE WAS INSULATION DAMAGE RESULTING IN THE LEAKAGE OF ELECTRIC CURRENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19955615 ↗
  • InjuryFeb 23, 2023Temperature Problem; Patient Device Interaction Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A PATIENT BURN.

    Patient problems: Burn(s)

    MAUDE report 16432611 ↗
  • InjuryJan 18, 2023Adverse Event Without Identified Device or Use Problem; Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THERE WAS A PATIENT BURN.

    Patient problems: Burn(s); Superficial (First Degree) Burn

    MAUDE report 16199930 ↗
  • MalfunctionOct 18, 2022Break; Leak/Splash

    IT WAS REPORTED THAT THE DEVICE HAD INSULATION DAMAGE RESULTING IN THE LEAKAGE OF ELECTRIC CURRENT. NO ADVERSE CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15620958 ↗
  • InjuryAug 4, 2022Energy Output Problem; Temperature Problem; Temperature Problem

    IT WAS REPORTED THAT THE PATIENT WAS BURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Superficial (First Degree) Burn

    MAUDE report 15163119 ↗
  • MalfunctionJun 29, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE DEVICE HAD HEAT SHRINK DAMAGE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12086166 ↗
  • MalfunctionMay 6, 2021Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT THE PATIENT WAS BURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Superficial (First Degree) Burn

    MAUDE report 11783188 ↗
  • MalfunctionNov 17, 2020Material Integrity Problem; Break; Material Integrity Problem

    IT WAS REPORTED THAT THE DEVICE HAD INSULATION DAMAGE. ALL PIECES WERE RETRIEVED.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 10857246 ↗