Stryker Corporation · Neptune E-SEP Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch · Class 2

All models of Neptune E-SEP (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Stryker Neptune smoke evacuation pencil has 140 adverse event reports, with no deaths and 14 injuries. Many describe the pencil activating without being pressed or arriving unsterile, while several injuries describe burns to patients, some severe. No recalls name this device. Reports fell to 10 over the last year from 19 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
140
Deaths reported
0
Injuries reported
14
Recalls naming it
0

Summary

The FDA's MAUDE database contains 140 adverse event reports that list this exact device (DI 07613327117189), received between Jun 12, 2018 and May 15, 2026 — classified as 14 injuries, 126 malfunctions.

The most frequently reported specific device problem is “Delivered as Unsterile Product” (29 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 3 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 5 reportsMar 2024: 0 reportsApr 2024: 4 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 1 reportsSep 2024: 1 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 2 reportsFeb 2025: 0 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 5 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 3 reportsOct 2025: 0 reportsNov 2025: 2 reportsDec 2025: 3 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Delivered as Unsterile Product29
  • Unintended Power Up26
  • Fracture12
  • Device Remains Activated11
  • Packaging Problem10
  • Output below Specifications8
  • Energy Output Problem7
  • Sparking6

Patient problems

  • No Clinical Signs, Symptoms or Conditions61
  • No Patient Involvement39
  • Burn(s)12
  • No Consequences Or Impact To Patient9
  • Superficial (First Degree) Burn7
  • Full thickness (Third Degree) Burn5
  • Insufficient Information4
  • Hematoma3

Recent adverse event reports

The 25 most recent of 140 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 15, 2026Unintended Power Up

    IT WAS REPORTED THAT DURING A PROCEDURE THE ESEP SAFEAIR PENCIL ACTIVATED WITHOUT PRESSING ANY BUTTONS BY THE SURGEON. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR INJURY WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25183877 ↗
  • InjuryMar 25, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A HYSTERECTOMY PROCEDURE AT THE USER FACILITY, THE ELECTROSURGICAL PENCIL BECAME VERY HOT AND CAUSED A BURN TO THE PATIENT¿S SKIN. IT WAS REPORTED THAT THE BURN WAS ASSESSED AS THIRD TO FOURTH DEGREE DUE TO THE DEPTH OF TISSUE INVOLVEMENT. IT WAS FURTHER REPORTED THAT THE EVENT REQUIRED MEDICAL INTERVENTION. AFFECTED SKIN AND UNDERLYING ADIPOSE TISSUE WERE REMOVED, ADDITIONAL ANTIBIOTICS WERE ADMINISTERED, AND THE BURN MEASURED APPROXIMATELY 8.5 CM BY 2.5 CM. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 24688279 ↗
  • InjuryMar 25, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A HYSTERECTOMY PROCEDURE AT THE USER FACILITY, THE ELECTROSURGICAL PENCIL BECAME VERY HOT AND CAUSED A BURN TO THE PATIENT¿S SKIN. IT WAS REPORTED THAT THE BURN WAS ASSESSED AS THIRD TO FOURTH DEGREE DUE TO THE DEPTH OF TISSUE INVOLVEMENT. IT WAS FURTHER REPORTED THAT THE EVENT REQUIRED MEDICAL INTERVENTION. AFFECTED SKIN AND UNDERLYING ADIPOSE TISSUE WERE REMOVED, ADDITIONAL ANTIBIOTICS WERE ADMINISTERED, AND THE BURN MEASURED APPROXIMATELY 8.5 CM BY 2.5 CM. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 24688255 ↗
  • MalfunctionJan 28, 2026Unintended Power Up

    IT WAS REPORTED THAT DURING A PROCEDURE THE SMOKE EVAC PENCIL ACTIVATED WITHOUT PRESSING ANY BUTTONS BY THE SURGEON. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR INJURY WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24198730 ↗
  • InjuryJan 16, 2026Unintended Power Up

    IT WAS REPORTED THAT, DURING USE AT THE USER FACILITY, THE USER DID NOT USE THE HOLSTER AND PLACED THE DEVICE ON THE PATIENT¿S LEG, CAUSING A THIRD-DEGREE BURN. IT WAS FURTHER REPORTED THAT THE PENCIL WAS NOT TURNED ON MANUALLY WHEN PLACED ON THE PATIENT¿S LEG. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT DELAY.

    Patient problems: Burn(s)

    MAUDE report 24094985 ↗
  • MalfunctionDec 9, 2025Defective Device

    BOVIE WAS PLUGGED INTO MACHINE PROPERLY BUT DID NOT FUNCTION OUT OF PACKAGE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23756205 ↗
  • MalfunctionDec 1, 2025Unintended Power Up; Unintended Power Up

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23683773 ↗
  • MalfunctionDec 1, 2025Use of Device Problem; Adverse Event Without Identified Device or Use Problem; Use of Device Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO ADDITIONAL INFORMATION PROVIDED, THE NON-STRYKER ELECTRODE WAS USED DURING THE SURGERY.

    Patient problems: Burn(s)

    MAUDE report 23679415 ↗
  • MalfunctionNov 21, 2025Unintended Power Up; Unintended Power Up

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23618281 ↗
  • InjuryNov 21, 2025Unintended Power Up; Unintended Power Up

    IT WAS REPORTED THAT, DURING USE AT THE USER FACILITY, THE DEVICE UNINTENTIONALLY ACTIVATED, CAUSING A THIRD DEGREE BURN, THAT REQUIRED POST OPERATION DRESSING CHANGE ORDERS. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT DELAY.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 23616826 ↗
  • InjurySep 18, 2025Fire; Fire

    IT WAS REPORTED THAT, DURING USE AT THE USER FACILITY, THE E-SEP PENCIL GOT CLOSE TO A PIECE OF HAIR WITH BETADINE AND THE HAIR CAUGHT ON FIRE, CAUSING A SECOND-DEGREE BURN TO THE PATIENT'S FOREHEAD AND EXTENDING ONTO THE FACE. IT WAS FURTHER REPORTED THAT THE BURNS REQUIRED ADDITIONAL TREATMENT IN ANOTHER HOSPITAL. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SIGNIFICANT DELAY.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 23084949 ↗
  • MalfunctionSep 18, 2025Fire; Fire

    IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, A FLAME CAME OUT OF THE PENCIL WHICH CAUSED THE PLASTIC TO MELT. NO INFORMATION WAS PROVIDED REGARDING ANY ADVERSE CONSEQUENCES, MEDICAL INTERVENTION, OR SURGICAL DELAY. IT WAS FURTHER REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS ALSO REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT SIGNIFICANT DELAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23084943 ↗
  • MalfunctionSep 15, 2025Device Remains Activated; Device Remains Activated

    NO NEW INFORMATION.

    Patient problems: Insufficient Information

    MAUDE report 23050467 ↗
  • MalfunctionAug 18, 2025Electrical /Electronic Property Problem; Mechanical Problem

    BOVIE SMOKE EVACUATOR WAS PLUGGED INTO BOVIE MACHINE TO BE USED DURING THE CASE BUT WHEN THE SURGEON WENT TO USE BOVIE IT DID NOT EMIT ELECTRICITY. BOVIE WAS REPLACED WITH A NEW ONE AND NEW BOVIE WORKED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22826762 ↗
  • MalfunctionJun 24, 2025Output below Specifications; Output below Specifications

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22321009 ↗
  • MalfunctionJun 24, 2025Output below Specifications; Output below Specifications

    DURING THE PROCEDURE, IT WAS REPORTED THAT THE COAG FUNCTION OR CUT FUNCTION ON THE E-SEP SAFEAIR PENCIL DID NOT WORK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22321003 ↗
  • MalfunctionJun 24, 2025Output below Specifications; Output below Specifications

    DURING THE PROCEDURE, IT WAS REPORTED THAT THE COAG FUNCTION OR CUT FUNCTION ON THE E-SEP SAFEAIR PENCIL DID NOT WORK. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22320792 ↗
  • MalfunctionJun 17, 2025Device Remains Activated; Device Remains Activated

    IT WAS REPORTED THAT DURING A PROCEDURE, THE BOVIE PENCIL REMAINED ACTIVATED DUE TO A STUCK BUTTON. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT ANY DELAYS, AND NO ADVERSE CONSEQUENCES OR INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22226374 ↗
  • MalfunctionJun 3, 2025Tear, Rip or Hole in Device Packaging; Tear, Rip or Hole in Device Packaging

    IT WAS REPORTED BY THE CUSTOMER THAT UPON RECEIVING, THE DEVICE PACKAGING WAS FOUND TO BE DAMAGED/HAVING HOLE. THERE WAS NO PATIENT INVOLVEMENT, NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22131245 ↗
  • InjuryApr 24, 2025Unintended Power Up; Unintended Power Up

    IT WAS REPORTED THAT DURING THE C-SECTION AT THE USER FACILITY, THE COAGULATION ACTIVATION ALARM WENT OFF, AND IT WAS MISINTERPRETED AS AS A GROUND FAULT ALARM. THE USER CONTINUED USING THE ELECTROSURGICAL PENCIL, RESULTING IN BURNS ON BOTH OF THE BABY'S BUTTOCKS, WHICH REQUIRED TRANSFER TO ANOTHER HOSPITAL FOR TREATMENT IN A BURN UNIT. THIS REPORT IS RELATED TO MFG REPORT # 3015967359-2025-00938.

    Patient problems: Full thickness (Third Degree) Burn

    MAUDE report 21899763 ↗
  • InjuryApr 23, 2025Unintended Power Up; Unintended Power Up

    IT WAS REPORTED THAT DURING THE C-SECTION AT THE USER FACILITY, THE COAGULATION ACTIVATION ALARM WENT OFF, AND IT WAS MISINTERPRETED AS A GROUND FAULT ALARM. THE USER CONTINUED TO USE THE PENCIL AND AS A RESULT THE MOTHER HAD BURNS ON HER INNER THIGHS WHICH REQUIRED TREATMENT. NO INFORMATION WAS PROVIDED REGARDING THE SEVERITY OF THE BURN OR THE TYPE OF THE TREATMENT.

    Patient problems: Burn(s)

    MAUDE report 21890058 ↗
  • MalfunctionMar 26, 2025Excessive Heating; Excessive Heating

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, A HOT CAUTERY TIP ACCIDENTALLY TOUCHED THE PATIENT AND LEFT A BURN MARK. IT WAS REPORTED THAT THERE WERE NO ADVERSE CONSEQUENCES AND NO MEDICAL INTERVENTION AS A RESULT OF THIS EVENT. IT WAS FURTHER REPORTED THAT THE PROCEDURE WAS COMPLETED SUCCESSFULLY, WITHOUT A DELAY.

    Patient problems: Burn(s)

    MAUDE report 21699658 ↗
  • MalfunctionMar 6, 2025Device Remains Activated; Device Remains Activated

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21534946 ↗
  • MalfunctionJan 2, 2025Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT DURING A PROCEDURE THE COAG FUNCTION ON THE E-SEP PENCIL DID NOT WORK BUT THE CUT FUNCTION DID. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21056144 ↗
  • MalfunctionJan 2, 2025Output below Specifications; Output below Specifications

    IT WAS REPORTED THAT DURING A PROCEDURE THE COAG FUNCTION ON THE E-SEP PENCIL DID NOT WORK BUT THE CUT FUNCTION DID. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21056141 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →