Skip to content

Neptune E-SEP 0703046002Recalls & reported problems

Stryker Corporation · Neptune E-SEP Smoke Evacuation Pencil, Push Button, Coated, Bulk Non-Sterile · Class 2

All models of Neptune E-SEP (19)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
9
Deaths reported
0
Injuries reported
1
Recalls naming it
1

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 07613327379488), received between Apr 6, 2020 and Jul 28, 2026 — classified as 1 injury, 8 malfunctions.

The most frequently reported specific device problem is “Unintended Power Up” (4 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 4, 2024Open, Classified

    There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

    Z-0370-2025 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

Most reported problems

Device problems

  • Unintended Power Up4
  • Fire3
  • Adverse Event Without Identified Device or Use Problem1
  • Device Remains Activated1
  • Improper or Incorrect Procedure or Method1
  • Particulates1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Burn(s)2
  • Superficial (First Degree) Burn2
  • No Patient Involvement1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 28, 2026Fire

    It was reported that, during a procedure at the user facility, the device tip caught fire. It was reported that the issue had no impact on the patient. It was further reported that there were no adverse consequences and no medical intervention as a result of this event. No information was provided regarding whether the procedure was completed successfully or whether the issue caused a surgical delay.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50052912 ↗
  • MalfunctionOct 7, 2025Unintended Power Up; Unintended Power Up

    NO NEW INFORMATION.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 23233313 ↗
  • InjuryFeb 10, 2025Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem; Improper or Incorrect Procedure or Method; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING USE IN A PROCEDURE AT THE USER FACILITY, A BURN TO THE PATIENT WHILE USING E-SEP PENCIL. IT WAS FURTHER REPORTED THAT THE BURN REQUIRED MEDICAL INTERVENTION. NO INFORMATION WAS AVAILABLE REGARDING SURGICAL DELAY OR IF PROCEDURE WAS SUCCESSFULLY COMPLETED.

    Patient problems: Burn(s)

    MAUDE report 21343424 ↗
  • MalfunctionJan 2, 2025Unintended Power Up; Unintended Power Up

    NO NEW INFORMATION

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21055089 ↗
  • MalfunctionJul 19, 2024Unintended Power Up; Unintended Power Up

    IT WAS REPORTED THAT DURING A PROCEDURE, THE E-SEP PENCIL ACTIVATED WITHOUT THE BUTTONS BEING PUSHED AND THE PATIENT GOT BURNED. THE PENCIL WAS PLUGGED INTO THE GENERATOR AND ACTIVATED WITHOUT EITHER OF THE BUTTONS BEING DEPLOYED. THE PENCIL WAS REMOVED FROM THE FIELD AND A NEW ONE WAS PROVIDED. IT WAS FURTHER REPORTED THAT THE BURN WAS TREATED WITH MARCAINE AND DERMABOND TO PREVENT INFECTION. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. IT WAS FURTHER REPORTED THAT THE USER DID NOT USE A HOLSTER WHILE THE PENCIL WAS NOT IN USE.

    Patient problems: Burn(s)

    MAUDE report 19780486 ↗
  • MalfunctionMar 20, 2024Fire; Fire

    IT WAS REPORTED THAT DURING AN ANERECTOPLASTY, A FLAME WAS NOTICED COMING FROM THE E-SEP TIP WHEN THE ELECTRODE WAS CHANGED. THEY DIDN'T USE THE STANDARD ELECTRODE THAT CAME IN THE BOVIE. THEY SWITCHED TO THE COLORADO NEEDLE ELECTRODE, A KIRWAN SURGICAL PRODUCT, WHEN THIS NEEDLE WAS PUT IN, IT LOOKED LIKE A LIGHTER EVERY TIME THEY PUSHED COAG. USUALLY THEY USE A PLUG TO ELIMINATE GAS COMING OUT, THEY WERE UNABLE TO DO SO IN THIS SITUATION, EVERY TIME THEY PRESSED COAG, IT FLICKERED LIKE A LIGHTER. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITHOUT A CLINICALLY SIGNIFICANT DELAY; NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18943320 ↗
  • MalfunctionAug 31, 2023Unintended Power Up; Device Remains Activated; Unintended Power Up; Device Remains Activated

    IT WAS REPORTED THAT DURING A PROCEDURE WHEN THE E-SEP PENCIL WAS PLUGGED INTO THE CONSOLE THE "CUT" FUNCTION ACTIVATED. THE PENCIL WAS UNPLUGGED AND RECONNECTED TO THE CONSOLE. THE SAME ISSUE OCCURRED. THE PENCIL WAS REMOVED FROM THE FIELD AND A REPLACEMENT PENCIL WAS USED. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY, MEDICAL INTERVENTION OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17663133 ↗
  • MalfunctionJun 3, 2021Fire; Fire

    IT WAS REPORTED THAT DURING A TOTAL HIP PROCEDURE, THE E-SEP PENCIL CAUGHT FIRE. IT WAS ALSO REPORTED THAT THE SURGEON RECEIVED A BURN ON THE RIGHT FORE-FINGER THAT DID NOT REQUIRE ANY MEDICAL INTERVENTION.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 11931200 ↗
  • MalfunctionApr 6, 2020Particulates; Particulates

    IT WAS REPORTED THAT PRIOR TO THE SURGICAL PROCEDURE, A HAIR WAS FOUND IN THE PACKAGE. IT WAS ALSO REPORTED THAT THIS EVENT DID NOT OCCUR DURING A SURGICAL PROCEDURE, AND THERE WAS NO DELAY OR ADVERSE CONSEQUENCES AS A RESULT OF THIS EVENT.

    Patient problems: No Patient Involvement

    MAUDE report 9929732 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →