HS AMICA AMICA-PROBE (REF APK14150T19V5)Recalls & reported problems
H.S. Hospital Service SPA · AMICA-PROBE 14GX150MM · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 08033055147163), received between Feb 25, 2020 and Oct 24, 2022 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Material Fragmentation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Fragmentation2
- Break1
Patient problems
- No Known Impact Or Consequence To Patient2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 24, 2022Break; Break
PATIENT UNDERGOING IMAGE GUIDED MICROWAVE ABLATION OF LIVER LESION. THE FIRST LESION WAS ABLATED AT 60WX10MIN. THE PROBE WAS THEN REPOSITIONED, AND THE ABLATION CONTINUED AT 80WX12MIN. THE PROBE WAS REMOVED AND POSITIONED IN THE SECOND LESION FOR AN ABLATION AT 80WX10MIN. THE PROBE WAS AGAIN REMOVED. CHAR WAS WIPED FROM THE SHAFT. NO DAMAGE TO THE TIP WAS NOTED. MULTIPLE PASSES WERE REQUIRED TO PLACE THE PROBE IN THE THIRD LESION, WHICH WAS ABLATED AT 80WX10MIN. UPON PROBE REMOVAL, IT WAS NOTED THAT TIP WAS NO LONGER ATTACHED. THE RETAINED TIP WAS VISIBLE IN CT IMAGING. WHAT WAS THE ORIGINAL INTENDED PROCEDURE? IMAGE GUIDED MICROWAVE ABLATION OF LIVER LESION. WHAT PROBLEM DID THE USER HAVE (CHECK ALL THAT APPLY): DEVICE FAILED (E.G. BROKE, COULDN'T GET IT TO WORK OR STOPPED WORKING).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15658023 ↗MalfunctionFeb 25, 2020Material Fragmentation; Material Fragmentation
MICROWAVE ABLATION OF PT WITH HCC, LIVER, SEG IV 4A, W/CIRRHOSIS, 1.0CM WAS ABLATED AT 100W/15M. TRACK ABLATION AT 40W / 4 S. AFTER AMICA 14G, 15CM WAS REMOVED, DR. NOTICED THAT THE DISTAL TIP REMAINED AT THE EDGE OF THE TUMOR ABLATION ZONE VERIFIED BY CT WAS NOT INTACT AND THE TIP OF THE PROBE WAS AT THE EDGE OF THE TUMOR ABLATION ZONE VERIFIED BY CT. THE POST CT SCAN W/CONTRAST SHOWED GOOD COVERAGE OF THE LESION. NO ADVERSE EVENTS NOTED FOR THE PATIENT A THIS TIME. FDA SAFETY REPORT ID#(B)(4)
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9748954 ↗MalfunctionFeb 25, 2020Material Fragmentation; Material Fragmentation
14GX15CM AMICA PROBE TIP MALFUNCTION FOLLOWING STANDARD SAFETY PROTOCOL RESULTING IN DISLODGMENT OF THE PROBE TIP IN THE LIVER DURING ABLATION. THE ABLATION WAS ABORTED WITHIN 15SEC OF THE FIRST ABLATION CYCLE. THE PROBE WAS REMOVED AND A NEW 14GX150MM AMICA PROBE WAS USED TO RE-ATTEMPT ABLATION. FOLLOWING STANDARD SAFETY PRACTICE THE NEW PROBE WAS INSERTED INTO THE AREA OF THE TUMOR. AFTER 1.5MIN INTO THE ABLATION CYCLE, THE MACHINE ABRUPTLY CUT OFF DUE TO SIMILAR ISSUES AS THE FIRST INCIDENT. THE PROBE TIP WAS ALSO FOUND TO BE DISLODGED IN THE LIVER. THE LOT NUMBERS OF THESE TWO DEVICES WERE THE SAME LOT #37210.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9748942 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page