LeVeen™ Standard M001262150Recalls & reported problems

Boston Scientific Corporation · Needle Electrode System · Class 2

All models of LeVeen Standard (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 08714729349594), received between Mar 25, 2024 and Feb 18, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Retraction Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Retraction Problem1
  • Therapeutic or Diagnostic Output Failure1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 18, 2025Retraction Problem; Retraction Problem

    IT WAS REPORTED THAT THE TINES OF THE NEEDLE WERE DIFFICULT TO RETRACT. THIS 3.5/15/25 PH LEVEEN STANDARD NEEDLE ELECTRODE WAS SELECTED FOR A RADIOFREQUENCY ABLATION PROCEDURE TO TREAT A POSSIBLE TUMOR. AFTER SUCCESSFULLY DEPLOYING AND PERFORMING ABLATION SIX TIMES USING THE DEVICE, IT WAS ATTEMPTED TO RETRIEVE THE NEEDLE, BUT THIS WAS UNSUCCESSFUL. MORE FORCE WAS APPLIED WITHOUT SUCCESS; THEREFORE, THEY CAPTURED THE NEEDLE BY PUSHING NON-IONIC WATER INTO THE SIDE OF THE NEEDLE TO CLEAR ANY OBSTACLES. THE PROCEDURE WAS COMPLETED, AND THE NEEDLE WAS REMOVED. ONCE OUTSIDE THE BODY, IT WAS NOTED THAT THERE WERE PARTICLES STUCK IN THE NEEDLE, AND ONCE REMOVED, THE TINES COULD BE RETRACTED NORMALLY. THERE WERE NO PATIENT COMPLICATIONS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21401271 ↗
  • MalfunctionMar 25, 2024Therapeutic or Diagnostic Output Failure; Therapeutic or Diagnostic Output Failure

    IT WAS REPORTED THAT INSUFFICIENT ROLL-OFF OCCURRED. A 3.5CM X 25CM LEVEEN STANDARD NEEDLE ELECTRODE WAS SELECTED FOR USE IN A RADIO FREQUENCY ABLATION PROCEDURE. THE DEVICE WAS USED IN THE ABDOMEN. THE IMPENDANCE DID NOT INCREASE DURING ABLATION, AND IT WAS THOUGHT THAT IT WAS POSSIBLY INSIDE THE BLOOD VESSELS. HOWEVER, IT WAS THEN CONFIRMED WITH ECHOCARDIOGRAPHY THAT IT WAS NOT INSIDE THE BLOOD VESSELS. SINCE THE DEVICE COULD NOT BE USED, THE DEVICE WAS EXCHANGED. ANOTHER SIZE WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18968582 ↗