Skip to content

MEDLINE SIL1455Recalls & reported problems

Medline Industries, Inc. · ELECTRODE, BLADE, 2.75', SIL-INS · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10193489015898), received between Sep 23, 2026 and Sep 24, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Crack” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 4 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack1
  • Material Protrusion/Extrusion1
  • Material Puncture/Hole1
  • Melted1
  • Peeled/Delaminated1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Burn(s)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 24, 2026Sparking

    It was reported that "the product started to spark while being used. The facility has deemed these unsafe for use now and would like to return them for a credit."

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50884876 ↗
  • MalfunctionSep 24, 2026Crack; Melted; Material Puncture/Hole

    It was reported that the side of electrode melted and exposed a hole in the insulation and noticed a small crack leading to the base of the tip. The location of the tip that failed: ¿right above the dark blue part of plastic¿.

    Patient problems: Burn(s)

    MAUDE report 50883283 ↗
  • MalfunctionSep 23, 2026Peeled/Delaminated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863073 ↗
  • MalfunctionSep 23, 2026Material Protrusion/Extrusion

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50863071 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page