Skip to content

MicroFrance® CEV133Recalls & reported problems

Integra Microfrance · MicroFrance® Botella bipolar forceps, 5 mm diameter, 360 mm length, triangular jaws · Class 2

All models of MicroFrance (22)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10381780091806), received between Oct 28, 2025 and Oct 28, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Mechanical Jam” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 28, 2025Mechanical Jam; Mechanical Jam

    A FACILITY REPORTED THAT DURING A PROCEDURE, AND THE FIRST VERY FIRST TIME THEY USED THE PILERS, THEY NOTICED THAT THE FORCEPS CEV133 BIPOLAR 350MM LRG BOTELLA (CEV133) WERE DIFFICULT TO CLOSE, "THEY ARE PARTICULARLY STIFF." IT WAS REPORTED THAT REPLACEMENT UNITS WERE USED TO COMPLETE THE SURGERY. THE PRODUCT WAS NOT IN CONTACT WITH THE PATIENT; THUS, NO INJURY OR DEATH WAS REPORTED. HOWEVER, THERE WAS A 30-MINUTE INCREASE IN SURGICAL TIME TO FIND AN ALTERNATIVE SUITABLE INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23408114 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page