The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780094678), received between Feb 11, 2025 and Dec 17, 2025 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Arcing1
  • No Apparent Adverse Event1
  • Smoking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Burn(s)1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionDec 17, 2025Break; Break

    A FACILITY REPORTED THAT BIPOLAR FORCEPS OF THE BIPOLAR INSERT CEV634-1A 350MM MOUIEL (CEV634-1A) BROKE INSIDE THE PATIENT DURING AN OPERATION. IT WAS REPORTED THAT THE BREAK OCCURRED AT THE TIP OF THE INSERT, A SINGLE PIECE DETACHED AND WAS RECOVERED, AND ANOTHER FORCEPS WERE USED TO COMPLETE THE SURGERY. THERE WAS NO INJURY, DEATH, AND IT WAS UNKNOWN IF THERE WAS ANY INCREASE IN SURGERY TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23834997 ↗
  • InjuryJun 17, 2025Smoking; Arcing; No Apparent Adverse Event; Smoking; Arcing; No Apparent Adverse Event

    THIS REPORT IS 3 OF 3 FOR THE DIA 5MM 350MM TUBE (CEV6795B) COMPONENT OF THE INSTRUMENT REPORTEDLY USED DURING THIS EVENT, AND IS RELATED MFG NUMBERS: 3003249645-2025-00022 AND 3003249645-2025-00023. A FACILITY REPORTED THAT DURING A COELIOSCOPY SURGERY, BLACK SMOKE APPEARED DURING FIRST USE OF THE LAPAROSCOPY SHEATH. AN ELECTRIC ARC OCCURRED WHICH BURNT THE WALL OF THE PATIENT'S ABDOMINAL CAVITY. THE INJURY WAS DESCRIBED AS A 1-CENTIMETER (CM) BURN OF THE PERITONEUM, AT THE OPPOSITE JUNCTION OF THE ANTERIOR AND LATERAL ABDOMINAL WALL OF THE ILIAC FOSSA. IT WAS REPORTED THAT NO TREATMENT WAS NEEDED; HOWEVER, BLACK DEBRIS FROM THE SHEATH THAT HAD SETTLED ON THE PERITONEUM WAS SUCTIONED OUT. A SINGLE-USE COELISCOPIC SCISSORS WAS USED AS A REPLACEMENT TO COMPLETE THE SURGERY. IT WAS REPORTED THAT THERE WAS NO FURTHER CONSEQUENCE TO PATIENT AT DAY 1 AND BEYOND. NO SURGICAL DELAY WAS REPORTED.

    Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions

    MAUDE report 22243399 ↗
  • MalfunctionFeb 11, 2025Break; Break

    IT WAS REPORTED THAT THE 350MM MOUIEL BIPOLAR INSERT ((B)(6)) BROKE IN THE PATIENT'S ABDOMEN AND WAS DIFFICULT TO RETRIEVE. ANOTHER DEVICE WAS AVAILABLE TO FINISH THE SURGERY. NO PATIENT INJURY WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21360611 ↗
  • MalfunctionFeb 11, 2025Break; Break

    N/A.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21360610 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices