MicroFrance® CEV634-1ARecalls & reported problems
Integra Microfrance · Mouiel Bipolar Insert Fenestrated Jaws for Tube CEV6795B · Class 2
All models of MicroFrance (22)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780094678), received between Feb 11, 2025 and Dec 17, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Arcing1
- No Apparent Adverse Event1
- Smoking1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Burn(s)1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 17, 2025Break; Break
A FACILITY REPORTED THAT BIPOLAR FORCEPS OF THE BIPOLAR INSERT CEV634-1A 350MM MOUIEL (CEV634-1A) BROKE INSIDE THE PATIENT DURING AN OPERATION. IT WAS REPORTED THAT THE BREAK OCCURRED AT THE TIP OF THE INSERT, A SINGLE PIECE DETACHED AND WAS RECOVERED, AND ANOTHER FORCEPS WERE USED TO COMPLETE THE SURGERY. THERE WAS NO INJURY, DEATH, AND IT WAS UNKNOWN IF THERE WAS ANY INCREASE IN SURGERY TIME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23834997 ↗InjuryJun 17, 2025Smoking; Arcing; No Apparent Adverse Event; Smoking; Arcing; No Apparent Adverse Event
THIS REPORT IS 3 OF 3 FOR THE DIA 5MM 350MM TUBE (CEV6795B) COMPONENT OF THE INSTRUMENT REPORTEDLY USED DURING THIS EVENT, AND IS RELATED MFG NUMBERS: 3003249645-2025-00022 AND 3003249645-2025-00023. A FACILITY REPORTED THAT DURING A COELIOSCOPY SURGERY, BLACK SMOKE APPEARED DURING FIRST USE OF THE LAPAROSCOPY SHEATH. AN ELECTRIC ARC OCCURRED WHICH BURNT THE WALL OF THE PATIENT'S ABDOMINAL CAVITY. THE INJURY WAS DESCRIBED AS A 1-CENTIMETER (CM) BURN OF THE PERITONEUM, AT THE OPPOSITE JUNCTION OF THE ANTERIOR AND LATERAL ABDOMINAL WALL OF THE ILIAC FOSSA. IT WAS REPORTED THAT NO TREATMENT WAS NEEDED; HOWEVER, BLACK DEBRIS FROM THE SHEATH THAT HAD SETTLED ON THE PERITONEUM WAS SUCTIONED OUT. A SINGLE-USE COELISCOPIC SCISSORS WAS USED AS A REPLACEMENT TO COMPLETE THE SURGERY. IT WAS REPORTED THAT THERE WAS NO FURTHER CONSEQUENCE TO PATIENT AT DAY 1 AND BEYOND. NO SURGICAL DELAY WAS REPORTED.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 22243399 ↗MalfunctionFeb 11, 2025Break; Break
IT WAS REPORTED THAT THE 350MM MOUIEL BIPOLAR INSERT ((B)(6)) BROKE IN THE PATIENT'S ABDOMEN AND WAS DIFFICULT TO RETRIEVE. ANOTHER DEVICE WAS AVAILABLE TO FINISH THE SURGERY. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21360611 ↗MalfunctionFeb 11, 2025Break; Break
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page