CODMAN® 9190002RPRecalls & reported problems
Integra LifeSciences Switzerland Sàrl · CODMAN® Integrated Bipolar Cord and Tubing Set Flying Leads, Rotary Pump · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780521778), received between Feb 21, 2025 and Apr 23, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical /Electronic Property Problem2
- Activation Problem1
- Insufficient Device Problem Information1
Patient problems
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
MEDWATCH#: (B)(4). A FACILITY REPORTED: "DURING PROCEDURE, THE MALLIS IRRIGATING TUBING FOR THE BIPOLAR WAS MALFUNCTIONING. MULTIPLE TUBING PULLED FROM STOCK WITH SAME LOT NUMBER WERE OPENED AND NON-FUNCTIONING. ALL THIS TUBING WAS DISCARDED. ONCE A NEW LOT NUMBER WAS OPENED, THERE WERE NO FURTHER ISSUES IDENTIFIED. THE TEAM PULLED ALL STOCK WITH AFFECTED LOT NUMBER AND WORKED WITH FIELD REPRESENTATIVE ON REPLACING USED STOCK ITEM. NO FURTHER EVENTS HAVE BEEN REPORTED TO DATE." NO PATIENT INJURY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21895206 ↗MalfunctionApr 4, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
MEDWATCH # (B)(4). A FACILITY REPORTED: "CODMAN IRRIGATING BIPOLAR WAS IRRIGATING BUT NOT CAUTERIZING. WHEN THE CORD WAS SWITCHED OUT, THE DEVICE (HANDPIECE AND CONSOLE) WORKED APPROPRIATELY." THE PRODUCT PROBLEM WAS DISCOVERED DURING THE PROCEDURE WHEN TRYING TO STOP A BLEED.
Patient problems: Insufficient Information
MAUDE report 21773368 ↗MalfunctionFeb 27, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem
MEDWATCH: "CODMAN IRRIGATING BIPOLAR SYSTEM WAS IRRIGATING BUT NOT CAUTERIZING. RESOLVED WHEN CORD AND TUBING SET WAS CHANGED. SURGEON SAYS THIS IS AN ONGOING ISSUE, STATING, "IT'S BEEN 50/50 WITH THESE LATELY".
Patient problems: Insufficient Information
MAUDE report 21483105 ↗MalfunctionFeb 21, 2025Activation Problem; Activation Problem
A FACILITY REPORTED: "ATTEMPTED TO USE A CODMAN IRRIGATING BIPOLAR, BUT THE DEVICE ONLY IRRIGATED AND DID NOT CAUTERIZE. WHEN THE TUBING/LEAD SET WAS REPLACED, IT WORKED APPROPRIATELY."
Patient problems: Insufficient Information
MAUDE report 21432805 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.