CODMAN® 9190002RPRecalls & reported problems

Integra LifeSciences Switzerland Sàrl · CODMAN® Integrated Bipolar Cord and Tubing Set Flying Leads, Rotary Pump · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10381780521778), received between Feb 21, 2025 and Apr 23, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Electrical /Electronic Property Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 2 reportsMar 2025: 0 reportsApr 2025: 2 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Electrical /Electronic Property Problem2
  • Activation Problem1
  • Insufficient Device Problem Information1

Patient problems

  • Insufficient Information3
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 23, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDWATCH#: (B)(4). A FACILITY REPORTED: "DURING PROCEDURE, THE MALLIS IRRIGATING TUBING FOR THE BIPOLAR WAS MALFUNCTIONING. MULTIPLE TUBING PULLED FROM STOCK WITH SAME LOT NUMBER WERE OPENED AND NON-FUNCTIONING. ALL THIS TUBING WAS DISCARDED. ONCE A NEW LOT NUMBER WAS OPENED, THERE WERE NO FURTHER ISSUES IDENTIFIED. THE TEAM PULLED ALL STOCK WITH AFFECTED LOT NUMBER AND WORKED WITH FIELD REPRESENTATIVE ON REPLACING USED STOCK ITEM. NO FURTHER EVENTS HAVE BEEN REPORTED TO DATE." NO PATIENT INJURY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21895206 ↗
  • MalfunctionApr 4, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    MEDWATCH # (B)(4). A FACILITY REPORTED: "CODMAN IRRIGATING BIPOLAR WAS IRRIGATING BUT NOT CAUTERIZING. WHEN THE CORD WAS SWITCHED OUT, THE DEVICE (HANDPIECE AND CONSOLE) WORKED APPROPRIATELY." THE PRODUCT PROBLEM WAS DISCOVERED DURING THE PROCEDURE WHEN TRYING TO STOP A BLEED.

    Patient problems: Insufficient Information

    MAUDE report 21773368 ↗
  • MalfunctionFeb 27, 2025Electrical /Electronic Property Problem; Electrical /Electronic Property Problem

    MEDWATCH: "CODMAN IRRIGATING BIPOLAR SYSTEM WAS IRRIGATING BUT NOT CAUTERIZING. RESOLVED WHEN CORD AND TUBING SET WAS CHANGED. SURGEON SAYS THIS IS AN ONGOING ISSUE, STATING, "IT'S BEEN 50/50 WITH THESE LATELY".

    Patient problems: Insufficient Information

    MAUDE report 21483105 ↗
  • MalfunctionFeb 21, 2025Activation Problem; Activation Problem

    A FACILITY REPORTED: "ATTEMPTED TO USE A CODMAN IRRIGATING BIPOLAR, BUT THE DEVICE ONLY IRRIGATED AND DID NOT CAUTERIZE. WHEN THE TUBING/LEAD SET WAS REPLACED, IT WORKED APPROPRIATELY."

    Patient problems: Insufficient Information

    MAUDE report 21432805 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →