Megadyne 0014Recalls & reported problems
Megadyne Medical Products, Inc. · E-Z Clean, Non-stick, PTFE Coated, Electrosurgical Blade Electrode, 6.5 inch (16.5 cm) · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10614559100141), received between May 3, 2021 and Dec 1, 2023 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Human-Device Interface Problem” (2 reports).
1 other recall cover devices cleared under the same FDA submission (K081791, K862221); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Human-Device Interface Problem2
- Material Integrity Problem2
- Break1
- Device Markings/Labelling Problem1
- No Audible Alarm1
Patient problems
- Burn(s)2
- No Clinical Signs, Symptoms or Conditions1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 1, 2023Device Markings/Labelling Problem; Device Markings/Labelling Problem
IT WAS REPORTED THAT IT WAS NOTICED THAT ONE OF THE ITEMS IN THEIR ORDER IS PACKAGED INCORRECTLY. THE OUTER WHITE PRODUCT BOX READS 0014 BUT THE PRODUCT INSIDE THE BOX READS 0014A. THERE WAS NO PROCEDURE OR PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18248660 ↗InjuryJul 27, 2023Human-Device Interface Problem; Material Integrity Problem; Human-Device Interface Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING A BILATERAL AUGMENTATION MAMMOPLASTY, DUAL PLANE, WITH RELEASE OF THE INFERIOR BREAST TISSUE PROCEDURE A LONG BOVIE TIP WAS OPENED ONTO THE SURGICAL FIELD. UPON START OF THE SURGERY, THE BOVIE MACHINE WAS SET TO CUT: 40 AND COAG: 40. THE PATIENT WAS BURNED ON THE RIGHT BREAST INCISION AND THE BOVIE TIP SEEMED TO HAVE BURNED AWAY PART OF THE OUTER COATING OF THE BOVIE TIP. THE USE OF THIS TIP WAS IMMEDIATELY DISCONTINUED, AND DR. STATED HE HAD TO CHANGE THE SURGERY DUE TO THE BURN. A NEW BOVIE TIP WAS OPENED FOR USE BUT WAS PASSED OFF DUE TO THE POTENTIAL OF ANOTHER BURN. PER DR.'S OPERATIVE REPORT: ¿OF NOTE, THE LONG INSULATED BOVIE TIP USED FOR THE DISSECTION IN THE POCKETS WAS DISCOVERED TO HAVE A HOLE IN THE INSULATION. THIS RESULTED IN SKIN BURNS ON THE INFERIOR PORTION OF THE INCISION ON THE RIGHT BREAST. THE BURNED SKIN WAS ABLE TO BE EXCISED AND THE INCISION CLOSED WITHOUT INCIDENT.¿
Patient problems: Burn(s)
MAUDE report 17405587 ↗InjuryJul 26, 2023Human-Device Interface Problem; Material Integrity Problem; Human-Device Interface Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE A LONG BOVIE TIP WAS OPENED ONTO THE SURGICAL FIELD. UPON START OF THE SURGERY, THE BOVIE MACHINE WAS SET TO CUT: 40 AND COAG: 40. THE PATIENT WAS BURNED ON THE RIGHT BREAST INCISION AND THE BOVIE TIP SEEMED TO HAVE BURNED AWAY PART OF THE OUTER COATING OF THE BOVIE TIP. THE USE OF THIS TIP WAS IMMEDIATELY DISCONTINUED, AND DR. STATED HE HAD TO CHANGE THE SURGERY DUE TO THE BURN. A NEW BOVIE TIP WAS OPENED FOR USE BUT WAS PASSED OFF DUE TO THE POTENTIAL OF ANOTHER BURN. PER DR.'S OPERATIVE REPORT: ¿OF NOTE, THE LONG INSULATED BOVIE TIP USED FOR THE DISSECTION IN THE POCKETS WAS DISCOVERED TO HAVE A HOLE IN THE INSULATION. THIS RESULTED IN SKIN BURNS ON THE INFERIOR PORTION OF THE INCISION ON THE RIGHT BREAST. THE BURNED SKIN WAS ABLE TO BE EXCISED AND THE INCISION CLOSED WITHOUT INCIDENT.¿.
Patient problems: Burn(s)
MAUDE report 17398012 ↗MalfunctionMay 3, 2021No Audible Alarm; Break
DURING THE SURGICAL PROCEDURE, IT WAS NOTED THAT THE PATIENT'S SKIN HAD SEVERAL BURN MARKS AROUND THE INCISION SITE. DOCTOR INSPECTED THE BOVIE TIP, WHICH HAD A BREAK IN THE INSULATION. NO ALARM SOUND WAS NOTED OR HEARD FROM THE VALLEY LAB UNIT. THE BOVIE TIP WAS ISOLATED REMOVED AND REPLACED. THE INCIDENT WAS REPORTED TO THE CHARGE NURSE AND THE BOVIE TIP WAS HANDED TO HER.
Patient problems: Superficial (First Degree) Burn
MAUDE report 11759685 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K081791, K862221. One submission often covers many models, so check the product and lot details in each record.
- Z-0486-2010 · Oct 13, 2009 · TerminatedE-Z CLEAN Non Stick Cautery Tip. Label reads in part: "MEGADYNE 11506 SOUTH STATE STREET DRAPER, UTAH 84020 USA***CAT/REF 0029M E-Z CLEAN NON STICK CAUTERY TIP***STERILE *** Expir
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page