Megadyne 0030Recalls & reported problems
Megadyne Medical Products, Inc. · Electrosurgical Pencil Rocker Switch E-Z Clean Coated Blade Electrode 10 ft (3 m) cord · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10614559100462), received between Feb 14, 2023 and Jun 12, 2024 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Arcing of Electrodes” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Arcing of Electrodes4
- Human-Device Interface Problem3
- Adverse Event Without Identified Device or Use Problem2
Patient problems
- Burn(s)3
- No Clinical Signs, Symptoms or Conditions2
- Partial thickness (Second Degree) Burn2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 12, 2024Arcing of Electrodes; Arcing of Electrodes
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE WITH THE APPROPRIATE PARAMETERS ALREADY MODIFIED (POWER IN THE GENERATOR), IT IS EVIDENT THAT IT GENERATES AN UNUSUAL FLAME IN THE SURGICAL FIELD, THE ELECTROCOAGULATION CONFIGURATION IS REDUCED TO AVOID BURNS. IN THE PATIENT. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19518399 ↗MalfunctionMar 12, 2024Arcing of Electrodes; Human-Device Interface Problem; Arcing of Electrodes; Human-Device Interface Problem
IT WAS REPORTED THAT DURING A TENDON REPAIR THE ASEPSIS IS PERFORMED IN THE AREA TO BE OPERATED ON (WITH 4% CHLORHEXIDINE), PLACEMENT OF STERILE SURGICAL DRAPES, FIRST INCISION IS MADE WITH AN ELECTROSCALPEL TIP, ON CONTACT WITH THE SKIN IT PRESENTS A SPARK AND SLIGHT FLAME, A CHANGE OF DEVICE IS PERFORMED, THE AFFECTED AREA WAS CHECKED AND THE BURN OF THE EDGES OF THE SKIN WAS CORRECTED, WITH NO CONSEQUENCES FOR THE PATIENT, THERE WAS NO DELAY IN THE SURGERY. MEGAPOWER CONSOLE IS REMOVED FROM THE ROOM FOR PRIOR EVALUATION BY BIOMEDICS. THERE WAS NO SURGICAL DELAY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED. THERE WERE NO PATIENT CONSEQUENCES.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 18884918 ↗InjuryFeb 14, 2023Arcing of Electrodes; Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem; Arcing of Electrodes; Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A RESECTION OF SUBCUTANEOUS CELL TISSUE TUMOR IN THE POSTERIOR NECK THE PATIENT WAS IN THE FACE DOWN POSITION WITH OXYGEN CANNULA AND FACE MASK. THE PROCEDURE BEGAN AND THE ELECTIVE PENCIL GENERATED A SPARK, WHICH IS WHY ANOTHER PENCIL WAS REQUESTED, WHICH WHEN ACTIVATED GENERATED A FLAME OF FIRE ON THE PATIENT, CAUSING A BURN ON THE SKIN OF THE RIGHT CHEEKBONE OF THE PATIENT. THE FACE, BEHIND THE EAR AND THE HAIR. THE PERSONNEL CONTROL THE FIRE AND FINISH THE PROCEDURE WITH ANOTHER MEDTRONIC ELECTRO PENCIL AND THE VALLEYLAB CONSOLE. THE SURGERY WAS DELAYED 20 MINUTES. THE FLAMES WERE EXTINGUISHED, THE PATIENT WAS CLEANED AND THE PROCEDURE WAS FINISHED WITH A VALLEYLAB CONSOLE AND AN ELECTRO PEN FROM MEDTRONIC.
Patient problems: Burn(s); Partial thickness (Second Degree) Burn
MAUDE report 16375372 ↗InjuryFeb 14, 2023Arcing of Electrodes; Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem; Arcing of Electrodes; Human-Device Interface Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A RESECTION OF SUBCUTANEOUS CELL TISSUE TUMOR IN THE POSTERIOR NECK THE PATIENT WAS IN THE FACE DOWN POSITION WITH OXYGEN CANNULA AND FACE MASK. THE PROCEDURE BEGAN AND THE ELECTIVE PENCIL GENERATED A SPARK, WHICH IS WHY ANOTHER PENCIL WAS REQUESTED, WHICH WHEN ACTIVATED GENERATED A FLAME OF FIRE ON THE PATIENT, CAUSING A BURN ON THE SKIN OF THE RIGHT CHEEKBONE OF THE PATIENT. THE FACE, BEHIND THE EAR AND THE HAIR. THE PERSONNEL CONTROL THE FIRE AND FINISH THE PROCEDURE WITH ANOTHER MEDTRONIC ELECTRO PENCIL AND THE VALLEYLAB CONSOLE. THE SURGERY WAS DELAYED 20 MINUTES. THE FLAMES WERE EXTINGUISHED, THE PATIENT WAS CLEANED AND THE PROCEDURE WAS FINISHED WITH A VALLEYLAB CONSOLE AND AN ELECTRO PEN FROM MEDTRONIC.
Patient problems: Burn(s); Partial thickness (Second Degree) Burn
MAUDE report 16375193 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page