Megadyne 0038HRecalls & reported problems

Megadyne Medical Products, Inc. · Electrosurgical Pencil Rocker Switch Non Coated Blade Electrode 10 ft (3 m) cord Holster · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 10614559100844), received between Feb 22, 2021 and Jun 25, 2024 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “Arcing of Electrodes” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Arcing of Electrodes2
  • Adverse Event Without Identified Device or Use Problem1
  • Delivered as Unsterile Product1
  • Product Quality Problem1
  • Self-Activation or Keying1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 25, 2024Self-Activation or Keying; Self-Activation or Keying

    IT WAS REPORTED THAT DURING UNKNOWN PROCEDURE, THE DEVICE HAD BEEN TEMPORARILY STORED IN A NON-WOVEN POCKET AND SUDDENLY STARTED MOVING. THE DEVICE DID NOT STOP EVEN AFTER THE BUTTONS WERE PRESSED AGAIN, SO IT WAS REMOVED FROM THE MAIN BODY OF THE ELECTROCAUTERY SCALPEL AND ITS USE WAS STOPPED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19604165 ↗
  • MalfunctionOct 18, 2022Arcing of Electrodes; Arcing of Electrodes

    SEE ATTACHED MEDWATCH# 1721194-2022-00071.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15624578 ↗
  • MalfunctionOct 4, 2022Product Quality Problem; Product Quality Problem

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE AN INSECT WAS FOUND INSIDE THE PACKAGE. THE DEVICE WAS NOT USED FOR THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15536935 ↗
  • MalfunctionAug 18, 2022Arcing of Electrodes; Arcing of Electrodes

    IT WAS REPORTED THAT DURING AN APPENDECTOMY THE BOVIE ARCED NO STEPS WERE TAKEN TO ADJUST SETTINGS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15252329 ↗
  • InjuryApr 29, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A MASTECTOMY PROCEDURE A PATIENT WAS BURNED AT THE SURGICAL SITE WHILE THE MONOPLANE PENCIL WAS USED. THE CUSTOMER IS KEEPING THE PRODUCT UNTIL THE INCIDENT REPORT IS CLEAR. THEY HAVE ALSO HAD TO CONTACT THE ELECTROSURGICAL GENERATOR COMPANY, MEDTRONIC, TO SEE WHAT STEPS NEED TO BE TAKEN.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 11747206 ↗
  • MalfunctionFeb 22, 2021Delivered as Unsterile Product; Delivered as Unsterile Product

    IT WAS REPORTED THAT WHEN THE DEVICE ARRIVED, THE INNER BOX WAS OPENED. THE BLADE TIP WAS NOT BROKEN OFF AND NO PIECES FELL INTO THE PATIENT. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT. NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11357743 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices