Skip to content

Megadyne 0855CLRecalls & reported problems

Megadyne Medical Products, Inc. · Patient Return Electrode, Single-Use, Contact Quality Monitoring, Adult, With 15 ft (4.5 m) Cord. For Patients Weighing Over 33lbs. (15kg) · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10614559104019), received between Feb 7, 2019 and Mar 28, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Arcing of Electrodes” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Arcing of Electrodes1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1
  • Partial thickness (Second Degree) Burn1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 28, 2025Arcing of Electrodes; Arcing of Electrodes

    IT WAS REPORTED THAT DURING A COLOSTOMY TAKE DOWN PROCEDURE THAT A DRY SPONGE CAUGHT ON FIRE AFTER ENERGY FROM THE TIP OF THE PENCIL "ARC'D" TOWARDS THE SPONGE DURING A PROCEDURE INVOLVING A MEGEN1 AND TELESCOPING SMOKE PENCIL (10 FT). THE DRY SPONGE WAS REPORTEDLY 2 INCHES AWAY FROM THE TIP OF THE PENCIL WHILE ACTIVATING. THE BRAND OF SMOKE EVACUATOR WAS NOT REPORTED ON THE CALL. THE CUSTOMER SENT THE MEGEN1 TO BIOMED FOR TROUBLESHOOTING BUT THEY THREW AWAY THE GROUNDING PAD AND TELESCOPING PENCIL. NO PATIENT CONSEQUENCE OR PROCEDURE DELAY REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21719253 ↗
  • InjuryFeb 7, 2019Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING AN ATRIAL FIBRILLATION ABLATION PROCEDURE A SECOND-DEGREE PATIENT BURN OCCURRED. FOLLOWING THE PROCEDURE, A SECOND-DEGREE BURN WAS LOCATED AT THE SITE OF THE DISPOSABLE PATIENT RETURN ELECTRODE.

    Patient problems: Partial thickness (Second Degree) Burn

    MAUDE report 8318585 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page