Megadyne 004125Recalls & reported problems
Megadyne Medical Products, Inc. · Suction Coagulator, Handswitching, 10 Fr, 6 inch (15.24cm) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10614559105146), received between Nov 29, 2018 and Dec 31, 2020 — classified as 1 injury, 11 malfunctions.
The most frequently reported specific device problem is “Device Remains Activated” (6 reports).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K072559); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jun 15, 2023Class IIOpen, Classified
An issue with the Suction Coagulator product family may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit (ESU).
Z-2257-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Device Remains Activated6
- Self-Activation or Keying3
- Adverse Event Without Identified Device or Use Problem1
- Failure to Fire1
- Failure to Shut Off1
Patient problems
- No Consequences Or Impact To Patient10
- No Clinical Signs, Symptoms or Conditions1
- Partial thickness (Second Degree) Burn1
- Tissue Damage1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 31, 2020Failure to Fire
COAGULATER WOULD NOT FIRE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11100340 ↗InjuryOct 31, 2019Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE DOCTOR PERFORMED A TONSILECTOMY AND ADENOIDECTOMY ON A PATIENT AND, AFTER THE PROCEDURE WAS COMPLETED, NOTICED A SECOND DEGREE BURN WITH BLISTERS TO THE RIGHT CORNER OF THE PATIENT'S MOUTH, ON BOTH THE LIP AND MOUTH. THE PATIENT WAS TREATED WITH BACITRACIN OINTMENT AND RELEASED.
Patient problems: Partial thickness (Second Degree) Burn; Tissue Damage
MAUDE report 9264015 ↗MalfunctionDec 10, 2018Device Remains Activated
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED. ALL DEVICES WERE DISCARDED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148694 ↗MalfunctionDec 10, 2018Self-Activation or Keying
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148638 ↗MalfunctionDec 10, 2018Device Remains Activated
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148601 ↗MalfunctionDec 10, 2018Self-Activation or Keying
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148573 ↗MalfunctionDec 10, 2018Device Remains Activated
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148565 ↗MalfunctionDec 10, 2018Device Remains Activated
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148546 ↗MalfunctionDec 10, 2018Device Remains Activated
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148544 ↗MalfunctionDec 10, 2018Device Remains Activated
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148515 ↗MalfunctionDec 10, 2018Self-Activation or Keying
IT WAS REPORTED THAT DURING A NASAL PROCEDURE, THE BUTTONS ON THE HAND HELD SUCTION COAGULATOR WOULD CONTINUE TO ACTIVATE THE DEVICE, AFTER THE DOCTOR RELEASED THE BUTTON THE DEVICE. THE DOCTOR USED FOOTSWITCH ACTIVATION TO COMPLETE THE PROCEDURE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8148380 ↗MalfunctionNov 29, 2018Failure to Shut Off
IT WAS REPORTED THAT DURING A TONSILS ADENOIDS, THE SURGEON WAS USING A SUCTION COLAGATOR AND THEY WERE ON A FOOT PEDAL, HE WAS REMOVING THE DEVICE FROM THE PATIENT AND ENDING UP BURNING THE PATIENT ON THE LIP. SMALL ELECTROCAUTERY BURN (4X4MM) ON THE RIGHT SUPERIOR LIP BORDER NEAR ORAL COMMISSURE. WHEN THE SURGEON RELEASED HIS FOOT FROM THE PETAL, THE ENERGY DID NOT STOP TO THE DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8118201 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072559. One submission often covers many models, so check the product and lot details in each record.
- Z-1151-2014 · Feb 27, 2014 · TerminatedHand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek polyester peel pouch, 25 pouches per box. This device i
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page