Megadyne 004225Recalls & reported problems
Megadyne Medical Products, Inc. · Suction Coagulator, Handswitching, 8 Fr, 6 inch (15.24cm) · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10614559105153), received between Dec 12, 2024 and Dec 12, 2024 — classified as 1 injury.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
1 other recall cover devices cleared under the same FDA submission (K072559); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Recalls naming this device
- Jun 15, 2023Class IIOpen, Classified
An issue with the Suction Coagulator product family may lead to fluid ingress into the handpiece. Fluid ingress can lead to intermittent device activation, non-activation, or self-activation when plugged into the electrosurgery unit (ESU).
Z-2257-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Patient Device Interaction Problem1
Patient problems
- Hemorrhage/Blood Loss/Bleeding1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryDec 12, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED VIA JOURNAL ARTICLE TITLE: DIRECT TUMORAL PUNCTURE EMBOLIZATION OF ONYX BEFORE ENDOSCOPIC ENDONASAL RESECTION OF A MASSIVE JUVENILE ANGIOFIBROMA FOR A PATIENT WITH VON WILLEBRAND DISEASE ¿ A CASE REPORT AND REVIEW OF LITERATURE AUTHOR(S): LAURA-ELISABETH GOSSELIN, PIERRE-OLIVIER CHAMPAGNE, PASCALE LAVOIE, SYLVIE NADEAU. CITATION: INTERNATIONAL JOURNAL OF SURGERY CASE REPORTS 120 (2024) 109784. THE AIM OF THIS STUDY IS TO DEMONSTRATE THE SAFETY AND THE USEFULNESS OF THIS EMBOLIZATION TECHNIQUE ON A PEDIATRIC CASE OF JUVENILE ANGIOFIBROMA AND TO COMPARE OUR RESULTS TO PREVIOUSLY REPORTED CASES. THIS IS A CASE OF A 12-YEAR-OLD PATIENT SUFFERING FROM VON WILLEBRAND DISEASE PRESENTED WITH VOLUMINOUS JUVENILE ANGIOFIBROMA WITH INTRACRANIAL EXTENSION. INTERNAL CAROTID ARTERY BRANCHES PARTIALLY VASCULARIZED THE TUMOR. THE PATIENT UNDERWENT SURGICAL INTERVENTION WITH ENDONASAL ENDOSCOPIC APPROACH. SURGICAL RESECTION OF THE ANTERIOR TUMORAL SECTION AND A PARTIAL DEBULKING OF THE RIGHT ORBIT WAS ACHIEVED WITH MEGADYNE¿ SUCCION COAGULATORS (ETHICON©). HOWEVER, AFTER A PARTIAL RESECTION OF THE NASOPHARYNX PORTION, AN IMPORTANT TUMORAL BLEEDING COMPELLED SURGEONS TO STOP THE SURG
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20912695 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K072559. One submission often covers many models, so check the product and lot details in each record.
- Z-1151-2014 · Feb 27, 2014 · TerminatedHand Control 12FR Suction Coagulator with 10 ft cord and holster, Sterile, Disposable, Model Number 0043-25, packaged Tyvek polyester peel pouch, 25 pouches per box. This device i
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.