MEGADYNE 251015JRecalls & reported problems
Megadyne Medical Products, Inc. · Megadyne Telescoping Smoke Evacuation Pencil - 15 ft tubing with Universal Connector and Holster · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10614559105580), received between Nov 8, 2021 and May 1, 2025 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Flaked” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available3
- Flaked3
- Delivered as Unsterile Product1
- Difficult or Delayed Activation1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Burn(s)1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 1, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
IT WAS REPORTED THAT WHILE STOCKING SHELVES IT WAS FOUND THAT THE DEVICE HAS A DAMAGED STERILE BLISTER PACKAGE. THERE WAS NO PROCEDURE OR PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21941482 ↗MalfunctionNov 21, 2022Flaked; Appropriate Device Problem Term/Code Not Available; Flaked; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A LAP BAND REMOVAL DUE TO GASTRIC NECROSIS PROCEDURE WHILE ACTIVATING THE PENCIL MADE A ¿POP¿ SOUND AND THEN A FLASH OF FLAME WAS SEEN THAT EXTINGUISHED ON ITS OWN WITH NO HARM TO THE PATIENT. THE STICKY PAD WAS THEN REMOVED AND REPLACED, AND SURGEON THEN SWITCHED TO HARMONIC TO CONTINUE WITH A PORTION OF THE CASE. SURGEON THEN LATER WENT BACK TO USING THE BOVIE WITH NOT FURTHER ISSUE. THERE WERE NO ERROR CODES REPORTED ON THE GENERATOR. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15834199 ↗MalfunctionNov 21, 2022Flaked; Appropriate Device Problem Term/Code Not Available; Flaked; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING A LAP BAND REMOVAL DUE TO GASTRIC NECROSIS PROCEDURE THAT WHILE ACTIVATING THE BOVIE PENCIL THE PENCIL MADE A ¿POP¿ SOUND THEN THERE WAS A FLAME WHICH IGNITED THE SPONGE THAT WAS NEARBY. FIRE WAS EXTINGUISHED WITH NO HARM TO THE PATIENT. THE STICKY PAD WAS THEN REMOVED AND REPLACED, AND SURGEON THEN SWITCHED TO HARMONIC TO CONTINUE WITH A PORTION OF THE CASE. SURGEON THEN LATER WENT BACK TO USING THE BOVIE WITH NO FURTHER ISSUE. THERE WERE NO ERROR CODES REPORTED ON THE GENERATOR. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15834124 ↗InjuryNov 8, 2021Flaked; Difficult or Delayed Activation; Appropriate Device Problem Term/Code Not Available; Flaked; Difficult or Delayed Activation; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE BOVI ARCED, CAUGHT THE DRAPE ON FIRE, AND THE PATIENT SUFFERED A SEVERE BURN. NO OTHER INFORMATION IS AVAILABLE.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 12770991 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page