CONMED 130317Recalls & reported problems
Conmed Corporation · CONMED ELECTROSURGICAL PENCIL, REUSABLE, WITH BUTTON SWITCH AND DISPOSABLE BLADE ELECTRODE · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10653405002208), received between Jun 30, 2025 and Jun 30, 2025 — classified as 1 injury.
The most frequently reported specific device problem is “Electrical Shorting” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Electrical Shorting1
- Therapy Delivered to Incorrect Body Area1
Patient problems
- Full thickness (Third Degree) Burn1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJun 30, 2025Therapy Delivered to Incorrect Body Area; Electrical Shorting; Therapy Delivered to Incorrect Body Area; Electrical Shorting
A CUSTOMER REPORTED THAT THE 130317, ELECTROSURGICAL PENCIL, REUSABLE DEVICE WAS USED IN A TONSILLECTOMY PROCEDURE ON (B)(6)2025, AND ¿THE ITEM BURNED THE PT. THE PROCEDURE WAS COMPLETED WITH ANOTHER PEN.¿. PER FURTHER ASSESSMENT, THE DEVICE "APPEARS TO HAVE SENT AN ELECTRICAL CURRENT THROUGH THE SIDE OF THE HANDPIECE NEAR THE BOTTOM.¿. THE SURGEON INDICATED THE INJURY ¿APPEARS TO BE THIRD DEGREE" BURN. ¿THE SURGICAL INTERVENTION THAT WAS REQUIRED WAS A REPAIR TO THE COMMISSURE AND INSIDE THE PATIENT'S MOUTH. IT TOOK ABOUT 30 MINUTES FOR THE DOCTOR TO REPAIR. SINCE THE BURN, THE PATIENT HAS HAD ONE ADDITIONAL SURGERY ON THE COMMISSURE WITH A MICROFLAP CLOSURE TO HELP PREVENT ANY LASTING DAMAGE TO THEIR SMILE. THAT WAS ALSO DONE AS AN OUTPATIENT PROCEDURE, AND THE PATIENT IS HOME RECOVERING AND BEING FOLLOWED BY THE SURGEON.¿. THE SURGERY WAS COMPLETED WITH THE USE OF AN ESU-VALLEYLAB FORCE FX DEVICE AND A DELAY OF 30 MINUTES. NOTICE WAS RECEIVED ON 23JUN2DDDDD 2023 THAT THE CUSTOMER HAS REPORTED THIS INCIDENT UNDER MEDWATCH REPORT NO. MW5171323. A SEARCH OF THE COMPLAINT SYSTEM HAS NOT FOUND A COMPLAINT REPORTED FOR THIS DEVICE/LOT NUMBER DURING THIS TIMEFRAME. THIS REPORT IS BEING
Patient problems: Full thickness (Third Degree) Burn
MAUDE report 22357069 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page