UltraClean 139104EXTRecalls & reported problems
Conmed Corporation · UltraClean Accessory Electrode 1" (2.54 cm) Coated Blade with Extended Insulation · Class 2
All models of UltraClean (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10653405007197), received between Feb 4, 2016 and Jan 22, 2025 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Delivered as Unsterile Product” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Delivered as Unsterile Product1
- Device Packaging Compromised1
- Packaging Problem1
Patient problems
- No Patient Involvement2
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 22, 2025Delivered as Unsterile Product; Delivered as Unsterile Product
DURING INCOMING INSPECTION, THE DISTRIBUTOR REJECTED THIS DEVICE, 139104EXT ACCESSORY ELECTRODE COATED BLADE WITH EXTENDED INSULATION, FOR AN INSUFFICIENT HEATSEAL. THERE WAS NO CONTACT WITH THE PATIENT AS THIS WAS FOUND DURING INCOMING INSPECTION. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21207564 ↗MalfunctionFeb 13, 2018Packaging Problem; Packaging Problem
THE DISTRIBUTOR IN (B)(6) REPORTED A "SCRATCH" ON THE STERILE POUCH OF THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT AS THIS DEFECT WAS DISCOVERED DURING INCOMING INSPECTION. THIS REPORT IS RAISED ON THE BASIS OF A STERILITY BREACH FOUND DURING THE DEVICE EVALUATION.
Patient problems: No Patient Involvement
MAUDE report 7267916 ↗MalfunctionFeb 4, 2016Device Packaging Compromised
THE DISTRIBUTOR IN (B)(4) REPORTED THAT DURING RECEIVING AND INSPECTION OF INCOMING PRODUCTS, "INSUFFICIENT HEAT SEAL" WAS FOUND IN THE SEALING AREA OF A STERILE PACKAGE CONTAINING THE ULTRACLEAN 1" COATED BLADE WITH EXTENDED INSULATION. TESTING RESULTS CONFIRMED THAT THE RETURNED PACKAGE EXHIBITING CREASES IN THE FINAL MANUFACTURING SEAL FAILED THE DYE LEAK TEST ON (B)(6) 2016. THERE WAS NO PATIENT INVOLVEMENT WITH THIS REPORTED DEVICE, AS THE PACKAGING ANOMALY WAS DISCOVERED DURING INSPECTION AT THE DISTRIBUTOR FACILITY PRIOR TO DISTRIBUTION TO AN END-USER.
Patient problems: No Patient Involvement
MAUDE report 5410537 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.