Skip to content

HYFRECATOR 2000 7-900-115 (UDI 10653405031598)Recalls & reported problems

Conmed Corporation · HYFRECATOR 2000 150VAC - 50/60 HZ · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10653405031598), received between Oct 16, 2025 and Oct 16, 2025 — classified as 1 injury.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information1

Patient problems

  • Burn(s)1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 16, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THE CUSTOMER REPORTED THAT THE DEVICE, 7-900-115, HYFRECATOR 2000, 115V/50-60HZ, WAS BEING USED DURING LIPOMA EXCISION PROCEDURE ON (B)(6) 2025 WHEN IT WAS REPORTED, ¿DURING THE PROCEDURE, DR. JING USED OUR HYFRECATOR WHICH CAUSED TWO BURN MARKS ON THE PATIENTS LEFT LEG. A PT WAS INJURED TODAY DURING A LIPOMA EXCISION OF HIS RIGHT TEMPORAL. THIS DEVICE HAS BEEN USED SEVERAL TIMES BY MULTIPLE PROVIDERS AND IS UPDATE WITH INSPECTION. IT IS KEPT SECURELY IN A DRY LOCATION. I AM NOT SURE WHAT COULD BE THE REASON FOR THE MALFUNCTION.". FURTHER ASSESSMENT WAS REQUESTED FROM THE REPORTER, AND THEY STATED THAT NO FURTHER INFORMATION WAS AVAILABLE. THE TYPE OF MALFUNCTION OR RELATIONSHIP BETWEEN THE DEVICE AND HOW THE BURN WAS CAUSED IS NOT KNOWN AT THIS TIME. THIS REPORT IS BEING RAISED DUE TO THE REPORTED INJURY UNKNOWN DEGREE OF BURN TO PATIENT.

    Patient problems: Burn(s)

    MAUDE report 23308901 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page