COOLIEF* Cooled Radiofrequency Pain Management Generator 42297Recalls & reported problems
Avanos Medical, Inc. · PMP,ADVANCED,RFGEN,115230,V4.11,1 · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10680651422979), received between Apr 3, 2024 and Sep 1, 2026 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Excessive Heating” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Excessive Heating1
- Sparking1
Patient problems
- Burn(s)1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 1, 2026Sparking
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50545024 ↗InjuryApr 3, 2024Excessive Heating; Excessive Heating
IT WAS REPORTED THE "MACHINE WAS NOT ABLE TO ACHIEVE BURNING TEMPERATURE, WITH IMPEDANCE RISING FROM 500 TO 2200 OHMS PRIOR TO SELF-TERMINATING AFTER APPROXIMATELY 40 SEC ATTEMPTED FURTHER 2 ATTEMPTS AT RF AT ABOVE SETTINGS WITH SAME ISSUE PULLED BACK BLANKET TO ENSURE DIATHERMY PAD WAS WELL ADHERED TO THE LEG IN CASE THAT WAS THE ISSUE AND NOTED CHARRING. IMMEDIATELY TURNED OFF THE MACHINE AND INSTRUCTED [THE PHYSICIAN] TO TERMINATE THE PROCEDURE." THE PATIENT'S INJURIES NOTED TWO POINTS OF BURN ON THE CORRESPONDING TO DIAGONAL POINTS OF THE DIATHERMY PAD - TWO BURNS ON THIGH WITH CHARRING. "UNCLEAR IF ARCED OVER SKIN OR THROUGH MUSCLE. MEDICAL TREATMENT WAS PROVIDED SECONDARY TO BURNS TO THIGHS. THE CLINICIANS "IMPRESSION IS THAT THERE WAS POOR CONTACT OF THE SKIN ADHESIVE TO THE SKIN DUE TO INTERFERENCE FROM HAIR." MEDICAL TREATMENT "WAS PROVIDED SECONDARY TO BURNS TO THIGHS."
Patient problems: Burn(s)
MAUDE report 19037430 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page