EM3 Monopolar Adapter, Universal EM3-60Recalls & reported problems
Encision, Inc. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10815288012983), received between Sep 21, 2020 and Sep 23, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Sparking” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Sparking4
- Melted2
- Fire1
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 23, 2024Melted; Sparking; Melted; Sparking
THE EM3-60 CABLE, WHICH IS CONNECTED TO THE ESU AND AEM MONITOR, BURNED IN THE OPERATING ROOM DURING SURGERY. THERE WERE NO INJURIES TO PATIENTS OR USERS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20278381 ↗MalfunctionMay 31, 2023Melted; Sparking; Melted; Sparking
THE EM3-60 CABLE, WHICH IS CONNECTED TO THE ESU AND AEM MONITOR, BURNED IN THE OPERATING ROOM DURING SURGERY. THERE WERE NO INJURIES TO PATIENTS OR USERS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17030528 ↗MalfunctionNov 23, 2021Sparking; Sparking
THE EM3-60 CABLE ADAPTER DISPLAYED BURNED OUTER INSULATION, INDICATING THE WIRE MAY HAVE BROKE AND BURNED THROUGH THE INSULATION. THE CABLE SPARKED IN THE OPERATING ROOM. THE LOCATION OF THE FAILURE OCCURED ON THE FOOT SWITCHING SIDE OF THE CABLE. THERE WAS NO INDICATION OF HARM TO PATIENTS OR USERS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12863169 ↗MalfunctionJul 27, 2021Sparking; Sparking
AN EM3-60 CABLE ADAPTER, WHICH IS CONNECTED TO AN ESU AND MONITORING DEVICE, DISPLAYED SMOKE AND SPARKS DURING A LAPAROSCOPIC ELECTROSURGICAL APPENDECTOMY PROCEDURE. POWER FROM THE ESU WAS SHUT OFF WHEN THE EVENT WAS DETECTED. NO PATIENTS OR USERS WERE HARMED DURING THIS EVENT OR AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12233399 ↗MalfunctionSep 21, 2020Fire
A TRANSFER CABLE USED IN ELECTROSURGERY HAD AN INSULATION BREACH WHICH RESULTED IN A VISIBLE FLAME. THIS IS THE THIRD TIME A CABLE HAS FAILED IN THE FIELD, THOUGH THIS IS THE ONLY ONE THAT RESULTED IN A FLAME. THIS FAILURE MODE IS BEING TRENDED. THE CABLE IS NON-PATIENT CONTACT, THE FAILURE RESULTED IN NO INJURIES ONLY A DELAY IN SURGERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10561301 ↗
Other Electrosurgical, Cutting & Coagulation & Accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page