LightWave IA-2000-SRecalls & reported problems
Conmed Corporation · LINVATEC SUCTION ABLATOR · Class 2
All models of LightWave (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 10845854040348), received between Oct 19, 2017 and Mar 20, 2026 — classified as 2 injuries, 8 malfunctions.
The most frequently reported specific device problem is “Melted” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Melted5
- Self-Activation or Keying2
- Burst Container or Vessel1
- Detachment Of Device Component1
- Human-Device Interface Problem1
- Inappropriate/Inadequate Shock/Stimulation1
- Material Fragmentation1
Patient problems
- No Known Impact Or Consequence To Patient4
- No Clinical Signs, Symptoms or Conditions2
- No Consequences Or Impact To Patient2
- Device Embedded In Tissue or Plaque1
- Partial thickness (Second Degree) Burn1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Self-Activation or Keying
THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE DEVICE, IA-2000-S, LINVATEC SUCTION ABLATOR, WAS USED DURING AN UNKNOWN PROCEDURE ON (B)(6) 2026 WHEN IT WAS REPORTED, ¿AFTER THE ELECTRODE IS CONNECTED TO THE HOST, IT WILL CONTINUE TO WORK CONTINUOUSLY WITHOUT THE NEED FOR PRESSING ANY BUTTONS.¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. FURTHER ASSESSMENT WAS REQUESTED AND ON (B)(6) 2026, THE SALES REPRESENTATIVE DISCOVERED THAT THE DEVICE HAD AUTO ACTIVATED WHEN FIRST PLUGGED IN FOR USE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24647656 ↗MalfunctionMar 13, 2026Self-Activation or Keying
THE DISTRIBUTOR REPORTED ON BEHALF OF THEIR CUSTOMER THAT THE DEVICE, IA-2000-S, LINVATEC SUCTION ABLATOR, WAS USED DURING A ROTATOR CUFF PROCEDURE ON (B)(6) 2026 WHEN IT WAS REPORTED, ¿THE BUTTON ACTIVATES WITHOUT BEING PRESSED (UPON CONTACT WITH WATER.)¿. THERE WAS NO REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED AS PLANNED USING AN ALTERNATE DEVICE WITH NO DELAY REPORTED. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24597098 ↗InjuryMay 15, 2019Material Fragmentation; Material Fragmentation
THE CONMED REPRESENTATIVE REPORTED ON BEHALF OF THE FACILITY THAT THE IA-2000-S, LIGHTWAVE SUCTION ABLATOR, BROKE DURING AN ELBOW ARTHROSCOPY. A PIECE OF PLASTIC DETACHED FROM THE ELECTRODE HEAD. THE INCIDENT OCCURRED AT FIRST USE OF THE DEVICE. THE SETTING WAS SET AT THE STANDARD 80. FURTHER INFORMATION WAS RECEIVED STATING THE PIECE WAS NOT REMOVED FROM THE PATIENT DUE TO ITS SMALL SIZE (2X1MM). IT COULD NOT BE RETRIEVED THROUGH THE SCOPE AND THE DOCTOR COULD NOT JUSTIFY A CONVERSION TO AN OPEN PROCEDURE. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. THERE WAS NO REPORTED DELAY THIS REPORT IS BEING RAISED BASED PATIENT INJURY DUE TO THE PIECE NOT BEING RETRIEVED.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 8613540 ↗InjuryNov 2, 2018Human-Device Interface Problem; Human-Device Interface Problem
THE CUSTOMER REPORTED THAT DURING A DIAGNOSTIC ARTHROSCOPY, PARTIAL SYNOVECTOMY IN THE GLENOHUMERAL JOINT THE IA-2000S, LIGHTWAVE SUCTION ABLATOR CAUSED A MINOR BURN. AFTER ADDITIONAL INFORMATION WAS RECEIVED, IT WAS DETERMINED THE PATIENT RECEIVED A 1ST TO 2ND DEGREE BURN ON THE LEFT UPPER ABDOMEN THAT HAD TO BE MEDICALLY TREATED WITH A PRESCRIPTION (FLAMMAZINE) OINTMENT. THE PROCEDURE WAS COMPLETED AS PLANNED WITH THE SAME DEVICE. THERE WAS NO SURGICAL DELAY REPORTED. THE PATIENT WAS RELEASED BUT WILL HAVE TO CONTINUE WITH DRESSING CHANGES AND APPLICATION OF OINTMENT BY THE FAMILY DOCTOR. THIS REPORT IS BEING RAISED ON THE BASIS OF A REPORTED PATIENT INJURY.
Patient problems: Partial thickness (Second Degree) Burn; Superficial (First Degree) Burn
MAUDE report 8032171 ↗MalfunctionApr 20, 2018Melted; Melted
A CUSTOMER REPORTED WHILE USING AN IA-2000-S, LIGHTWAVE SUCTION ABLATOR, IN AN ACL ARTHROSCOPY PROCEDURE, THE DEVICE MELTED. A 15-MINUTE SURGICAL DELAY OCCURRED AS THE SURGICAL SITE WAS EXPLORED TO ASSURE NO DEBRIS WAS LEFT BEHIND. A NEW CAUTERY TIP WAS USED TO COMPLETE THE PROCEDURE AS INTENDED. THIS INCIDENT DID NOT CAUSE HARM OR INJURY TO THE PATIENT. THIS REPORT IS BEING RAISED ON THE BASIS OF A REPORTED DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY WITH REOCCURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7447607 ↗MalfunctionJan 17, 2018Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
THE USER FACILITY REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE IA-2000-S LIGHTWAVE SUCTION ABLATOR WAS SHOCKING THE DOCTOR DURING USE. UPON GATHERING ADDITIONAL INFORMATION, FROM THE CONMED SALES REPRESENTATIVE AND THE USER FACILITY, THIS ALLEGED "SHOCK" DID NOT HARM THE DOCTOR OR THE PATIENT. THE PROCEDURE WAS COMPLETED WITH AN ALTERNATE DEVICE WITH MINIMAL SURGICAL DELAY. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7201862 ↗MalfunctionDec 20, 2017Burst Container or Vessel; Detachment Of Device Component; Melted; Burst Container or Vessel; Detachment Of Device Component; Melted
THE USER FACILITY REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE DISTAL TIP OF THE IA-2000-S LIGHTWAVE SUCTION ABLATOR MELTED DURING USE. THE TIP THEN BURST IN THE SHOULDER JOINT, ALL THE PIECES WERE REMOVED. THE PROCEDURE WAS COMPLETED WITH A SURGICAL DELAY OF 15 MINUTES. NO PATIENT INJURY REPORTED. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7135941 ↗MalfunctionDec 20, 2017Melted; Melted
THE USER FACILITY REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE TIP OF THE IA-2000-S LIGHTWAVE SUCTION ABLATOR MELTED. THE PROCEDURE WAS COMPLETED WITH NO SURGICAL DELAYS AND NO PATIENT INJURY WAS REPORTED. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7135927 ↗MalfunctionDec 20, 2017Melted; Melted
THE USER FACILITY REPORTED THAT DURING A SHOULDER ARTHROSCOPY, THE TIP OF THE IA-2000-S LIGHTWAVE SUCTION ABLATOR MELTED. THE PROCEDURE WAS COMPLETED WITH NO SURGICAL DELAYS AND NO PATIENT INJURY WAS REPORTED. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7135904 ↗MalfunctionOct 19, 2017Melted; Melted
A CONMED SALES REPRESENTATIVE REPORTED ON BEHALF OF A USER FACILITY THAT THE IA-2000-S ABLATORS' TIP MELTED DURING A ROTATOR CUFF REPAIR. THE PROCEDURE WAS COMPLETED AS PLANNED WITH NO SURGICAL DELAY REPORTED. THE PATIENT DID NOT SUFFER ANY CONSEQUENCES FROM THIS ALLEGED MALFUNCTION. THIS REPORT IS RAISED ON THE BASIS OF A REPORTED DEVICE MALFUNCTION WITH POTENTIAL FOR INJURY WITH RECURRENCE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6961715 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page