PlumePen PLP1020 (UDI 10845854061268)Recalls & reported problems
Conmed Corporation · PlumePen Elite Surgical Smoke Evacuation Pencil, 7/8" (22 mm), 10 ft. (3 m) · Class 2
All models of PlumePen (8)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10845854061268), received between Aug 2, 2023 and Jun 4, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Mechanical Jam1
- Self-Activation or Keying1
- Unintended Electrical Shock1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Burn(s)1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 4, 2026Unintended Electrical Shock
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLP1020 ELITE SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS BEING USED ON (B)(6) 2026 DURING A CIRCUMCISION PROCEDURE WHEN IT WAS REPORTED ¿DR. HAD BEEN USING A BOVIE FOR APPROXIMATELY FOR 1 HOUR AND THEN SHE FELT A ZAP AND QUICKLY PUT DOWN THE BOVIE. HER HAND FINGER WAS HURTING AND WHEN SHE REMOVED HER GLOVE, SHE NOTICED SMALL BURN MARK AND A REDDENED AREA.¿. FURTHER ASSESSMENT FOUND THAT SURGEON WAS NOT DIAGNOSED WITH BURN SEVERITY (NO PATIENT IMPACT ONLY SURGEON). BURN LOCATED ON RIGHT INDEX FINGER. NOT TREATMENT REPORTED - MENTIONED REDNESS AND SORE. PATIENT STATUS - UNAFFECTED (SURGEON EXPERIENCED BURN). THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: Burn(s)
MAUDE report 25392236 ↗MalfunctionSep 5, 2024Mechanical Jam; Mechanical Jam
A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PLP1020, 20/CA PLUMEPEN ELITE (S) DEVICE WAS BEING USED ON (B)(6) 2024, AND ¿BUTTON IS STUCK IN "ON" POSITION, WON'T TURN OFF.¿. FOLLOW-UP ASSESSMENT WAS ATTEMPTED BUT UPON RESPONSE, THE CUSTOMER PROVIDED NO FURTHER INFORMATION; THEREFORE, IT COULD NOT BE DETERMINED IF THIS WAS AN EVENT OF SELF-ACTIVATION AND WHETHER IT OCCURRED INTRA-OPERATIVELY OR DURING PRE-OPERATIVE TESTING. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR HOSPITALIZATION FOR THE USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20147670 ↗MalfunctionAug 2, 2023Self-Activation or Keying; Self-Activation or Keying
THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLP1020, 20/CA PLUMEPEN ELITE (S), WAS BEING USED DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE WHEN IT WAS REPORTED, ¿THE PLUME PEN ELITE HAND PIECES MUST HAVE AN ISSUE WITH THE CUT BUTTON AS THE STAFF SAY THAT THEY ARE ACTIVATING THE DIATHERMY AS SOON AS THEY ARE CONNECTED.¿. THERE WAS NO INITIAL REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED AS PLANNED. FURTHER QUESTIONING WAS SENT TO THE REPORTER TO CLARIFY IF SELF-ACTIVATION IS BEING REPORTED; HOWEVER, WE HAVE NOT RECEIVED AN ANSWER TO DATE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17441350 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page