Skip to content

PlumePen PLP1020 (UDI 10845854061268)Recalls & reported problems

Conmed Corporation · PlumePen Elite Surgical Smoke Evacuation Pencil, 7/8" (22 mm), 10 ft. (3 m) · Class 2

All models of PlumePen (8)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10845854061268), received between Aug 2, 2023 and Jun 4, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Mechanical Jam” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Mechanical Jam1
  • Self-Activation or Keying1
  • Unintended Electrical Shock1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Burn(s)1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 4, 2026Unintended Electrical Shock

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLP1020 ELITE SURGICAL SMOKE EVACUATION PENCIL DEVICE WAS BEING USED ON (B)(6) 2026 DURING A CIRCUMCISION PROCEDURE WHEN IT WAS REPORTED ¿DR. HAD BEEN USING A BOVIE FOR APPROXIMATELY FOR 1 HOUR AND THEN SHE FELT A ZAP AND QUICKLY PUT DOWN THE BOVIE. HER HAND FINGER WAS HURTING AND WHEN SHE REMOVED HER GLOVE, SHE NOTICED SMALL BURN MARK AND A REDDENED AREA.¿. FURTHER ASSESSMENT FOUND THAT SURGEON WAS NOT DIAGNOSED WITH BURN SEVERITY (NO PATIENT IMPACT ONLY SURGEON). BURN LOCATED ON RIGHT INDEX FINGER. NOT TREATMENT REPORTED - MENTIONED REDNESS AND SORE. PATIENT STATUS - UNAFFECTED (SURGEON EXPERIENCED BURN). THERE WAS NO REPORT OF MEDICAL INTERVENTION, OR EXTENDED HOSPITALIZATION FOR THE PATIENT OR USER. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: Burn(s)

    MAUDE report 25392236 ↗
  • MalfunctionSep 5, 2024Mechanical Jam; Mechanical Jam

    A SALES REPRESENTATIVE REPORTED ON BEHALF OF A CUSTOMER THAT THE PLP1020, 20/CA PLUMEPEN ELITE (S) DEVICE WAS BEING USED ON (B)(6) 2024, AND ¿BUTTON IS STUCK IN "ON" POSITION, WON'T TURN OFF.¿. FOLLOW-UP ASSESSMENT WAS ATTEMPTED BUT UPON RESPONSE, THE CUSTOMER PROVIDED NO FURTHER INFORMATION; THEREFORE, IT COULD NOT BE DETERMINED IF THIS WAS AN EVENT OF SELF-ACTIVATION AND WHETHER IT OCCURRED INTRA-OPERATIVELY OR DURING PRE-OPERATIVE TESTING. THERE WAS NO REPORT OF INJURY, MEDICAL/SURGICAL INTERVENTION, OR HOSPITALIZATION FOR THE USER. THIS REPORT IS BEING RAISED ON THE BASIS OF MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20147670 ↗
  • MalfunctionAug 2, 2023Self-Activation or Keying; Self-Activation or Keying

    THE SALES REPRESENTATIVE REPORTED ON BEHALF OF THE CUSTOMER THAT THE PLP1020, 20/CA PLUMEPEN ELITE (S), WAS BEING USED DURING AN UNKNOWN PROCEDURE ON AN UNKNOWN DATE WHEN IT WAS REPORTED, ¿THE PLUME PEN ELITE HAND PIECES MUST HAVE AN ISSUE WITH THE CUT BUTTON AS THE STAFF SAY THAT THEY ARE ACTIVATING THE DIATHERMY AS SOON AS THEY ARE CONNECTED.¿. THERE WAS NO INITIAL REPORT OF INJURY, MEDICAL INTERVENTION, OR HOSPITALIZATION FOR THE PATIENT OR USER. THE PROCEDURE WAS COMPLETED AS PLANNED. FURTHER QUESTIONING WAS SENT TO THE REPORTER TO CLARIFY IF SELF-ACTIVATION IS BEING REPORTED; HOWEVER, WE HAVE NOT RECEIVED AN ANSWER TO DATE. THIS REPORT IS BEING RAISED DUE TO THE REPORTED MALFUNCTION WITH POTENTIAL FOR INJURY UPON REOCCURRENCE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17441350 ↗

Other Electrosurgical, cutting & coagulation & accessories devices

All Electrosurgical, cutting & coagulation & accessories devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page