Force Triad ForceTriad ZERecalls & reported problems
Covidien LP · Energy Platform · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10884521133310), received between Feb 8, 2018 and Feb 3, 2026 — classified as 1 injury, 12 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem7
- Device Displays Incorrect Message3
- Energy Output Problem2
- Self-Activation or Keying2
- Connection Problem1
- Defective Alarm1
- Failure to Align1
- Inaccurate Delivery1
Patient problems
- Burn(s)4
- No Clinical Signs, Symptoms or Conditions3
- No Patient Involvement2
- Disseminated Intravascular Coagulation (DIC)1
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
- No Code Available1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 3, 2026Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER DURING PROCEDURE, THERE WERE BURNS WHILE USING REM (RETURN ELECTRODE MONITORING) PADS. CUT SETTINGS WERE 25/25 CUT AND COAG.
Patient problems: Burn(s)
MAUDE report 24244270 ↗MalfunctionMar 8, 2022Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, THE UNIT REACTIVE ENDPOINT DID NOT REACH SEAL. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13688540 ↗MalfunctionSep 21, 2021Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem; Device Displays Incorrect Message; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY OF A VETERINARY CASE, THE UNIT BEGAN TO BEEP LOUDLY AND THE SCREEN STARTED FLASHING WITH A MESSAGE THAT SAID INSTRUMENT COMBINATION NOT ALLOWED AND THE E275 SUPERVISOR ERROR. A BLEEDING OCCURRED BUT NO TRANSFUSION PERFORMED. THE PROCEDURE WAS COMPLETED USING A GOOD OLD SCALPEL BLADE AND SUTURE MATERIAL TO CUT THROUGH TISSUE AND LIGATE VESSELS APPROPRIATELY. THERE WAS NO PATIENT INJURY.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 12505611 ↗MalfunctionSep 12, 2021Self-Activation or Keying; Device Displays Incorrect Message; Self-Activation or Keying; Device Displays Incorrect Message
ACCORDING TO THE REPORTER, DURING USE, THE UNIT DISPLAYED MULTIPLE ERROR CODES WITHIN A COUPLE MINUTES OF EACH OTHER AND WAS ACTIVATING ON ITS OWN. THE BIOMEDICAL TECHNICIAN UNABLE TO DUPLICATE THE REPORTED ISSUES HOWEVER IT WAS ABLE TO CONFIRM ERROR CODES E266-E300-E260-E275-E302 THROUGH THE ERROR LOGS. NO FURTHER INFORMATION AVAILABLE.
Patient problems: Insufficient Information
MAUDE report 12457772 ↗MalfunctionJul 7, 2021Failure to Align; Adverse Event Without Identified Device or Use Problem; Failure to Align; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING USE, THE UNIT HAD A REM/HANDPIECE ISSUE. THE PROCURE WAS DELAYED AND THE PATIENT HAD A BURN.
Patient problems: Burn(s)
MAUDE report 12127297 ↗MalfunctionFeb 15, 2021Inaccurate Delivery; Inaccurate Delivery
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT'S PATIENT RETURN ELECTRODES (REM) WOULD NOT TURN GREEN. IT STAYED RED WHEN THE HAND PIECE WAS ACTIVATED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11327594 ↗MalfunctionJan 5, 2021Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC APPENDECTOMY, THE DEVICE CAUSED BURN.
Patient problems: Burn(s)
MAUDE report 11120966 ↗MalfunctionJul 28, 2020Energy Output Problem; Device Displays Incorrect Message; Connection Problem; Energy Output Problem; Device Displays Incorrect Message; Connection Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE INSTRUMENT IN LIGASURE PORT 1 WAS NOT SEALING AND THE UNIT ALSO HAD AN ERROR. ANOTHER UNIT WAS USED AND WORKED FINE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 10331511 ↗MalfunctionDec 14, 2018Self-Activation or Keying; Self-Activation or Keying
ACCORDING TO THE REPORTER, THE GENERATOR WAS RESETTING CUT AND COAGULATION TO ONE ON ITS OWN.
Patient problems: No Information
MAUDE report 8164292 ↗InjuryDec 11, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE LITERATURE SOURCE STUDY, PRESENTING THE EXPERIENCE WITH SURGICAL MANAGEMENT OF PLACENTA PRAEVIA PERCRETA. A TOTAL OF 58 PATIENTS UNDERWENT ULTRASOUND MAPPING OF THE PLACENTAL AREA BEFORE SURGERY. CESAREAN SECTION DELIVERY WAS PERFORMED FOR PERCRETA PATIENTS OVER THE 34TH WEEK OF GESTATION IF NO OTHER EMERGENT SITUATION (BLEEDING, UTERINE CONTRACTION, MEMBRANE RUPTURE) WAS PRESENT. FOUR PATIENTS (6.9 %) HAD BLADDER INJURY AS A RESULT OF DENSE ADHESIONS BETWEEN THE POSTERIOR WALL OF THE BLADDER AND UTERUS DURING DISSECTION. ALL THESE INJURY CASES ARE MANAGED TOGETHER WITH EXPERT UROLOGIST. ONE PATIENT (1.7 %) REQUIRED CYSTOTOMY, AND ONE PATIENT (1.7 %) REQUIRED RE-OPERATION BECAUSE OF POSTOPERATIVE HEMORRHAGE. FOUR PATIENTS (6.9%) HAD DISSEMINATED INTRAVASCULAR COAGULATION (DIC), TWO PATIENTS (3.4%) HAD PULMONARY EMBOLISM AND FOUR PATIENTS (6.9%) WITH WOUND INFECTION.
Patient problems: Disseminated Intravascular Coagulation (DIC); No Code Available; Post Operative Wound Infection; Pulmonary Embolism
MAUDE report 8150400 ↗MalfunctionOct 2, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE PATIENT GOT BURN INJURIES WHILE DOING BREAST OPERATION.
Patient problems: Burn(s)
MAUDE report 7925663 ↗MalfunctionMar 28, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE DEVICE DID NOT GIVE A PROPER SEAL. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7378704 ↗MalfunctionFeb 8, 2018Defective Alarm; Material Deformation; Defective Alarm; Material Deformation
ACCORDING TO THE REPORTER, EXTERNAL INSPECTION FOUND MINOR SCRATCHES ON THE COVER. IT WAS ALSO NOTED THAT WHEN THE ADAPTER WAS INSERTED AND REMOVED, OCCASIONALLY THE REM INDICATOR WOULD QUICKLY FLASH FROM RED TO GREEN BACK TO RED. THIS SHOULD NOT OCCUR; THE REM INDICATOR SHOULD REMAIN RED UNTIL THE CORRECT IMPEDANCE IS SEEN. TRACED ISSUE DOWN TO A FAULT IN THE REM CIRCUITRY ON THE RF BOARD. UNIT WILL NEED THE RF BOARD REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 7252636 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.