Skip to content

Edge E1450XNSBRecalls & reported problems

Covidien LP · Hex-locking Blade Electrode;Caution: For manufacturing, processing, or repacking. · Class 2

All models of Edge (43)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 10884521175426), received between Mar 24, 2021 and Oct 27, 2023 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Detachment of Device or Device Component1
  • Peeled/Delaminated1

Patient problems

  • Insufficient Information2
  • Burn(s)1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A BELOW THE KNEE AMPUTATION, THE TEAM WAS USING THE CAUTERY TO OBTAIN HEMOSTASIS AS PER USUAL. THERE WAS A NON-MEDTRONIC GAUZE SPONGE AT THE WOUND BUT NOT WHERE THEY WERE CAUTERIZING (MAYBE 3CM AWAY) AND THEY NOTICED THAT IT TURNED COLOR AND THEN A SMALL FLAME APPEARED. THE SPONGE WAS IMMEDIATELY REMOVED FROM THE FIELD. THERE WAS NO OBVIOUS FLAMMABLE SUBSTANCE ON THE FIELD, THE PREP WAS DONE AT THE BEGINNING OF THE CASE, AND THE PATIENT WAS INTUBATED, SO THERE WAS NO OXYGEN SOURCE ON THE FIELD. THERE WAS NO FIRE ON THE FIELD, WOUND, AND DRAPES AND THERE WAS NO SOOT OR CHAR. THE BOVIE PEN, GROUNDING PAD, AND MACHINE WERE ALL REMOVED FROM THE CASE. THE SPONGE WAS SECURED AS WELL. THEY WERE ABLE TO COMPLETE THE PROCEDURE SAFELY. THERE WAS NO PATIENT/STAFF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18023420 ↗
  • InjuryJul 10, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A TONSILLECTOMY PROCEDURE, IT DID NOT GO ALL THE WAY DOWN INTO THE PENCIL AND A PATIENT WAS BURNED ON THE LIP BECAUSE OF THE EXPOSED METAL AT THE BASE. FOLLOW UP CARE WAS PROVIDED TO THE PATIENT AND PATIENT WAS READMITTED TO THE HOSPITAL ON TWO OCCASIONS AFTER THE PROCEDURE. PATIENT CURRENTLY DOING FINE.

    Patient problems: Burn(s)

    MAUDE report 17295295 ↗
  • MalfunctionApr 12, 2021Peeled/Delaminated; Peeled/Delaminated

    ACCORDING TO THE REPORTER, THE DEVICE TIP WAS DAMAGED. A PHOTO WAS ATTACHED SHOWING THE DEVICE INSULATION WAS DAMAGED.

    Patient problems: Insufficient Information

    MAUDE report 11654210 ↗
  • MalfunctionMar 24, 2021Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE TEFLON BOVIE WAS COMING OFF.

    Patient problems: Insufficient Information

    MAUDE report 11562248 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates