Barrx TTS-1100 (UDI 10884521176621)Recalls & reported problems
Covidien LP · Channel RFA Endoscopic Catheter,Esophageal TTS, 120 Applications · Class 2
All models of Barrx (19)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10884521176621), received between Mar 14, 2019 and May 12, 2026 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Detachment of Device or Device Component3
- Adverse Event Without Identified Device or Use Problem1
- Connection Problem1
- Insufficient Device Problem Information1
- Material Twisted/Bent1
Patient problems
- Foreign Body In Patient2
- Arrhythmia1
- Dysphagia/ Odynophagia1
- No Clinical Signs, Symptoms or Conditions1
- No Consequences Or Impact To Patient1
- Sore Throat1
- Unspecified Tissue Injury1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryMay 12, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT WHILE PERFORMING RADIOFREQUENCY ABLATION AT 48 WATTS AND 12 JOULES, AN EPISODE OF SINUS ARRHYTHMIA WAS NOTED. IMMEDIATE STABILIZATION PROTOCOLS WERE FOLLOWED TO WHICH THE ANESTHESIOLOGIST ADMINISTERED MAGNESIUM, LIDOCAINE, ESMOLOL, AND BUTYL HYOSCINE, ALLOWING FOR THE COMPLETION OF THE PROCEDURE. POST-PROCEDURAL EVALUATION CONFIRMED THE PATIENT WAS STABLE. THERE WERE NO LOOSE COMPONENTS.
Patient problems: Arrhythmia
MAUDE report 25149989 ↗InjuryApr 18, 2025Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
IT WAS REPORTED THAT UPON REMOVAL, THE TIP OF A DEVICE BROKE OFF IN THE ESOPHAGUS. THE BROKEN TIP WAS RETRIEVED VIA BIOPSY FORCEPS. THE SAME DEVICE WAS USED TO COMPLETE THE ABLATION.
Patient problems: Foreign Body In Patient
MAUDE report 21866301 ↗InjuryDec 29, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, A PORTION OF THE ABLATION CATHETER WAS STUCK IN THE PATIENT'S MUCOSA AND AS IT WAS MANIPULATED, THE CATHETER TORE. A SNARE WAS USED TO RETRIEVE THE STUCK ELECTRODE. THE PATIENT WAS FINE.
Patient problems: Foreign Body In Patient
MAUDE report 18422609 ↗InjuryOct 20, 2023Break; Detachment of Device or Device Component; Break; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, THE CATHETER BROKE OFF INSIDE THE PATIENT. THE CUSTOMER WAS ABLE TO RETRIEVE THE CATHETER FROM THE PATIENT.
Patient problems: Unspecified Tissue Injury
MAUDE report 17981042 ↗MalfunctionJun 28, 2023Material Twisted/Bent
CHANNEL RFA (RADIO FREQUENCY ABLATION) CATHETER WOULD NOT GO THROUGH THE SCOPE WORKING CHANNEL. THE TYPE OF CATHETER USED WAS THE BARRX CHANNEL RFA ENDOSCOPIC CATHETER. NEVER HEARD OF AN ISSUE WITH THIS PARTICULAR INSTRUMENT BEFORE THIS OCCURRENCE. DOCTOR AND THE SURGICAL TECHNOLOGIST IN THE ROOM TRIED MULTIPLE ROUTES OF GETTING DOWN THE WORKING CHANNEL, AND WHEN IT WAS EVENTUALLY INTRODUCED INTO THE ESOPHAGUS, IT WAS BENT AND NOT WORKING PROPERLY. THE TISSUE ABLATIONS WERE NOT 100% BECAUSE THE TISSUE WAS NOT GETTING THE FULL CONTACT OF THE PADDLE. THE FAULTY CATHETER WAS TAKEN OUT OF THE PATIENT AND A NEW ONE WAS USED AND HAD NO DIFFICULTIES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17216046 ↗MalfunctionSep 24, 2019Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, DURING THE EGD (ESOPHAGOGASTRODUODENOSCOPY) OF THE BARRETT'S PROCEDURE, THE CABLE WOULD NOT CONNECT TO ANY OF THE CATHETERS. THE CUSTOMER OPENED 4 PACKAGES. THE PATIENT WAS PREPPED FOR THE PROCEDURE AND WAS UNDER ANESTHESIA. THE PROCEDURE WAS ABORTED AND A REPEAT PROCEDURE WAS NECESSARY. THERE WAS NO PATIENT AND USER HARM.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9112619 ↗InjuryMar 14, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
ACCORDING TO THE REPORTER, A WEEK POST INLET PATCH RADIO FREQUENCY ABLATION PROCEDURE, THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO THROAT PAIN AND DYSPHAGIA. THE PATIENT WAS GIVEN PAIN REVILER BOTH ORALLY AND INTRAVENOUSLY AND PLACED ON LIQUID DIET TO REGULATE THE PAIN AND DYSPHAGIA. THE CUSTOMER NOTED THAT THERE WERE NO ERROR CODES FLAGGED DURING THE PROCEDURE. THE PHYSICIAN WAS NOT ATTRIBUTING THE HOSPITALIZATION TO THE DEVICE AND WAS CONFIDENT THAT IT WAS DUE TO THE PROCEDURE. THE PATIENT'S THROAT PAIN AND DYSPHAGIA WERE RESOLVED, AND THE PATIENT WAS DISCHARGED.
Patient problems: Dysphagia/ Odynophagia; Sore Throat
MAUDE report 8422947 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page