LigaSure LF3225 (UDI 10884521197183)Recalls & reported problems
Covidien LP · Curved Jaw, Open Sealer/Divider Electrode · Class 2
All models of LigaSure (32)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 20
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 20 adverse event reports that list this exact device (DI 10884521197183), received between May 9, 2018 and Apr 6, 2026 — classified as 20 malfunctions.
The most frequently reported specific device problem is “Retraction Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem6
- Retraction Problem5
- Detachment of Device or Device Component4
- Difficult to Open or Close2
- Protective Measures Problem2
- Connection Problem1
- Energy Output Problem1
- Flare or Flash1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Superficial (First Degree) Burn5
- No Consequences Or Impact To Patient4
- Insufficient Information3
- Burn(s)2
Recent adverse event reports
The 20 most recent of 20 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 6, 2026Retraction Problem
IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE'S KNIFE BLADE JAMMED AND COULD NOT RETRACT.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 24790374 ↗MalfunctionApr 6, 2026Retraction Problem
IT WAS REPORTED THAT DURING A VAGINAL HYSTERECTOMY PROCEDURE, THE DEVICE'S KNIFE BLADE JAMMED AND COULD NOT RETRACT. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 24790243 ↗MalfunctionApr 21, 2025Detachment of Device or Device Component; Detachment of Device or Device Component
IT WAS REPORTED THAT WHEN SETTING UP THE DEVICE, THE DISPOSABLE WAS ATTACHED TO THE REUSABLE CLAMP IN THE PRESCRIBED MANNER, AND NO ISSUES OR ABNORMALITIES WERE NOTICED. HOWEVER, DURING A DEBULKING SURGERY, THE OPERATING TEAM NOTICED A GREY PIECE OF PLASTIC IN THE PATIENT¿S ABDOMEN. WHEN THE STAFF TOOK IT OUT OF THE PATIENT, INVESTIGATED AND FOUND OUT IT WAS ONE OF THE PINS OF THE DEVICE¿S INLAY. THE SURGEON DECIDED TO FINISH THE SURGERY WITH THE SAME DEVICE, SINCE THE PROCEDURE WERE CLOSE TO FINISH, AND IT DID SEEM TO STILL WORK PROPERLY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21873834 ↗MalfunctionOct 17, 2024Difficult to Open or Close; Protective Measures Problem; Difficult to Open or Close; Protective Measures Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT WAS UNABLE TO MOUNT/CLIP THE ELECTRODE TO SEAL THE OPEN DIVIDER ON THE LIGASURE CLAMP; THE BLADE WAS MORE OUT THAN USUAL. THE DEVICE WAS CONNECTED TO THE GENERATOR, AND THE PROCEDURE WAS COMPLETED WITH THE USE OF ANOTHER LIGASURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20470345 ↗MalfunctionJul 3, 2023Connection Problem; Detachment of Device or Device Component; Connection Problem; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, PRIOR TO USE THE END OF THE DEVICE BROKE OFF ON APPLICATION AND WOULD NOT CONNECT TO THE HAND PIECE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17246453 ↗MalfunctionJun 16, 2023Protective Measures Problem; Protective Measures Problem
ACCORDING TO THE REPORTER, PRIOR TO A C SECTION, THE BLADE ON THE LEFT JAW OF LIGASURE WAS EXPOSED. UPON OPENING, IT APPEARED THAT WHOLE LEFT JAW WAS NOT PROPERLY CONNECTED TO LIGASURE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17142730 ↗MalfunctionFeb 22, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE LIGASURE DISPOSABLE UNIT HAD A BLUE PLASTIC PIECE TO SECURE THE CABLE. WHEN THE SURGEON BROKE THIS OPEN ALONG THE PERFORATIONS, A VERY SMALL PIECE OF PLASTIC BROKE OFF INTO THE ABDOMEN. PHYSICIAN WAS ABLE TO FIND SMALL PIECE AND REMOVE FROM PATIENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13576900 ↗MalfunctionFeb 15, 2021Difficult to Open or Close; Separation Problem; Difficult to Open or Close; Separation Problem
DURING THE PROCEDURE, AT THE BEGINNING OF THE PROCEDURE, THE JAWS OF THE DISSECTOR WERE PARTICULARLY ATTACHING TO THE TISSUE (TISSUE STICKING TO THE JAWS). AS A CONSEQUENCE, IT WAS DIFFICULT TO REOPEN THE JAWS AND CONTINUE WITH THE PROCEDURE. THERE WAS NO CONSEQUENCE FOR THE PATIENT, BUT IT INCREASED THE OPERATING TIME. ANOTHER DISSECTOR WAS USED WITHOUT ANY PROBLEM ENCOUNTERED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11327318 ↗MalfunctionJan 21, 2021Retraction Problem; Retraction Problem
ACCORDING TO THE REPORTER, IT WAS IMPOSSIBLE TO RETRACT THE BLADE PREVENTING THE BUTTON FROM BEING USED.
Patient problems: Insufficient Information
MAUDE report 11207813 ↗MalfunctionOct 31, 2020Detachment of Device or Device Component; Detachment of Device or Device Component
ACCORDING TO THE REPORTER, DURING A TOTAL ABDOMINAL HYSTERECTOMY, ADHESIOLYSIS, CONSERVE OVARIES, PART OF THE DEVICE FELL OFF (NOT INTO THE PATIENT). THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10767210 ↗MalfunctionSep 15, 2020Retraction Problem; Retraction Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE KNIFE BLADE DID NOT RETRACT. THE SURGEON OPENED A NEW HAND PIECE TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10538768 ↗MalfunctionAug 28, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A GYNECOLOGICAL PROCEDURE, THE PATIENT HAD A FIRST-DEGREE BURN OUTSIDE THE VAGINA IN THE (B)(6) WITH THE DEVICE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 10469060 ↗MalfunctionAug 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A GYNECOLOGICAL PROCEDURE, THE PATIENT HAD A FIRST DEGREE BURN OUTSIDE THE VAGINA IN THE LABIA MAJORA WITH THE DEVICE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 10452802 ↗MalfunctionAug 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A GYNECOLOGICAL PROCEDURE, THE PATIENT HAD A FIRST-DEGREE BURN OUTSIDE THE VAGINA IN THE LABIA MAJORA WITH THE DEVICE.
Patient problems: Superficial (First Degree) Burn
MAUDE report 10452795 ↗MalfunctionJul 23, 2020Retraction Problem; Retraction Problem
ACCORDING TO THE REPORTER, THE DEVICE'S KNIFE BLADE DID NOT RETRACT. IT WAS DISCOVERED THAT THE LEVER THAT OPERATES THE BLADE WAS BLOCKED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10315910 ↗MalfunctionJan 6, 2020Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, DURING A HEMORRHOIDECTOMY, THE DEVICE HAD AN INCOMPLETE SEAL, THEY ADJUSTED IT AND RE-TRIED IT THREE TIMES, THEN THEY ABANDONED THE DEVICE AND USED A CAUTERY DEVICE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9552666 ↗MalfunctionNov 20, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, POST OPERATIVELY, THE PATIENT HAD A FIRST DEGREE VULVAR BURN.
Patient problems: Burn(s)
MAUDE report 8091896 ↗MalfunctionNov 15, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, POST OPERATIVELY, THE PATIENT HAD A FIRST DEGREE VULVAR BURN.
Patient problems: Burn(s)
MAUDE report 8076046 ↗MalfunctionNov 9, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, POST OPERATIVELY, THE PATIENT GOT A FIRST DEGREE VULVAR BURN.
Patient problems: Superficial (First Degree) Burn
MAUDE report 8054662 ↗MalfunctionMay 9, 2018Flare or Flash; Flare or Flash
ACCORDING TO THE REPORTER, POST-OPERATIVELY, THE PATIENT EXPERIENCE FIRST DEGREE BURN (REDNESS) AT THE VULVAR AREA.
Patient problems: Superficial (First Degree) Burn
MAUDE report 7499039 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page