12
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521550803), received between May 22, 2020 and Oct 27, 2025 — classified as 1 injury, 11 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 7 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520267
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component9
  • Adverse Event Without Identified Device or Use Problem1
  • Arcing1
  • Entrapment of Device1
  • Failure to Deliver Energy1
  • Intermittent Energy Output1
  • Sparking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • Burn, Thermal1
  • Insufficient Information1
  • No Information1
  • Pain1
  • Tissue Breakdown1
  • Unintended Radiation Exposure1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 27, 2025Failure to Deliver Energy; Entrapment of Device; Intermittent Energy Output; Failure to Deliver Energy; Entrapment of Device; Intermittent Energy Output

    IT WAS REPORTED THAT DURING PROCEDURE, THE DEVICE STOPPED WORKING. IT DID NOT BURN/WORK PROPERLY. THE TISSUE NEEDS TO BE TORN TO GET IT OFF. IT WAS NOT EASILY REMOVED.

    Patient problems: Tissue Breakdown

    MAUDE report 23397337 ↗
  • MalfunctionDec 26, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING THE PROCEDURE, THE SURGEON DID A FINAL LOOK INTO THE ABDOMEN AND IDENTIFIED A BLUE CAP IN THE PATIENT. THE SURGICAL TECHNICIAN (ST) REMOVED 2 CAPS (1 FROM EACH END) AND DISCARDED IT PRIOR TO CASE START.

    Patient problems: Insufficient Information

    MAUDE report 21018467 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297486 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297454 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297425 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297415 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297393 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297373 ↗
  • MalfunctionSep 24, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, THE ELECTRODE'S TEFLON FELL OFF OF L-HOOK. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20297294 ↗
  • InjuryFeb 22, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, DURING DIAGNOSTIC LAPAROSCOPY, THE SURGICAL TECH DID NOT REMOVE THE PLASTIC COVER, IT ENDED UP IN PATIENT AND WAS RECOVERED 18 MONTHS LATER AND IT WAS FULLY RETRIEVED. AN X-RAY WAS REQUIRED TO ASSIST IN LOCATING THE PIECE, THERE WAS REOPERATION TO REMOVE THE PLASTIC COVER, INVESTIGATION REVEALED THE RETAINED FOREIGN OBJECT (RFO) WAS LIKELY THE COVER TO AN L-HOOK ELECTRODE USED IN THE APPENDECTOMY CASE. PATIENT COMPLAINED OF PAIN FROM FOREIGN PROJECT.

    Patient problems: Pain; Unintended Radiation Exposure

    MAUDE report 16419192 ↗
  • MalfunctionJul 8, 2020Arcing; Adverse Event Without Identified Device or Use Problem; Arcing; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC SURGERY, THE DEVICE SPARKED AND CAUSED A SMALL BURN UNDER THE TROCAR.

    Patient problems: No Information

    MAUDE report 10246842 ↗
  • MalfunctionMay 22, 2020Sparking

    PATIENT WAS HAVING SURGERY LAPAROSCOPICALLY WITH A BOVIE AND L HOOK ELECTRODE. THE L-HOOK ELECTRODE SPARKED AND CAUSED A SMALL BURN UNDER THE TROCAR. FDA SAFETY REPORT ID# (B)(4).

    Patient problems: Burn, Thermal

    MAUDE report 10086681 ↗