Valleylab FT10 VLFT10GENZDRecalls & reported problems
Covidien LP · FT Series Energy Platform · Class 2
All models of Valleylab FT10 (3)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 10884521640474), received between Aug 6, 2018 and Oct 19, 2025 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Energy Output Problem” (8 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Nov 16, 2023Open, Classified
Due to software issue, New systems may exhibit an error which may prohibit use of new systems resulting in system being inoperable
Z-0631-2024 · Root cause: Software Manufacturing/Software Deployment · Recall details · FDA record ↗
Most reported problems
Device problems
- Energy Output Problem8
- Connection Problem4
- Device Displays Incorrect Message3
- Self-Activation or Keying3
- Adverse Event Without Identified Device or Use Problem2
- Activation Problem1
- Failure to Align1
- Failure to Deliver Energy1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Patient Involvement2
- Unspecified Tissue Injury2
- Burn(s)1
- Hemorrhage/Blood Loss/Bleeding1
- Insufficient Information1
- No Consequences Or Impact To Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2025Energy Output Problem; Device Displays Incorrect Message; Energy Output Problem; Device Displays Incorrect Message
IT WAS REPORTED THAT DURING TOTAL LAPAROSCOPIC HYSTERECTOMY, THE DEVICE HAD ALARM ISSUE. IT SHOWED ERROR CODE 404 ENERGY ACTIVATION / SEALING ISSUE ON THE GENERATOR. THE DEVICE HAD INADEQUATE SEALING. THERE WAS EVIDENCE OF ENERGY DELIVERY BUT BLEEDS AFTER CUTTING AND END TONES WAS HEARD. THE JAWS WERE NOT CLEANED BECAUSE IT HAPPENED IMMEDIATELY OUT OF STERILE PACKAGING. THERE WAS ZERO GOOD SEALS. THE TISSUES THICKNESS WAS NOT VERY COMPRESSED TO LESS THAN 1.5MM OF FATTY TISSUE / OMENTUM. THE DEVICE WAS CONNECTED TO A GENERATOR WITH SOFTWARE VERSION 5.0.0.4025. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR'S DEVICE. THE ISSUE OCCURRED DURING LYSIS OF ADHESIONS ON PERITONEUM AT THE ONSET OF THE PROCEDURE. THE PATIENT HAD A BLEEDING OF 2ML THAT DID NOT REQUIRE A BLOOD TRANSFUSION.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 23333799 ↗MalfunctionOct 19, 2025Energy Output Problem; Energy Output Problem
IT WAS REPORTED THAT DURING A TOTAL LAPAROSCOPIC HYSTERECTOMY, THE GENERATOR WAS DELIVERING THE IMPROPER AMOUNT OF ENERGY WITH A SEAL-CYCLE COMPLETE TONE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23333656 ↗MalfunctionSep 26, 2024Failure to Deliver Energy; Adverse Event Without Identified Device or Use Problem; Failure to Deliver Energy; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, THE UNIT HAD A BURN ISSUE.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 20311894 ↗MalfunctionSep 11, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, PATIENT HAD URETER INJURY. FOUND TWO DAYS AFTER TOTAL LAP HYSTERECTOMY CASE COMPLETED. NO ISSUES WERE IDENTIFIED DURING PROCEDURE. PATIENT HAD NEPHROSTOMY TUBE PLACEMENT AND DELAYED URETERAL RE-IMPLANTATION. INJURY WAS THREE CM ABOVE THE URETEROVESICAL JUNCTION. UROLOGIST WAS UNABLE TO PASS A STENT AND RETROGRADE PYELOGRAM SHOWED EXTRAVASATION OF DYE AT THAT LEVEL. NO EVIDENCE OF SUTURE LIGATION. UNABLE TO ELABORATE ON ETIOLOGY OF INJURY.
Patient problems: Unspecified Tissue Injury
MAUDE report 17726297 ↗MalfunctionJun 19, 2023Energy Output Problem; Connection Problem; Energy Output Problem; Connection Problem
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC SIGMOIDECTOMY, IT WAS NOT A NOTICEABLE DEFECT, BUT IT SEEMS LIKE THERE WERE SEVERAL OCCASIONS WHEN THE SEAL OUTPUT FELT WEAK. THERE WAS NO REGRASP ALERT. END TONE WAS HEARD AND THERE WAS EVIDENCE OF ENERGY DELIVERY. THE OPERATION WAS COMPLETED SUCCESSFULLY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17158721 ↗MalfunctionApr 15, 2022Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, THE UNIT WAS NOT SEALING.
Patient problems: Insufficient Information
MAUDE report 14118106 ↗MalfunctionOct 20, 2021Self-Activation or Keying; Smoking; Inaccurate Delivery; Connection Problem; Self-Activation or Keying; Smoking; Inaccurate Delivery; Connection Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE HANDPIECE ACTIVATED BY ITSELF, THERE WAS NO REM ALARM AND REM RECEPTACLE STARTED SMOKING. THE PATIENT SUFFERED FROM MILD SECOND-DEGREE BURN THAT THEY BELIEVED WAS CAUSED BY THE BOVIE TIP. THE PATIENT WAS TREATED BY APPLYING BANDAGES TO THE AFFECTED AREAS.
Patient problems: Partial thickness (Second Degree) Burn; Unspecified Tissue Injury
MAUDE report 12666513 ↗MalfunctionAug 19, 2021Energy Output Problem; Failure to Align; Device Displays Incorrect Message; Connection Problem; Energy Output Problem; Failure to Align; Device Displays Incorrect Message; Connection Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE UNIT HAD STOPPED SEALING WHEN ON A PATIENT. BIOMED SPENT 25 MINUTES TROUBLESHOOTING THE UNIT WHICH ALSO HAD A REM ERROR. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12343765 ↗MalfunctionMar 15, 2021Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, DURING USE, THE SEAL FUNCTION WAS NOT WORKING CORRECTLY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11484884 ↗MalfunctionJan 22, 2021Self-Activation or Keying; Self-Activation or Keying
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT'S MONOPOLAR PORT 1 WAS ACTIVE WHEN NOTHING WAS PLUGGED INTO IT. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11213729 ↗MalfunctionMar 5, 2020Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE UNIT WAS NOT SEALING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Patient Involvement
MAUDE report 9795765 ↗MalfunctionAug 6, 2018Energy Output Problem; Self-Activation or Keying; Device Displays Incorrect Message; Connection Problem; Intermittent Energy Output; Activation Problem; Energy Output Problem; Self-Activation or Keying; Device Displays Incorrect Message; Connection Problem; Intermittent Energy Output
ACCORDING TO THE REPORTER, THE MONOPOLAR CUT MODE WAS ACTIVATED WHEN COAGULATE BUTTON WAS PRESSED.
Patient problems: No Consequences Or Impact To Patient; No Known Impact Or Consequence To Patient; No Patient Involvement
MAUDE report 7754011 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page