Edge E1452-6Recalls & reported problems
Covidien LP · Needle Electrode · Class 2
All models of Edge (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521713161), received between Jan 17, 2019 and Jan 11, 2024 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Arcing” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Arcing1
- Energy Output Problem1
- Material Fragmentation1
Patient problems
- No Consequences Or Impact To Patient2
- Partial thickness (Second Degree) Burn1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 11, 2024Energy Output Problem; Adverse Event Without Identified Device or Use Problem; Energy Output Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE MONOPOLAR DEVICE MADE THE ELECTRO WAVES JUMP. THE PROCEDURE WAS COMPLETED BUT PATIENT HAD A 2ND DEGREE SMALL BURN ON THE RIGHT CORNER OF THE LIP AND THE BURN WAS TREATED WITH CREAMS.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 18499174 ↗MalfunctionMar 4, 2019Arcing; Adverse Event Without Identified Device or Use Problem; Arcing; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, WHEN THE ELECTRODE NEEDLE WAS HOOKED ON TO THE GENERATOR, THE TIP SPARKED AND PRODUCED AN OPEN FLAME AND HAD TO BE REMOVED AND REPLACED. THE PATIENT WAS ON THE TABLE, BUT WAS NOT DIRECTLY INVOLVED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8387361 ↗MalfunctionJan 17, 2019Material Fragmentation; Material Fragmentation
ACCORDING TO THE REPORTER, DURING PROCEDURE, THE ORIGINAL BLADE ELECTRODE WAS CHANGED TO NEEDLE ELECTRODE. IT WAS NOTED THAT THE ELECTRODE WAS ASSEMBLED CORRECTLY PRE-OPERATIVELY. THE ELECTRODE DETACHED FROM THE PENCIL AND FELL INTO THE PATIENT CAVITY BUT IT WAS RETRIEVED. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8259270 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page