Edge E1465-4Recalls & reported problems
Covidien LP · Insulated Needle Electrode · Class 2
All models of Edge (43)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521713239), received between Nov 12, 2018 and May 19, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated2
- Detachment of Device or Device Component1
Patient problems
- Burn(s)1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 19, 2026Peeled/Delaminated; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING THE PROCEDURE, THE CLEAR INSULATED BOVIE TIP SHEATH HAS FALLEN OFF. THERE WAS NO PATIENT INJURY.
Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 25218136 ↗MalfunctionNov 12, 2018Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, THE SURGEON STARTED THE PROCEDURE WITH A SMOKE EVACUATION PENCIL AND AT ONE POINT DECIDED TO PUT THE ELECTRODE ON A STANDARD MONOPOLAR PENCIL. THE ELECTRODE COULD NOT BE REMOVED BY SURGEON'S HAND AND OPTED TO USE A KELLY CLAMP. SHORTLY AFTER THE ELECTRODE WAS SWITCHED BACK TO THE SMOKE EVACUATING PENCIL AND THIS TIME THE TECH REMOVED THE ELECTRODE USING THE KELLY CLAMP AS WELL. THIS TIME, THERE WERE BURNS NOTICED ON PATIENT'S RIGHT NIPPLE WITH FOUR SMALL BURNS A LESS THAN HALF OF ONE CENTIMETER IN SIZE EACH SO THE SURGEON INSPECTED THE ELECTRODE AND SAW CLAMP MARKS IN THE AREA OF THE FAILED PART OF THE INSULATION. A NEW ELECTRODE WAS OPENED AND USED WITHOUT ANY FURTHER ISSUE. BURNS SITE WERE TREATED WITH NEOSPORIN MEDICATION AFTER THE PROCEDURE.
Patient problems: Burn(s)
MAUDE report 8060923 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page