Valleylab E3781-28Recalls & reported problems
Covidien LP · Laparoscopic Curved Spatula Electrode Hollow · Class 2
All models of Valleylab (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884521767867), received between Feb 15, 2021 and Apr 25, 2023 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Electrical Shorting” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Electrical Shorting1
Patient problems
- Bowel Burn1
- Burn(s)1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 25, 2023Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC LOW ANTERIOR SURGERY, WHILE PERFORMING THE INTESTINAL MEMBRANE PROCESSING, MAKING AN INTESTINAL INCISION, THERE WAS OUTPUT (POWER LEAKAGE) FROM THE TIP OF THE INSULATED PART OF THE DEVICE. THE FAT OF THE PATIENT WAS BURNED/CAUTERIZED, WITH A SIZE OF 1MM, THE DEGREE OF THE BURN COULD NOT BE MEASURED, BUT IT WAS AN INCISION SITE SO THERE WAS NO BLEEDING, AND SINCE IT WAS IN THE POSITION WHERE THE INCISION WAS PLANNED, NO TREATMENT WERE PERFORMED FOR THE BURN. THE DEVICE WAS USED WITH CARE TO COMPLETE THE SURGERY. THERE WAS NO ADHESION OF ESCHAR AND THE LIKE.
Patient problems: Burn(s)
MAUDE report 16811084 ↗MalfunctionFeb 15, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC INTERSPHINCTERIC RESECTION (ISR), THE CYLINDRICAL ELECTRODE WAS EXPOSED EXCEPT FOR THE TIP ELECTRODE PART. WHEN THE TIP OF THE ELECTRODE TOUCHED THE CYLINDRICAL PART, THE INTESTINE AND MESENTERY WERE BURNED BY ABOUT 1 CM AND THERE WAS NO NEED TO REPAIR.
Patient problems: Bowel Burn
MAUDE report 11327457 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page