Valleylab LS10 VLLS10GEN (UDI 10884521780583)Recalls & reported problems
Covidien LP · LS Series Single Channel Vessel Sealing Generator · Class 2
All models of Valleylab LS10 (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 10884521780583), received between Jul 14, 2023 and Jun 20, 2025 — classified as 7 malfunctions.
The most frequently reported specific device problem is “Energy Output Problem” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Energy Output Problem7
- No Apparent Adverse Event3
- Connection Problem1
Patient problems
- Hemorrhage/Blood Loss/Bleeding5
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 20, 2025Energy Output Problem; No Apparent Adverse Event; Energy Output Problem; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE, THE DEVICE WAS NOT SEALING. THE LS10 EQUIPMENT AND POWER CABLE WERE EXCHANGED, BUT THE PROBLEM PERSISTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information
MAUDE report 22299193 ↗MalfunctionJun 20, 2025Energy Output Problem; No Apparent Adverse Event; Energy Output Problem; No Apparent Adverse Event
IT WAS REPORTED THAT DURING A NEPHRECTOMY PROCEDURE, THE DEVICE WAS NOT SEALING. THE LS10 EQUIPMENT AND POWER CABLE WERE EXCHANGED, BUT THE PROBLEM PERSISTED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; Insufficient Information
MAUDE report 22297025 ↗MalfunctionDec 10, 2024Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC RADICAL GASTRECTOMY, AT THE BEGINNING, THE DEVICE WAS UNABLE TO COAGULATE THE MESENTERIC BLOOD VESSEL ALTHOUGH THE HARDWARE SCREEN SHOWED THE CIRCLE WAS COMPLETED. VESSEL THICKNESS WAS 2-3 MILLIMETERS. ENERGY DELIVERY WAS NOTED AND NO ERROR TONES WERE HEARD. TRIED SEVERAL TIMES TO CHANGE THE LOTS OF DEVICE, GENERATORS, AND EVEN POWER CHARGING LINES BUT THE EFFECT WAS THE SAME. THERE WAS A BLOOD LOSS OF 20 MILLILITERS; NO BLOOD TRANSFUSION WAS REQUIRED. THE JAWS WERE CLEAN AND FRESH, WITHOUT ESCHAR. THE PROCEDURE WAS NOT COMPLETED IN A VERY PLEASANT MOOD. THE GENERATOR HAS BEEN UPDATED TO SOFTWARE 4.0.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 20892470 ↗MalfunctionApr 26, 2024Energy Output Problem; No Apparent Adverse Event; Energy Output Problem; No Apparent Adverse Event
ACCORDING TO THE REPORTER, DURING HEMICOLECTOMY, INABILITY TO COMPLETE SEALING AND BLEEDING ARE OBSERVED. THE DEVICE CONSTANTLY GAVE THE WARNING THAT THE SEALING HAS BEEN COMPLETED, BUT SEALING WAS NOT COMPLETED AND THE PATIENT BLEEDS AFTER CUTTING. THE VESSEL WAS NOT FULLY TRANSECTED USING THE FIRST DEVICE, WHILE VESSEL WAS FULLY TRANSECTED ON THE SECOND DEVICE. BLEEDING OBSERVED DIRECTLY FROM THE SEAL. NO SEAL WAS COMPLETED. NO BLOOD TRANSFUSION REQUIRED. ANOTHER DEVICE USED SUCCESSFULLY.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 19198209 ↗MalfunctionApr 25, 2024Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE LIGASURE DID NOT COAGULATE AND HAD NO SEAL. THERE WAS NO EVIDENCE OF ENERGY DELIVERY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19181677 ↗MalfunctionJan 26, 2024Energy Output Problem; Connection Problem; Energy Output Problem; Connection Problem
ACCORDING TO THE REPORTER, DURING OPEN PROCEDURE, THE HANDPIECE HAD NO SEAL (NO EVIDENCE OF ENERGY DELIVERY) AND WAS DIFFICULT TO ACTIVATE. THERE WAS A RE-GRASP ALERT AND ERROR TONE HEARD. THE JAWS WERE CLEANED WHEN THE HANDPIECE IS NOT AT USE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18589080 ↗MalfunctionJul 14, 2023Energy Output Problem; Energy Output Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, SURGEON STATED THAT THE FORCEPS DID NOT MAKE AN ADEQUATE SEAL AND AFTER SEALING THE TISSUE REMAINED BLEEDING. NO RE-GRASP ALERT, THERE WAS AN END TONE AND ENERGY DELIVERED. THE VLLS10GEN TEAM DID NOT REPORT INCOMPLETE SEAL.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17326028 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page