Valleylab VLFT10FXGENRecalls & reported problems
Covidien LP · FT10 Electrosurgical Generator · Class 2
All models of Valleylab (97)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 10884521829527), received between Jan 31, 2025 and Mar 21, 2025 — classified as 1 injury, 2 malfunctions.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
Patient problems
- Burn(s)1
- Partial thickness (Second Degree) Burn1
- Superficial (First Degree) Burn1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING AN EXCISION OF BASAL CELL CARCINOMA LESION ON THE RIGHT TEMPLE AREA OF THE PATIENT, THE PATIENT SUFFERED SUPERFICIAL BURNS DUE TO A FIRE. WHEN THE PATIENT WAS PLACED ON THE OPERATING TABLE, THE PATIENT RECEIVED LOW FLOW OXYGEN BY VENTIMASK. THE SURGEON INJECTED LIDOCAINE TO INCISION LINE. SURGEON USED CAUTERY TO BEGIN EXCISION OF SKIN CANCER. AFTER USING THE CAUTERY FOR THE INITIAL INCISION, A FIRE BROKE OUT IN THE VICINITY OF THE DRAPE AND VENTIMASK. IMMEDIATELY THE VENTIMASK WAS REMOVED AND THE OXYGEN WAS SHUT OFF. THE FIRE WAS QUICKLY EXTINGUISHED. ALL EQUIPMENT THAT WAS USED WERE REMOVED. SURGERY CONTINUED WITH NEW BOVIE, DRAPE AND INSTRUMENTATION. THE LESION OF CONCERN WAS REMOVED AND SENT FOR FROZEN SECTION UNTIL ALL MARGINS WERE CLEAR. BACITRACIN OINTMENT WAS APPLIED TO THE FACIAL BURN. THE PATIENT WAS DISCHARGED TO HOME IN GOOD CONDITION WITH PRESCRIPTIONS.
Patient problems: Superficial (First Degree) Burn
MAUDE report 21662430 ↗MalfunctionJan 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING PROCEDURE, THE SURGEON WAS WORKING ON THE HEAD AREA AND IT GOT BURNT. PATIENT HAD TO GO TO OR AFTER SURGERY TO TAKE CARE OF IT.
Patient problems: Burn(s)
MAUDE report 21285099 ↗InjuryJan 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING EXCISION LESION RIGHT TEMPLE UNDER MODERATE SEDATION, THE DOCTOR WAS REMOVING LESION USING NEEDLE TIP CAUTERY. SCRUB TECH STATED THAT THEY SAW A SPARK ¿JUMP¿ FROM THE CAUTERY TO THE DRAPE ON THE PATIENT¿S LEFT SIDE. THE DRAPE IGNITED, AND IMMEDIATELY THE STAFF AND SURGEON PULLED ALL DRAPES OFF THE PATIENT. PATIENT RECEIVED BURNS TO FOREHEAD AND LEFT CHEEK AREA. THE PATIENT WAS SENT TO THE ER (EMERGENCY ROOM) AFTER THE INITIAL PROCEDURE. AN INTERVENTION WAS NEEDED, AND THE PATIENT WAS REFERRED TO THE WOUND CENTER FOR A FOLLOW UP. ER (EMERGENCY ROOM) DOCTOR NOTED THE BURNS TO BE ¿PARTIAL THICKNESS, WITH NO AIRWAY OR OCULAR INVOLVEMENT.¿
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 21283193 ↗
Other Electrosurgical, cutting & coagulation & accessories devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page