Endo Shears 176643Recalls & reported problems

Covidien LP · Shears · Class 2

Quick summary

This Covidien Endo Shears is a hand-held surgical scissor used during laparoscopic surgery to cut and seal tissue. The 34 reports mostly describe the instrument failing to open or close, breaking, or not cutting properly, with a few mentions of the shaft tip burning and insulation flaking off. Two reports involve injuries; no deaths are described. There are no recalls naming this device, and the report count is holding steady at 3 over each of the past two 12-month periods.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
34
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 10884523000733), received between May 15, 2017 and Jan 21, 2026 — classified as 2 injuries, 32 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 2 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close13
  • Break6
  • Failure to Cut5
  • Mechanical Problem4
  • Adverse Event Without Identified Device or Use Problem3
  • Flare or Flash3
  • Dull, Blunt2
  • Melted2

Patient problems

  • No Clinical Signs, Symptoms or Conditions15
  • No Known Impact Or Consequence To Patient9
  • No Consequences Or Impact To Patient3
  • Burn(s)2
  • No Patient Involvement2
  • Unspecified Tissue Injury2
  • Burn, Thermal1
  • No Code Available1

Recent adverse event reports

The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 21, 2026Material Split, Cut or Torn

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, THE SHAFT SURFACE OF THE DEVICE WAS FOUND TO BE CRACKED. ANOTHER SAME TYPE OF DEVICE WAS USED TO COMPLETE THE CASE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24140550 ↗
  • MalfunctionJan 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC PROCEDURE, SPLIT INSULATION WAS OBSERVED ON THE DEVICE AND A HOLE WAS CAUSED IN THE GASTRIC ARTERY BY THE ENDO SHEARS. AN INSULATION ISSUE WAS ALSO IDENTIFIED WITH THE RETRACTABLE L HOOK, WHERE INSULATION FELL OFF OR DETACHED FROM THE SHAFT WHILE THE DEVICE WAS PLUGGED INTO A GENERATOR. ANOTHER LIGASURE WAS USED WITH THE SAME GENERATOR AND IT WORKED FINE. THE PATIENT EXPERIENCED A TEMPORARY INJURY AS A RESULT OF THE HOLE IN THE GASTRIC ARTERY, WHICH WAS MEDICALLY MANAGED BY CLIPPING THE ARTERY.

    Patient problems: Tissue Breakdown

    MAUDE report 24018178 ↗
  • MalfunctionDec 19, 2025Difficult to Open or Close; Difficult to Open or Close

    IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY, WHEN TESTED PRIOR TO ENTERING THE BODY. THE SHEARS PART OF THE INSTRUMENT WOULD NOT OPEN AND CLOSE. A NEW DEVICE WAS OPENED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23855073 ↗
  • MalfunctionJul 7, 2025Break; Break

    IT WAS REPORTED THAT WHEN THE DEVICE WAS OPERATED IN MONOPOLAR MODE, THE END OF THE SHAFT WAS BURNED. THIS ISSUE OCCURRED ON TWO DEVICES. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22427917 ↗
  • MalfunctionJul 7, 2025Break; Break

    IT WAS REPORTED THAT WHEN THE DEVICE WAS OPERATED IN MONOPOLAR MODE, THE END OF THE SHAFT WAS BURNED. THIS ISSUE OCCURRED ON TWO DEVICES. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22427807 ↗
  • MalfunctionJun 20, 2025Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE SHEARS CANNOT BE OPENED. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22300449 ↗
  • MalfunctionMar 26, 2024Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, PRIOR TO USE OF THE DEVICE, THE SURGEON REALIZED THAT THERE WAS DIFFICULTY IN CLOSING THE JAWS OF BOTH DEVICE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18980842 ↗
  • MalfunctionMar 26, 2024Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, PRIOR TO USE OF THE DEVICE, THE SURGEON REALIZED THAT THERE WAS DIFFICULTY IN CLOSING THE JAWS OF BOTH DEVICE. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18979138 ↗
  • MalfunctionOct 10, 2023Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC GASTRIC BAND REMOVAL, WHEN CUTTING ADHESION, THE DEVICE WOULD NOT CLOSE. THE DEVICE WAS REPLACED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17908682 ↗
  • MalfunctionJul 26, 2023Difficult to Open or Close; Physical Resistance/Sticking; Difficult to Open or Close; Physical Resistance/Sticking

    ACCORDING TO THE REPORTER, PRIOR TO A LAPAROSCOPIC UMBILICAL HERNIA REPAIR, THE HANDLES OF THE DEVICE WERE BINDING, IT WAS ALSO REPORTED THAT THE JAWS OF THE DEVICE WAS TIGHT AND WOULD NOT OPEN. ANOTHER DEVICE WAS USED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17402270 ↗
  • MalfunctionJun 30, 2023Flare or Flash; Flare or Flash

    PLI 10: ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HYSTERECTOMY, THE DEVICE WAS USED IN CUTTING THE VAGINAL CUFF AND RELEASING THE UTERUS, AND WHEN THE DOCTOR WAS FINISHING THE PROCEDURE, A SMALL FLAME IN THE CAVITY WAS OBSERVED. THE DEVICE WAS BURNED, AND A SMALL PART OF THE UTERUS (SPECIMEN) HAD A FIRST DEGREE BURN. THIS RESULTED IN TISSUE DAMAGE. IT WAS NOTED THAT THE BURN WAS NOT TREATED SINCE IT WAS ON THE SPECIMEN. TO COMPLETE THE CASE, AN ULTRASONIC DEVICE WAS USED.

    Patient problems: Burn(s); Unspecified Tissue Injury

    MAUDE report 17238234 ↗
  • MalfunctionJun 2, 2023Break; Break

    ACCORDING TO THE REPORTER, PREOPERATIVELY, THE TIP OF THE JAW WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17054471 ↗
  • InjuryJan 17, 2023Break; Misfire; Flare or Flash; Break; Misfire; Flare or Flash

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE INSULATED SHAFT HAD A BREAK IN THE INSULATION WHICH RESULTED IN A MISFIRE AND BURN ON THE PATIENT'S UTERUS.

    Patient problems: Burn(s); Unspecified Tissue Injury

    MAUDE report 16186914 ↗
  • MalfunctionJun 16, 2022Sparking; Sparking

    ACCORDING TO THE REPORTER, PRE OPERATIVELY, WHEN THE DEVICE WAS USED, A SPARK OCCURRED IN THE PROTECTION AREA THAT WAS MADE OF PLASTIC FROM THE TIP. ANOTHER PRODUCT WAS UTILIZED TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14715749 ↗
  • MalfunctionJul 19, 2021Mechanical Problem; Difficult to Open or Close; Mechanical Problem; Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING NEPHRECTOMY, WHEN TRYING TO USE THE DEVICE, IT WOULD NOT GRASP PROPERLY. THE DEVICE WOULD NOT CLOSE, LOCKED UP AND WOULD NOT CLOSE. THE CLAMP WAS REPLACED TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12194184 ↗
  • MalfunctionJul 6, 2021Break; Break

    ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC INGUINAL HERNIORRHAPHY, PRIOR TO USE THE DEVICE IN PATIENT, IT CAME DAMAGED INSIDE THE CLOSED PACKAGE. WHEN OPENING THE DEVICE IT WAS BROKEN AND MAKING OPENING AND CLOSING IMPOSSIBLE. THE DEVICE WAS REPLACED WITH A NEW PRODUCT TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12124600 ↗
  • MalfunctionMar 8, 2021Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, PRE-OPERATIVELY, THE DEVICE WAS UNABLE TO OPEN OR CLOSE. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11436783 ↗
  • MalfunctionMar 8, 2021Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, THE JAWS OF THE ENDOSCOPIC SHEARS WERE NOT OPENING. THERE WAS NO PATIENT INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11436740 ↗
  • MalfunctionNov 15, 2019Break; Break

    ACCORDING TO THE REPORTER, DURING THE USE OF THE MONOPOLAR ON CHOLECYSTECTOMY, THE PLASTIC COVERING THE BLADES BURNS. ANOTHER SCISSOR WAS USED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Information

    MAUDE report 9329626 ↗
  • MalfunctionJul 25, 2019Mechanical Problem; Difficult to Open or Close; Mechanical Problem; Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING GENERAL LAPAROSCOPIC PROCEDURE, WHILE CUTTING OF TISSUE, THE DEVICE SHEERS WOULD NOT OPEN. IT WAS ALSO NOTED THAT THE DEVICE DID NOT GRASP PROPERLY. ANOTHER SHEER WAS OPENED AND THE CASE CONTINUED WITHOUT INCIDENT. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8828040 ↗
  • MalfunctionJul 17, 2019Detachment of Device or Device Component; Detachment of Device or Device Component

    ACCORDING TO THE REPORTER, PREOPERATIVELY, THE CAP OF THE ELECTRIC DEVICE CONNECTION TERMINAL PART OF THE SURGICAL SHEARS WAS FOUND TO BE DISCONNECTED. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE.

    Patient problems: No Patient Involvement

    MAUDE report 8800948 ↗
  • MalfunctionApr 29, 2019Failure to Cut; Failure to Cut

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE DEVICE SCISSORS NOT CUT NOT CLOSE. ANOTHER DEVICE WAS OPENED TO COMPLETE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8560047 ↗
  • MalfunctionMar 4, 2019Mechanical Problem; Difficult to Open or Close; Mechanical Problem; Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC CHOLECYSTECTOMY, THE DEVICE WAS USED ON THE PATIENT, THE TIP WOULD NOT CLOSE. THE DEVICE DID NOT GRASP PROPERLY. THEY DISCARDED AND OPENED ANOTHER DEVICE TO COMPLETE THE CASE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8386162 ↗
  • MalfunctionDec 3, 2018Difficult to Open or Close; Difficult to Open or Close

    ACCORDING TO THE REPORTER, DURING A LAPAROSCOPIC HYSTERECTOMY, ON THE MOBILIZATION OF THE UTERUS, THE SURGEON REPORTED THE JAWS WERE TOO STIFF TO OPEN AND CLOSE. THE SURGEON OPENED A SECOND UNIT FROM THE SAME BOX WHICH ALSO HAD THE SAME PROBLEM. THEY OPENED A COMPETITOR¿S PRODUCT TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8127207 ↗
  • MalfunctionNov 12, 2018Melted; Melted

    ACCORDING TO THE REPORTER, DURING LAPAROSCOPIC CHOLECYSTECTOMY, THE SURGEON NOTICED MELTING ON THE SHAFT OF THE DEVICE AFTER ITS USE IN THE PROCEDURE. THE CASE WAS FINISHED WHEN IT WAS NOTICED. THERE WAS NO PATIENT INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8060156 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →