Valleylab E1502Recalls & reported problems
Covidien LP · Straight Extension · Class 2
Quick summary
The Valleylab E1502 is a Covidien straight extension used in electrosurgical cutting and coagulation. Most of the 33 reports describe the plastic or insulation surface peeling, cracking or delaminating — often noticed after sterilization — along with arcing and electrical shorting; several mention patients being burned during procedures. There are 5 injuries and no deaths. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 33
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 33 adverse event reports that list this exact device (DI 10884524000305), received between Oct 12, 2017 and Mar 18, 2024 — classified as 5 injuries, 28 malfunctions.
The most frequently reported specific device problem is “Peeled/Delaminated” (23 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Peeled/Delaminated23
- Adverse Event Without Identified Device or Use Problem14
- Electrical Shorting6
- Arcing5
- Contamination /Decontamination Problem5
- Scratched Material4
- Break3
- Detachment of Device or Device Component3
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- Burn(s)13
- Partial thickness (Second Degree) Burn8
- Full thickness (Third Degree) Burn5
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 25 most recent of 33 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 18, 2024Adverse Event Without Identified Device or Use Problem; Scratched Material; Adverse Event Without Identified Device or Use Problem; Scratched Material
ACCORDING TO THE REPORTER, DURING PROCEDURE, PATIENT WAS BURNT RELATIVE TO TWO OF THE SAME PART NUMBER ELECTRODES HAVING BEEN USED. ONE DEVICE HAD A BURN HOLE ON IT AND ANOTHER HAD ONE THAT WAS SMALLER IN VISUAL DETECTION.
Patient problems: Burn(s)
MAUDE report 18927514 ↗MalfunctionMar 18, 2024Adverse Event Without Identified Device or Use Problem; Scratched Material; Adverse Event Without Identified Device or Use Problem; Scratched Material
ACCORDING TO THE REPORTER, DURING PROCEDURE, PATIENT WAS BURNT RELATIVE TO TWO OF THE SAME PART NUMBER ELECTRODES HAVING BEEN USED. ONE DEVICE HAD A BURN HOLE ON IT AND ANOTHER HAD ONE THAT WAS SMALLER IN VISUAL DETECTION.
Patient problems: Burn(s)
MAUDE report 18927513 ↗MalfunctionSep 25, 2023Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE REALIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17808052 ↗MalfunctionSep 15, 2023Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE REALIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17761042 ↗MalfunctionSep 15, 2023Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE RILIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17761033 ↗MalfunctionSep 15, 2023Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STERILIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17761027 ↗MalfunctionSep 15, 2023Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STERILIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17761007 ↗MalfunctionSep 14, 2023Peeled/Delaminated; Contamination /Decontamination Problem; Peeled/Delaminated; Contamination /Decontamination Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE RILIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17749441 ↗MalfunctionSep 14, 2023Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE REALIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17749063 ↗MalfunctionSep 14, 2023Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE RILIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17745973 ↗MalfunctionAug 25, 2023Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, PRIOR TO USE, THE SAME CLIENT HAS OBSERVED THAT THE INSULATION WAS ALSO BEING CRACKED AFTER THE FIRST STE RILIZATION. THE DEVICE WAS NOT USED AS THE CRACKS WERE VISIBLE. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17626777 ↗MalfunctionJul 13, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING GYNECOLOGICAL SURGERY, PATIENT WAS OPERATED ON FOR THE FOLLOWING PROCEDURE "CONIZATION OF THE UTERINE CERVIX". SKIN BURN (BURN ON LEFT VULVAR LIP) DUE TO THERMAL DIFFUSION, USE OF ELECTRIC SCALPEL EXTENSION DURING CONIZATION PROBABLY DEFECTIVE. NO CONTACT BETWEEN THE TIP OF THE INSTRUMENT AND THE PATIENT'S SKIN. NO SPECIFIC ACTIONS. THE DEVICE CAUSED A BURN DUE TO A BREAK IN THE PROTECTIVE SHEATH. STERILIZATION ENSURES THAT THE SHEATH WAS VISUALLY VERIFIED PRIOR TO AUTOCLAVING. THE DEGREE OF BURN NOT KNOWN.
Patient problems: Burn(s)
MAUDE report 17316529 ↗MalfunctionMay 12, 2023Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, DURING BREAST AUGMENTATION AND MASTOPEXY, THE INSULATION HAD CRACKS AND THE PATIENT HAD A 2CM SECOND DEGREE BURN AT THE BREAST LINE. THE PATIENT WAS NOW FINE HOWEVER, IT HAD TO HAD THE BURN SURGICALLY RESECTED.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 16920605 ↗MalfunctionMay 12, 2023Peeled/Delaminated; Peeled/Delaminated
ACCORDING TO THE REPORTER, DURING BREAST AUGMENTATION AND MASTOPEXY, THE INSULATION HAD CRACKS AND THE PATIENT HAD A 2CM SECOND DEGREE BURN AT THE BREAST LINE. THE PATIENT WAS NOW FINE HOWEVER, IT HAD TO HAD THE BURN SURGICALLY RESECTED.
Patient problems: Partial thickness (Second Degree) Burn
MAUDE report 16920483 ↗MalfunctionFeb 6, 2023Break; Peeled/Delaminated; Device Damaged Prior to Use; Scratched Material; Break; Peeled/Delaminated; Device Damaged Prior to Use; Scratched Material
ACCORDING TO THE REPORTER, DURING PROCEDURE, THE DEVICE'S INSULATION WAS CHIPPED ON ONE PART. THE DEVICE WAS ALSO SCRATCHED AND HAD DENTS. THE CRACK WAS PROBABLY THERE AFTER THE STERILIZATION, BUT THE NURSE NOTICED THAT ON SURGERY. AFTER NOTICING THE ISSUE, THE DEVICE WAS REPLACED. NOTHING FELL ON PATIENT'S CAVITY AND THERE WAS NO MEDICAL INTERVENTION DUE TO THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16308574 ↗MalfunctionJan 4, 2023Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A LOOP OR LARGE LOOP EXCISION OF THE TRANSFORMATION ZONE (LLETZ) CONIZATION, THE PATIENT SUSTAINED A BURN FROM THE DEVICE BY ENERGY LEAKAGE ON A HEALTHY TISSUE AT THE VAGINAL WALL. THE DEGREE OF BURN WAS UNKNOWN. THERE WAS NO TREATMENT DONE; INSTEAD IT WAS ALLOWED FOR NATURAL HEALING.
Patient problems: Burn(s)
MAUDE report 16096465 ↗MalfunctionDec 29, 2022Arcing; Adverse Event Without Identified Device or Use Problem; Arcing; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A MAMMA AUGMENTATION, THE STRAIGHT ELECTRODE WAS DIFFERENT FROM THE PREVIOUS STRAIGHT ELECTRODE THEY BOUGHT OVER THE PAST 10 YEARS, THE NEW ELECTRODE LACKED INSULATION AND CAUSED LITTLE BURNINGS ON THE SKIN OF THE PATIENT DUE TO THE SPARKING/ARCING OF THE ELECTRODE, THE CUSTOMER ALSO STATED THAT THE NEW ELECTRODE HAVE A COLORING IN THE SAME BLUE PATTERN, WHEREAS THE ¿OLD ELECTRODES¿ HAVE TWO DIFFERENT BLUE COLORINGS (ON THE SHAFT DARKER THE REST NORMAL BLUE), TWO ELECTRODES WERE USED WITHIN THE PROCEDURE WHILE THE REST ARE BACK-TO-STOCK. THE BURN SITE WAS TREATED WITH NORMAL SKIN LOTION.
Patient problems: Burn(s)
MAUDE report 16074073 ↗MalfunctionDec 29, 2022Arcing; Adverse Event Without Identified Device or Use Problem; Arcing; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A MAMMA AUGMENTATION, THE STRAIGHT ELECTRODE WAS DIFFERENT FROM THE PREVIOUS STRAIGHT ELECTRODE THEY BOUGHT OVER THE PAST 10 YEARS, THE NEW ELECTRODE LACKED INSULATION AND CAUSED LITTLE BURNINGS ON THE SKIN OF THE PATIENT DUE TO THE SPARKING/ARCING OF THE ELECTRODE, THE CUSTOMER ALSO STATED THAT THE NEW ELECTRODE HAVE A COLORING IN THE SAME BLUE PATTERN, WHEREAS THE ¿OLD ELECTRODES¿ HAVE TWO DIFFERENT BLUE COLORINGS (ON THE SHAFT DARKER THE REST NORMAL BLUE), TWO ELECTRODES WERE USED WITHIN THE PROCEDURE WHILE THE REST ARE BACK-TO-STOCK. THE BURN SITE WAS TREATED WITH NORMAL SKIN LOTION.
Patient problems: Burn(s)
MAUDE report 16073702 ↗InjuryDec 26, 2022Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING PROCEDURE, FIVE ELECTRODES HAD PARTIALLY BROKEN INSULATION, SMALL CRACKS AND DENTS. THE INSULATION OF AN ELECTROCAUTERY EXTENSION TURNED OUT TO BE LEAKING, WHICH LED TO A SMALL-AREA CONTACT BURN IN THE AREA OF THE RIGHT BREAST IN A PATIENT. THERE COULD POTENTIALLY BE A WOUND HEALING DISORDER ON THE BREAST. IN THE WORST CASE IT COULD BREAST IMPLANT EXPOSED, WHICH WOULD RESULT IN AT LEAST TWO REVISION SURGERIES. MOSTLY MIXED SUPERFICIAL AND DEEP PARTIAL THICKNESS BURNS (SECOND DEGREE) WITH VERY SMALL AREAS OF THIRD DEGREE INJURIES. TREATMENT DONE WAS CONSERVATIVE TREATMENT.
Patient problems: Burn(s); Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn
MAUDE report 16046717 ↗InjuryDec 21, 2022Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING PROCEDURE, FIVE ELECTRODES HAD PARTIALLY BROKEN INSULATION, SMALL CRACKS AND DENTS. THE INSULATION OF AN ELECTROCAUTERY EXTENSION TURNED OUT TO BE LEAKING, WHICH LED TO A SMALL-AREA CONTACT BURN IN THE AREA OF THE RIGHT BREAST IN A PATIENT. THERE COULD POTENTIALLY BE A WOUND HEALING DISORDER ON THE BREAST. IN THE WORST CASE IT COULD BREAST IMPLANT EXPOSED, WHICH WOULD RESULT IN AT LEAST TWO REVISION SURGERIES.
Patient problems: Burn(s); Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn
MAUDE report 16028552 ↗InjuryDec 21, 2022Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING A PROCEDURE, 5 ELECTRODES HAD PARTIALLY BROKEN INSULATION, SMALL CRACKS AND DENTS. THE INSULATION OF AN ELECTROCAUTERY EXTENSION TURNED OUT TO BE LEAKING, WHICH LED TO A SMALL-AREA CONTACT BURN IN THE AREA OF THE RIGHT BREAST IN A PATIENT. THERE COULD POTENTIALLY BE A WOUND HEALING DISORDER ON THE BREAST. IN THE WORST CASE IT COULD BREAST IMPLANT EXPOSED, WHICH WOULD RESULT IN AT LEAST 2 REVISION SURGERIES. MOSTLY MIXED SUPERFICIAL AND DEEP PARTIAL THICKNESS BURNS (2ND DEGREE) WITH VERY SMALL AREAS OF 3RD DEGREE INJURIES. TREATMENT DONE WAS CONSERVATIVE TREATMENT.
Patient problems: Burn(s); Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn
MAUDE report 16028402 ↗InjuryDec 21, 2022Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING PROCEDURE, FIVE ELECTRODES HAD PARTIALLY BROKEN INSULATION, SMALL CRACKS AND DENTS. THE INSULATION OF AN ELECTROCAUTERY EXTENSION TURNED OUT TO BE LEAKING, WHICH LED TO A SMALL-AREA CONTACT BURN IN THE AREA OF THE RIGHT BREAST IN A PATIENT. THERE COULD POTENTIALLY BE A WOUND HEALING DISORDER ON THE BREAST. IN THE WORST CASE IT COULD BREAST IMPLANT EXPOSED, WHICH WOULD RESULT IN AT LEAST TWO REVISION SURGERIES. MOSTLY MIXED SUPERFICIAL AND DEEP PARTIAL THICKNESS BURNS (SECOND DEGREE) WITH VERY SMALL AREAS OF THIRD DEGREE INJURIES. TREATMENT DONE WAS CONSERVATIVE TREATMENT.
Patient problems: Burn(s); Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn
MAUDE report 16026420 ↗InjuryDec 21, 2022Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem; Peeled/Delaminated; Electrical Shorting; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, DURING PROCEDURE, FIVE ELECTRODES HAD PARTIALLY BROKEN INSULATION, SMALL CRACKS AND DENTS. THE INSULATION OF AN ELECTROCAUTERY EXTENSION TURNED OUT TO BE LEAKING, WHICH LED TO A SMALL-AREA CONTACT BURN IN THE AREA OF THE RIGHT BREAST IN A PATIENT. THERE COULD POTENTIALLY BE A WOUND HEALING DISORDER ON THE BREAST. IN THE WORST CASE IT COULD BREAST IMPLANT EXPOSED, WHICH WOULD RESULT IN AT LEAST TWO REVISION SURGERIES.
Patient problems: Burn(s); Full thickness (Third Degree) Burn; Partial thickness (Second Degree) Burn
MAUDE report 16026386 ↗MalfunctionNov 17, 2022Peeled/Delaminated; Arcing; Peeled/Delaminated; Arcing
ACCORDING TO THE REPORTER, DURING LIVER RESECTION, WHEN PROCESSING LIGAMENT AROUND LIVER, A SPARK OCCURRED ON THE EXTENSION AXIS ITSELF, AND THERE WERE TRACES OF ABLATION ALONG THE LONG AXIS FOR SEVERAL CENTIMETERS ON THE EXTENSION AXIS. THERE WAS SOMETHING LIKE A BLACK CHARRED LINE THAT SEEMED TO HAVE FORMED WHEN IT CAME INTO CONTACT WITH THE EXTENSION SHAFT. IT WAS REPLACED TO ANOTHER EXTENSION SHAFT AND USED. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15815419 ↗MalfunctionOct 24, 2022Peeled/Delaminated; Scratched Material; Peeled/Delaminated; Scratched Material
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, WHEN THE DEFECTIVE EXTENSION SHAFT WAS USED IN THE OP, IT SEEMS THAT THE PHYSICIAN CON FIRMED THAT THERE WAS A SCRATCH MARK ON THE SIDE. THERE WAS NO PATIENT INJURY. MEDTRONIC INITIAL INVESTIGATION FOUND THAT VISUAL INSPECTION FOUND SCRATCHES AND CRACKS TO THE INSULATION AND TROUBLESHOOTING CAN CONFIRM THE INSULATION DAMAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15660829 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.