Valleylab E2515HDBRecalls & reported problems

Covidien LP · Rocker Switch Pencil Holster · Class 2

All models of Valleylab (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 10884524000527), received between Jul 16, 2019 and Jul 20, 2024 — classified as 5 malfunctions.

The most frequently reported specific device problem is “Device Remains Activated” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Device Remains Activated2
  • Arcing1
  • Communication or Transmission Problem1
  • Connection Problem1
  • Fire1
  • Loose or Intermittent Connection1
  • Self-Activation or Keying1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient2
  • Burn(s)1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 20, 2024Device Remains Activated; Communication or Transmission Problem; Device Remains Activated; Communication or Transmission Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE WAS NOT DETECTED BY THE GENERATOR. THE DEVICE HAD ERROR CODE E322. THERE WAS NO PATIENT INVOLVEMENT. MEDTRONIC'S INITIAL EVALUATION OF THE INCIDENT, IT WAS IDENTIFIED THAT ONCE THE COAGULATION BUTTON WAS PRESSED, THE PENCIL WILL REMAINED ACTIVATED UNTIL THE PENCIL WAS UNPLUGGED FROM THE GENERATOR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19786029 ↗
  • MalfunctionDec 1, 2022Self-Activation or Keying; Connection Problem; Self-Activation or Keying; Connection Problem

    ACCORDING TO THE REPORTER, PRIOR TO USE, THE DEVICE WAS CONNECTED TO THE GENERATOR BUT THE GENERATOR DID NOT RECOGNIZED THE DEVICE. THERE WAS NO PATIENT INVOLVEMENT. THE MEDTRONIC INITIAL EVALUATION STATED THAT THE PENCIL HAD SELF-ACTIVATION. THE DEVICE HAD AN E322 ALARM, INDICATING THE GENERATOR BELIEVES A BUTTON IS BEING PRESSED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15898995 ↗
  • MalfunctionMay 27, 2020Device Remains Activated; Device Remains Activated

    ACCORDING TO THE REPORTER, THE DEVICE'S COAGULATION BUTTON STICKS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10090666 ↗
  • MalfunctionDec 2, 2019Fire; Loose or Intermittent Connection; Arcing; Adverse Event Without Identified Device or Use Problem; Fire; Loose or Intermittent Connection; Arcing; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING A PROCEDURE, 6.5" ELECTRODE AND PENCIL WAS USED DURING CAUTERY WHEN ARCING OCCURRED CAUSING FIRE ON THE SPONGE THAT WAS ON NEARBY METAL RETRACTOR. THE ELECTRODE WAS NOT FULLY SEATED IN THE PENCIL AND FORMED A SMALL GAP, WHERE A PORTION OF THE METAL ELECTRODE WAS EXPOSED BETWEEN PENCIL AND THE ELECTRODE INSULATION. THERE WAS NO PATIENT INJURY.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9397036 ↗
  • MalfunctionJul 16, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING TONSILLECTOMY, THE PENCIL WITH A DIFFERENT ELECTRODE HAD ARCING AND CAUSED A MILD MUCOSAL BURN NEAR THE LIP AREA OF THE PATIENT. ANOTHER ELECTRODE WAS USED TO COMPLETE THE PROCEDURE. PATIENT WAS HEALING WELL.

    Patient problems: Burn(s)

    MAUDE report 8794400 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →