Cool-tip ACT1520Recalls & reported problems
Covidien LP · RFA Single Electrode Kit · Class 2
All models of Cool-tip (16)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 10884524001180), received between Mar 1, 2017 and Apr 14, 2026 — classified as 4 injuries, 9 malfunctions.
The most frequently reported specific device problem is “No Flow” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- No Flow5
- Temperature Problem5
- Adverse Event Without Identified Device or Use Problem1
- Appropriate Device Problem Term/Code Not Available1
- Device Alarm System1
- Device Damaged Prior to Receipt by User1
- Failure to Fire1
- Material Discolored1
Patient problems
- No Consequences Or Impact To Patient5
- No Clinical Signs, Symptoms or Conditions4
- No Code Available3
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryApr 14, 2026No Flow; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PROCEDURE WAS NOT ABLE TO BE COMPLETED AND WAS ABORTED DUE TO THE DEVICE FAILED TO CIRCULATE.
Patient problems: Insufficient Information
MAUDE report 24872964 ↗MalfunctionJun 3, 2025No Flow; No Flow
IT WAS REPORTED THAT THE DEVICE FAILED TO CIRCULATE DURING THE PROCEDURE. THE PROCEDURE WAS ABLE TO BE COMPLETED WITH A NEW DEVICE. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22131635 ↗MalfunctionAug 28, 2024Overheating of Device; No Flow; Overheating of Device; No Flow
ACCORDING TO THE REPORTER, DURING A PROCEDURE, THE CONSUMABLE COULD NOT BE WATER CIRCULATED AND THE ABLATION NEEDLE OVERHEATED. THE PHYSICIAN REPLACED THE DEVICE AND CONTINUED THE SURGERY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20090654 ↗MalfunctionDec 17, 2021Material Discolored; Device Damaged Prior to Receipt by User; Adverse Event Without Identified Device or Use Problem; Material Discolored; Device Damaged Prior to Receipt by User; Adverse Event Without Identified Device or Use Problem
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE DEVICE AFTER OPENING THE PACKAGE, IT BECAME MOLDY AND OXIDIZES AND TURNED BLACK. THE PROCEDURE WAS CANCELLED THE PATIENT WAS NOT UNDER ANESTHESIA AND SCHEDULED THE NEXT DAY. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13024723 ↗MalfunctionSep 21, 2021Temperature Problem; Temperature Problem
ACCORDING TO THE REPORTER, PRIOR TO USE, EVEN WHEN THE NEEDLE WAS CONNECTED, THE TEMPERATURE FLUCTUATED. NO PROBLEM WAS DETECTED IN THE PROCEDURE ITSELF BECAUSE THE DEVICE MOVED WITHOUT ANY PROBLEMS WHEN A DIFFERENT NEEDLE WAS CONNECTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12506775 ↗InjuryJul 1, 2019No Apparent Adverse Event; No Apparent Adverse Event
ACCORDING TO THE REPORTER, DURING ABLATION, THE ANTENNA WAS NORMAL DURING USE, THE RANGE OF THE ABLATION HAD BEEN IN 2.6CM AND 2.3CM (LENGTH AND WIDTH) IN ACCORDING THE ORIGINAL INFORMATION. BUT THIS EVENT WAS ABNORMAL BECAUSE THE RESULT OF TREATING THE TUMOR DID NOT REACH THE ABLATION RANGE OF THE PRODUCT. THE DOCTOR WILL ABLATION AGAIN IN ANOTHER TIME FOR THE PATIENT. AN IMAGE WAS PROVIDED WITH DETECTED TUMOR.
Patient problems: No Code Available
MAUDE report 8750722 ↗MalfunctionJun 18, 2019No Flow; No Flow
ACCORDING TO THE REPORTER, DURING A LIVER TUMOR ABLATION AND AFTER THE PUNCTURE, THE WATER CIRCULATION WAS BLOCKED. THEY REPLACED THE PRODUCT TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8708282 ↗InjuryApr 22, 2019Failure to Fire; Failure to Fire
ACCORDING TO THE REPORTER, DURING ABLATION, THE DEVICE HAD NO POWER NOR ENERGY DELIVERY. THE SAME MACHINE REPEATEDLY TURNED ON SEVERAL TIMES, SHOWING THE SAME PROBLEM, AND AFTER REPLACING A NEW SET OF ELECTRODE NEEDLES, IT RETURNS TO NORMAL. THE PROCEDURE WAS ABORTED.
Patient problems: No Code Available
MAUDE report 8537045 ↗MalfunctionDec 18, 2018No Flow; No Flow
ACCORDING TO THE REPORTER, INTRAOPERATIVELY, THE CIRCULATING WATER PIPES ARE NOT SMOOTH. THERE WAS NO PATIENT INJURY.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8176554 ↗MalfunctionDec 18, 2018Device Alarm System; Temperature Problem; Device Alarm System; Temperature Problem
ACCORDING TO THE REPORTER, DURING ABLATION, THE DEVICE HAD AN ALARM PROBLEM SHOWING AN EXCESSIVE TEMPERATURE. REPLACED THE NEW NEEDLE AND CONTINUED WITH THE SURGERY. THERE WAS NO PATIENT INJURY.
Patient problems: No Code Available; No Consequences Or Impact To Patient
MAUDE report 8176527 ↗InjuryMay 9, 2017Temperature Problem; Temperature Problem
THE CUSTOMER REPORTED THAT THE ELECTRODE WAS SET ACCORDING TO THE MANUAL. IT WAS ACTIVATED BUT THE UNIT WOULD NOT START AND AN ¿IMPEDANCE OVER¿ ERROR WAS DISPLAYED REPEATEDLY. THE PROCEDURE WAS CANCELLED. THERE WAS NO HARM TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6554333 ↗MalfunctionApr 4, 2017Temperature Problem; Temperature Problem
THE CUSTOMER REPORTED THE TEMPERATURE WAS NOT DISPLAYED AFTER PUNCTURING. THERE WAS NO PATIENT HARM. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6457050 ↗MalfunctionMar 1, 2017Temperature Problem; Temperature Problem
THE CUSTOMER REPORTED THAT PRIOR TO USE ON A PATIENT, TEMPERATURE WAS NOT SHOWN ON THE DISPLAY UPON CONNECTING SUBJECT NEEDLE TO THE GENERATOR. ANOTHER NEEDLE WAS USED TO COMPLETE THE CASE. NO PATIENT INVOLVEMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6368415 ↗
Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices
- Vasoview Hemopro 2 VH-4000Maquet Cardiovascular LLC · Electrosurgical, Cutting & Coagulation & Accessories6 recalls1 death3,232 reports
- LigaSure LF1937Covidien LP · Electrosurgical, cutting & coagulation & accessories3 deaths2,090 reports
- Olympus WB91051WOLYMPUS Winter & Ibe GmbH · Electrosurgical, Cutting & Coagulation & Accessories1 death1,951 reports
- Valleylab FT10 VLFT10GENCovidien LP · Electrosurgical, cutting & coagulation & accessories1 recall5 deaths1,062 reports
- LigaSure LF1837Covidien LP · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall967 reports
- LigaSure LF2019Covidien LP · Electrosurgical, cutting & coagulation & accessories948 reports
- VAPR 225028Depuy Mitek, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES895 reports
- LigaSure LF1923Covidien LP · Electrosurgical, cutting & coagulation & accessories855 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page