Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 10884524001333), received between Jan 21, 2022 and Oct 5, 2023 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Peeled/Delaminated” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Peeled/Delaminated1

Patient problems

  • Full thickness (Third Degree) Burn1
  • Hemorrhage/Blood Loss/Bleeding1
  • Superficial (First Degree) Burn1
  • Unspecified Tissue Injury1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 5, 2023Peeled/Delaminated; Peeled/Delaminated

    ACCORDING TO THE REPORTER, DURING LAPAROSCOPY WITH BIOPSY OF THE PERITONEUM, BOVIE EXTENDER BURNED THE PERITONEUM WHILE LYING ACROSS IT, INDICATING FAULTY INSULATION. THE PRODUCT HAD BEEN THROUGH STERILE PROCESSING PRIOR TO USE. THE PATIENT HAD SUPERFICIAL (FIRST DEGREE) BURN. THERE WAS NO TREATMENT DONE FOR THE BURN SITE.

    Patient problems: Superficial (First Degree) Burn

    MAUDE report 17876685 ↗
  • InjuryJan 21, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ACCORDING TO THE REPORTER, DURING AN ATRAUMATIC CORONARY ARTERY BYPASS, THE LIGASURE COULD NOT ACTIVATE AS THERE WAS AN ERROR MESSAGE. THE PROCEDURE WAS THEN CONTINUED WITH AN ELECTRODE TOGETHER WITH A COMPETITOR'S PENCIL AND A GENERATOR WHERE IT WAS REPORTED THAT THE COMPETITOR'S PENCIL AND THE GENERATOR WORKED WITHOUT BEING ACTIVATED AND THAT THE PENCIL WAS WARM. AN ELECTRO SURGICAL CUT WITH NO BURNT EDGES WAS VISIBLE ON THE AORTA THAT CAUSED 1 LITER OF BLOOD LOSS AND HAD TO BE CLOSED VIA SUTURING.

    Patient problems: Full thickness (Third Degree) Burn; Hemorrhage/Blood Loss/Bleeding; Unspecified Tissue Injury