Skip to content

Valleylab E0020VRecalls & reported problems

Covidien LP · Bipolar Forceps Cord,Reusable · Class 2

All models of Valleylab (97)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 10884524001852), received between Apr 3, 2026 and Apr 3, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Difficult or Delayed Activation” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult or Delayed Activation1
  • Nonstandard Device1

Patient problems

  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2026Nonstandard Device; Difficult or Delayed Activation

    IT WAS REPORTED THAT DURING THE INITIAL LAPAROSCOPIC ASSESSMENT, SIGNIFICANT INTRA-ABDOMINAL ADHESIONS WERE IDENTIFIED PRIOR TO ROBOTIC DOCKING FOR A ROBOTIC HYSTERECTOMY WITH BILATERAL SALPINGECTOMY. IN ORDER TO SAFELY ADDRESS THESE ADHESIONS, LAPAROSCOPIC ADHESIOLYSIS WAS PERFORMED BEFORE PROCEEDING WITH ROBOTIC DOCKING. TO FACILITATE LAPAROSCOPIC ADHESIOLYSIS, THE HUGO FT10 CONNECTIONS WERE REMOVED, AND MONOPOLAR AND BIPOLAR FOOT PEDALS WERE CONNECTED AND UTILIZED. ADHESIOLYSIS WAS COMPLETED LAPAROSCOPICALLY WITHOUT COMPLICATION. FOLLOWING COMPLETION OF THIS PORTION OF THE PROCEDURE, THE MONOPOLAR AND BIPOLAR FOOT PEDALS WERE DISCONNECTED, AND THE HUGO FT10 CONNECTIONS WERE RECONNECTED IN PREPARATION FOR ROBOTIC DOCKING. FOLLOWING ROBOTIC ADHESIOLYSIS, A BLEEDING SITE WAS ENCOUNTERED THAT REQUIRED BIPOLAR COAGULATION. THE SURGEON ATTEMPTED TO ACTIVATE THE BIPOLAR FENESTRATED GRASPER (BFG) FROM THE CONSOLE. THE BFG APPEARED ACTIVE AT THE CONSOLE, AND THE FT10 CONFIRMED BIPOLAR ENERGY WAS ENABLED WITH THE ENERGY LEVEL SET TO 30. DESPITE THIS, NO BIPOLAR ENERGY WAS DELIVERED WHEN THE SURGEON ATTEMPTED TO FIRE FROM THE SURGEON CONSOLE. TROUBLESHOOTING WAS INITIATED INTRAOPERATIVELY.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 24779429 ↗

Other ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices

All ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page